Development and Validation of Integrated RP-HPLC and HPTLC Strategies for the Concurrent Estimation of Flunarizine Dihydrochloride and Propranolol Hydrochloride in Pharmaceutical Matrices
Priyadharshini N1, Malathi S2*
1The Tamil Nadu Dr. M.G.R Medical University, Guindy, Chennai – 600032, Tamil Nadu, India.
2Department of Pharmaceutical Analysis, PSG College of Pharmacy, Coimbatore - 641004, Tamil Nadu, India.
*Corresponding Author E-mail: malathis@psgpharma.ac.in
ABSTRACT:
This study presents the development of facile, robust analytical methodologies: RP-HPLC and HPTLC methods for simultaneously determining Flunarizine Dihydrochloride and Propranolol Hydrochloride in bulk and fixed-dose combination. For RP-HPLC, separation was achieved using XBridgeTM Phenyl 5µm (4.6 × 250mm) stationary phase employing Methanol as the solvent and binary eluent system consist of Acetonitrile: Methanol (50:50) ratio, delivered at a flow rate of 1 ml/min. Detection occurred at 267nm (Isoabsorptive Point), yielding the retention times of 2.9 minutes for Flunarizine dihydrochloride and 5.26 minutes for Propranolol hydrochloride. The analytes exhibited excellent linearity across the concentration ranges of 2-10mcg/ml for Flunarizine also 8 - 40mcg/ml for Propranolol. The LOD values of Flunarizine and Propranolol determined to be 0.1 and 0.5mcg/ml. The HPTLC method employed a developing solvent of Acetone: Methanol: Chloroform: Toluene (0.5:1:8:0.5) to elute the drugs, with the Rf Value of 0.89 for Flunarizine dihydrochloride and 0.15 for Propranolol hydrochloride. Linearity was observed in the range of 200 - 600ng/Spot for Flunarizine dihydrochloride and 800-2400ng/spot for Propranolol hydrochloride. Detection was carried out using 254nm. LOD of Flunarizine dihydrochloride and Propranolol hydrochloride were beginning 20-400ng/spot. Comprehensive validation, encompassing linearity, LOD, LOQ and recovery studies, sustained the precision, reliability, and analytical performance of both techniques. The outcomes underscore the applicability of the proposed methodologies for routine quality control and quantitative assessment of these analytes in pharmaceutical formulations and bulk substances.
KEYWORDS: HPLC, HPTLC, Flunarizine dihydrochloride, Propranolol hydrochloride, Validation Parameters.
INTRODUCTION:
Flunarizine dihydrochloride (FLU), chemically known as 1-[bis (4-fluorophenyl) methyl]-4-[(2E)-3-phenylprop-2-en-1-yl] piperazine dihydrochloride, represents a difluoro substituted piperazinyl derivative.
Flunarizine primarily acts as a calcium and sodium influx inhibitor and has been widely prescribed for migraine prophylaxis for more than three decades. Its therapeutic effect is attributed to lowering intracellular calcium levels by reducing spasms in arterial and arteriolar smooth muscles, thereby improving cerebral oxygenation. The compound exhibits a molecular mass of 477.4 gram mole-1, and recent studies suggest its potential role in protecting children from recurrent hemiplegia. Propranolol hydrochloride (PRH), a non-selective beta adrenergic antagonist, it is identified as [(1-methylethyl) amino]-3-(naphthalen-1-yloxy) propan-2-ol hydrochloride. It exhibits antianginal, antihypertensive, and antiarrhythmic activities by competitively inhibiting beta 1 and beta 2 adrenergic receptors in cardiac tissue. Its molecular weight is 295.8 gram mole-1. The combined use of FLU and PRH has been reported for effective in the prophylactic treatment of migraine1,2. In recent times, no analytical methodologies have been delineated for the formulation and quantification of both drugs in tablet pharmaceutical form. However, a literature survey indicates that there have been reports on the individual drugs as well as combinations with other drugs using UV spectroscopy3-17, HPLC18-34, and HPTLC35.
MATERIALS AND METHODS:
Materials:
Working standards of pharmaceutical agents of Flunarizine dihydrochloride procured from Carbanio chemicals, Hyderabad, and Propranolol hydrochloride obtained from Sigma Aldrich, Chennai. The marketed formulation was bought from the local market separately (MIGRABETA PLUS), comprises of 10 and 40 milligrams of FLU and PRH. HPLC-grade Methanol and Acetonitrile were procured from Sigma Aldrich, Chennai.
Chromatographic Condition:
RP- HPLC Method:
The concurrent assessment was conducted employing a Waters high performance liquid chromatography system, comprising a Waters 515 HPLC pump, an X Bridge® phenyl (5µm, 4.6 ×250mm), and a Waters 2489 UV detector, operated through Waters EmpowerTM 3 software. The eluent system consisted of Acetonitrile and Methanol in a ratio of (50:50). Prior to use, eluent was subjected to filtration via whatmann filter paper (No. 41), and subsequently degassed in an Ultrasonic cleanser. The elution rate was maintained at 1 ml/min, and detection was performed in HPLC-UV detector.
HPTLC technique:
Separation based evaluation was executed on aluminium - backed TLC plates precoated with silica gel 60F254 (20 × 20cm),which were preconditioned by washing with methanol and subsequently rendered active at 60°C to function as the stationary phase. An eluent comprising Acetone: Methanol: Chloroform: Toluene in the ratio (0.5:1:8:0.5) employed as the developing system for the elution of both the formulation. Sample application was performed using a CAMAG LINOMAT 5, equipped with WINCATs software. Plate development was undertaken in a twin-trough chamber with a chamber saturation period of 15 minutes. Chromatographic detection was accomplished at 254nm, serves as an analytical wavelength for drug quantification and detection.
Preparation of stock solution for both methods:
Accurately weighed quantities corresponding to 10 milligrams of FLU and 40 milligrams of PRH were separately transferred into 10mL calibrated flasks. Each sample was solubilized in methanol and brought to volume to produce stock solutions of 1000mcg/ml for FLU and 4000mcg/ml for PRH. Derived from the primary stock solutions, a series of appropriate dilutions were subsequently prepared to facilitate the construction curves for both the RP-HPLC and HPTLC strategies.
Formulation of working standard solutions for calibration curve development: RP-HPLC and HPTLC methodologies:
The above working standard solutions, were adjusted to the desired concentrations in 10mL calibrated flask using methanol, yielding ranges of 2 to 10mcg/ml for FLU, and 8 to 40mcg/ ml for PRH. Corresponding aliquots from these standard solutions were applied onto precoated HPTLC plates. This resulted in concentration ranges from 200 - 600ng per spot for FLU and 800-2400 ng per spot for PRH.The aforementioned concentration ranges employed to construct a calibration curve. Chromatographic development of the plates was conducted using the previously optimized eluent system, and the calibration plots were plotted as peak areas versus respective concentrations.
Preparation of the sample solution using the marketed formulation:
Twenty tablets were precisely weighed and finely triturated. A portion of the powdered mass equivalent to 10 and 40 milligrams of FLU and PRH was transferred to a 100ml graduated flask, solubilized in methanol, and sonicated for 20minutes, then brought to the volume with the same solvent. The solution was filtered using Whatmann (No: 41) filter paper, and suitable aliquots were prepared. A 0.2 milliliter volume of this solution was transferred into 10 mL calibrated flask and diluted to the mark, resulting in a solution containing 2mcg/ml of FLU and 8 mcg/ml of PRH with solvent. This solution was subsequently injected into the analytical column, and the chromatogram was recorded for RP-HPLC analysis. From the same filtrate, a 2µl aliquot was applied onto a precoated HPTLC plate, corresponding to 200ng per application of FLU and 800ng per application of PRH, and the formulation analysis outcomes are presented in Table 1.
Method Development technique in RP-HPLC:
A series of experiments and a Literature survey were carried out on various solvent ratios and scanning depths (253nm, 288nm and 267nm) were performed. Comparing the UV spectra for FLU and PRH at a wavelength of 267nm shows good absorbance for analysis. A signal-to-noise ratio, good resolution, sharp chromatogram with less tailing and fronting peaks were obtained in Acetonitrile: Methanol (50:50) in an X Bridge® phenyl 5µm column (4.6 ×250mm). Elution efficiency was achieved at a velocity of 1ml/min, resulting in retention times of FLU at 2.9 minutes and for PRH 5.26 minutes. The Corresponding results are illustrated in figures 1 - 3.
Figure 1: Overlain Spectra in Methanol
Figure 2(a): Chromatogram for Flunarizine dihydrochloride
Figure 2 (b): Chromatogram for Propranolol hydrochloride
Figure 3. Chromatogram for Formulation
HPTLC Method:
For HPTLC system, enumerate solvents were tried based upon the literature survey and trials conducted for better separation and resolution, among these Acetone: Methanol: Chloroform: Toluene in the ratio of (0.5:1:8:0.5) mobile phase gives a better reproducibility and well resolved peaks for FLU and PRH with a migration distance of 7cm and the Rf value of FLU and PRH were found to be 0.89 and 0.15. The resulting Chromatogram is shown in figures 4 and 5.
Figure 4(a): Chromatogram for Flunarizine dihydrochloride
Figure 4(b): Chromatogram for Propranolol hydrochloride
Figure 5: Chromatogram for Formulation
RESULTS AND DISCUSSION:
Method validation of RP-HPLC and HPTLC:
The analytical procedure was successfully validated for the concurrent determination of FLU and PRH in accordance with ICH guidelines Q2 (R1)36.
Linearity:
Linearity was assessed by visual inspection of the calibration plot signal vs concentration of analyte. The drug substances were appropriately diluted to prepare solutions across the required concentration ranges. FLU and PRH exhibited linearity within the concentration intervals of 2-10mcg/ml and 8-40mcg/ml, respectively, for RP-HPLC. In HPTLC analysis, 200-600ng per spot for FLU and 800-2400ng per spot for PRH. The corresponding data and graphical representations are presented in Table 2.
Limit of detection and Quantification:
The Limit of detection (LOD) and Limit of quantification (LOQ) were established by administering and analyzing various concentrations of the drugs. These parameters were assessed using visual evaluation, signal - noise ratio level, and the calibration curve and subsequently computed using the following formulas:
LOD = 3.3× (measure of variability/calibration line slope)
LOQ = 10× (measure of variability/calibration line slope)
Accuracy:
For accuracy studies, was evaluated by the standard addition method, in which a defined amount of analyte was supplemented into the sample. The outcomes were compared with those obtained from a well-known secondary procedure. It is performed by analysing nine replicate determinations across a minimum of three concentration levels corresponding to 80%, 100% and 120%.The percentage recoveries and % RSD were calculated.
Precision:
Repeatability and intermediate precision studies were used to assess the precision studies for both approaches. By injecting six replicates of the same concentration on the same day, repeatability was assessed over time under the same operational conditions. To evaluate consistency, the percentage relative standard deviation and the statistical dispersion were computed. To confirm the approach reproducibility in a variety of scenarios, intermediate precision was used. A summary of the findings may be found in Table 3.
Robustness:
Robustness is the capability of an analytical technique to generate consistent results under marginally different settings without compromising the anticipated result. These conditions were designed to reduce issues that arose during interlaboratory research. An assessment of robustness shows that the approach created is stable and consistent even when small conditions change. The corresponding empirical findings are articulated in Table 4.
System suitability:
It is an important analytical procedure; the core concept is based on evaluating the instrument performance, method operation, and sample analysis under optimized conditions. It confirms that the selected method, instrument, and reagents are appropriate for routine analysis of testing. Various parameters such as plate count (N), chromatographic resolution (Rs), retention time (Rt), asymmetry factor (T), distribution ratio (K), and retardation factor (Rf) are used to ensure the method reproducibility. It also ensures the generation of reliable, accurate data integrity by evaluating the overall performance of the analytical system before and during analysis. The experimental findings were satisfactory and documented in Table 5.
Table 1: Pharmaceutical analysis of the dosage form - HPLC and HPTLC
|
Drug |
Label Claim |
Amount Found (mg) |
% Assay |
% RSD * |
|||
|
HPLC |
HPTLC |
HPLC |
HPTLC |
HPLC |
HPTLC |
||
|
FLU |
10 mg |
9.85 |
9.76 |
98.5 |
97.6 |
0.9 |
0.7 |
|
PRH |
40 mg |
38.62 |
39.17 |
96.55 |
97.92 |
0.78 |
0.4 |
* RSD of three determinations
Validation parameters:
Table 2: Linearity graph of FLU and PRH – HPLC and HPTLC
|
HPLC |
||||
|
S.No |
FLU |
PRH |
||
|
Concentration (mcg/ml) |
Peak area |
Concentration (mcg/ml) |
Peak area |
|
|
1 |
2 |
10746 |
8 |
14683 |
|
2 |
4 |
15230 |
16 |
24060 |
|
3 |
6 |
19720 |
24 |
29865 |
|
4 |
8 |
23659 |
32 |
36589 |
|
5 |
10 |
27432 |
40 |
42165 |
|
HPTLC |
||||
|
S.No |
Concentration (ng / spot) |
Peak area |
Concentration (ng / spot) |
Peak area |
|
1 |
200 |
1623.8 |
800 |
1235.1 |
|
2 |
300 |
1645.9 |
1200 |
1334.5 |
|
3 |
400 |
1662.3 |
1600 |
1537.1 |
|
4 |
500 |
1682.7 |
2000 |
1647.7 |
|
5 |
600 |
1698.3 |
2400 |
1739.4 |
Table 3: Summary of Validation Parameters
|
Parameters |
RP - HPLC |
HPTLC |
||
|
Compound name |
FLU |
PRH |
FLU |
PRH |
|
Linearity |
2 -10 mcg/ml |
8-40 mcg/ml |
200-600 ng/spot |
800 – 2400 ng/spot |
|
Slope |
2090.1 |
843.66 |
0.1858 |
0.329 |
|
Intercept |
6817.1 |
9224.5 |
1588.3 |
971.12 |
|
Regression Coefficient |
0.9983 |
0.9905 |
0.9969 |
0.982 |
|
LOD (mcg/ml) |
0.1 |
0.5 |
20 ng/spot |
400 ng/spot |
|
LOQ (mcg/ml) |
0.5 |
1 |
100 ng/spot |
500 ng/spot |
|
Accuracy ( mean ±SD) |
0.1 |
0.5 |
0.4 |
0.8 |
|
Precision ( %RSD *) |
0.57 |
0.26 |
0.7 |
0.45 |
|
Repeatability ( %RSD *) |
1.2 |
0.90 |
0.72 |
1.10 |
* RSD of three determinations
Table 4: Robustness of FLU and PRH in HPLC
|
Condition |
Methods |
Drug |
Rt |
N |
T |
Rs |
|
Flow rate - 0.8 ml/min |
HPLC |
FLU PRH |
3.84 8.06 |
2950 3019 |
1.35 1.12 |
5.116 |
|
1.2 ml/min |
FLU PRH |
3.03 5.919 |
2286 2756 |
1.33 1.147 |
4.651 |
Table 5: System Suitability
|
Method |
Parameters |
FLU |
PRH |
Reference Value |
|
HPLC |
Retention time (Rt) |
2.951 |
5.526 |
- |
|
Tailing factor (T) |
1.615 |
1.128 |
≤ 1.5 |
|
|
Capacity factor (K-) |
4.52 |
1.95 |
1–10 |
|
|
Theoretical plates (N) |
4937 |
3750 |
N > 2000 |
|
|
HPTLC |
Retardation factor (Rf) |
0.88 |
0.1 |
- |
CONCLUSION:
This study presents the design and validation of HPLC and HPTLC methods for the simultaneous estimation of Flunarizine dihydrochloride and Propranolol hydrochloride in their combined pharmaceutical dosage form. The proposed methodologies were systematically validated in accordance with ICH regulatory guidelines. A key advantage of this method to detect both drugs at a lower concentration than those reported in earlier studies. The reproducibility of results, along with the simplicity of the procedures and reduced solvent requirements, further highlight the practicality of the methods. The resulting chromatogram provides clear evidence of the originality of the work, supporting their application for routine quality control surveillance under the specified chromatographic conditions.
CONFLICT OF INTEREST:
No conflicts of interest are associated with this work.
REFERENCES:
1. Shivarkar NA, Dudhe PB, Nagras MA. Development and validation of an HPTLC method for the simultaneous quantitation of flunarizine dihydrochloride and propranolol hydrochloride in capsule dosage form. Indian Journal of Pharmaceutical Sciences 2013; 75:364-368. doi: 10.4103/0250-474X.117418.
2. M.V. Francis, Flunarizine: a review of its role in migraine prophylaxis. International Journal of Research in Medical Sciences 2020; 8: 786-793.http://dx.doi.org/10.18203/2320-6012.ijrms20200276.
3. Chapke S.W. and Game M.D., UV spectrophotometric method development and validation for simultaneous determination of flunarizine dihydrochloride and propranolol hydrochloride in combined capsule dosage form. Asian Journal of Research in Chemistry 2013; 6: 506-508.
4. Busaranon K., Suntornsuk W, and Suntornsuk L., Comparison of UV spectrophotometric method and high performance liquid chromatography for the analysis of flunarizine and its application for the dissolution test. Journal of Pharmaceutical and Biomedical Analysis. 2006; 41: 158-164. https://doi.org/10.1016/j.jpba.2005.11.008.
5. Joshi Pv, Walke Tv. Development and Validation of a UV Spectrophotometric method for the simultaneous estimation of propranolol hydrochloride and flunarizine dihydrochloride in Pharmaceutical Dosage Form. World Journal of Pharmaceutical Research 2016; 5: 1445-1455.doi: 10.20959/Wjpr20167-6612.
6. Savaja BV, Kumar Patidarb A, Rajc HA. Simultaneous determination of Propranolol hydrochloride and Hydrochlorothiazide in tablet formulation using spectrophotometric technique (first derivative Method). Journal of Multidisciplinary Scientific Research. 2015; 3:10-13.
7. PV DS. Estimation of Etizolam and Propranolol in Combined Dosage Form by Spectroscopic Method.Asian J. Pharm. Ana. 2021; 11: 263-269. doi:10.52711/2231-5675.2021.00045.
8. Savaj BV, Patidar AK, Raj HA. Simultaneous determination of Propranolol hydrochloride and Hydrochlorothiazide in tablet formulation using spectrophotometric technique (Simultaneous Equation Method). Asian J. Pharm. Res 2015; 5: 36-40. doi: 10.5958/2231-5691.2015.00005.2
9. Bhadresh V. Savaj, Ashutosh Kumar Patidar, Hashumati A. Raj. Simultaneous determination of Propranolol hydrochloride and Hydrochlorothiazide in Tablets formulation using spectrophotometric technique (Simultaneous Equation Method). Asian J. Pharm. Ana. 2014; 5(1): 36-40.
10. Ajay I. Patel, Shweta V. Gosai, Amikumar J. Vyas, Nilesh K. Patel, Ashok B. Patel. UV Spectrophotometric Estimation of Propranolol Hydrochloride and Rosuvastatin Calcium in Synthetic mixture using First Order Derivative Method. Asian J. Pharm. Ana. 2020; 10(4): 189-194
11. Susmeena Tabassum Kapatrala, Vinod Kumar Kondreddy, Swapna Kandlapalli, Tejaswi Male. Method Development and Validation for the Estimation of Etizolam and Propranolol Hydrochloride in bulk and Combined Dosage Forms by Simultaneous and Derivative Spectroscopic Methods. Asian Journal of Pharmaceutical Analysis. 2021; 11(4): 263-269.
12. Bhadresh V. Savaj, Ashutosh Kumar Patidar, Hashumati A. Raj. Simultaneous Determination of Propranolol Hydrochloride and Hydrochlorothiazide in Tablets Formulation using Spectrophotometric Technique (First Derivative Method). Asian J. Pharm. Res. 2015; 5(1): 31-36.
13. Sangita W. Chapke, Madhuri D. Game. UV Spectrophotometric Method Development and Validation for Simultaneous Determination of Flunarizine Dihydrochloride and Propranolol Hydrochloride in Combined Capsule Dosage Form. Asian J. Research Chem. 2013; 6(5): 506-508.
14. Khan Arshad Bashir, NG Nanjundaswamy. Formulation and Evaluation of Sustained Release Matrix Tablets of Propranolol Hydrochloride Using Hydroxyethyl Guar as Rate Sustaining Polymer. Research J. Pharma. Dosage Forms and Tech. 2009; 1(3): 236-239.
15. Amir Alhaj Sakur, May Karman. Three Spectrophotometric Methods Validated for Handling of Overlapped Spectra: Application for Determination Propranolol Hydrochloride and Hydrochlorothiazide in Dosage Forms and spiked biological fluids. Research Journal of Pharmacy and Technology. 2024; 17(11): 5291-5298.
16. Sangita W. Chapke, Madhuri D. Game. Absorption Correction Method for Simultaneous Spectrophotometric Estimation of Flunarizine Dihydrochloride and Propranolol HCl in Combined Capsule Dosage Form. Asian J. Research Chem. 2013; 6(5): 514-516.
17. M. A. Shende, P. K. Gupta. Formulation Development of Propranolol Hydrochloride Sustained Release Prunus amygdalus Gum Based Mucoadhesive Buccal Matrices. Asian J. Pharm. Tech. 2017; 7 (4): 181-188.
18. Mahesh. M, Sridhar Thandra, Shaik Muneer, B. Siva Sai Kiran, H. Mamatha. Method Development and Validation of RP-HPLC Method for the Simultaneous Estimation of Propranolol and Hydralazine in Pharmaceutical Dosage Form. Asian J. Pharm. Ana. 2019; 9(1): 37-42.
19. Akshay V. Dangariya, Nilesh K. Patel. Development and Validation of a Stability Indicating HPLC Method for the Simultaneous Determination of Flunarizine and Domperidone in Tablet Dosage Form. Asian Journal of Pharmaceutical Analysis. 2025; 15:25-0.doi: 10.52711/2231-5675.2025.00005.
20. O'Connor N, Geary M, Wharton M, Sweetman P. Development and validation of a rapid chromatographic method for the analysis of flunarizine and its main production impurities. J Pharm Anal. 2013; 3: 211-214. doi:10.1016/j.jpha.2012.12.005.
21. Lucida H, Yelli F. Development and Validation of Analytical Methods and Uniformity Content of Flunarizine Hydrochloride in Tablet Preparations. World Journal of Pharmacy and Pharmaceutical Sciences. 2020; 9: 55-66. doi: 10.20959/wjpps20207-16450.
22. Imam SS, Ahad A, Aqil M, Sultana Y, Ali A. A validated RP-HPLC method for the simultaneous determination of propranolol and valsartan in bulk drug and gel formulation. J Pharm Bioallied Sci. 2013; 5: 61-65. doi:10.4103/0975-7406.106573.
23. Srikanth MV, Ram BJ, Sunil SA, Rao NS, Murthy KR. Development and validation of an HPLC method for estimation of propranolol HCl in human plasma. J. Sci. Ind. Res. 2012; 71: 102-123.
24. SHAIK N. A precise RP-HPLC method for the simultaneous estimation of diazepam and hydrochloride in tablet dosage form. Journal of Drug Delivery and Therapeutics. 2012; 2:97-101.
25. Peraman R, Bandi J, Kondreddy VK, Kalva B, Kothakota SG, Paritala J, Nagappan K, Yirgamreddy PR. Analytical quality by design approach versus conventional approach: Development of HPLC-DAD method for simultaneous determination of etizolam and propranolol hydrochloride. Journal of Liquid Chromatography and Related Technologies. 2021; 44: 197–209. doi.org/10.1080/10826076.2021.1874982.
26. Shabir GA. Development and validation of an RP-HPLC method for the determination of methamphetamine and propranolol in tablet dosage form. Indian Journal of Pharmaceutical Sciences. 2011; 73: 430-435. doi.org/10.4103/0250-474X.95632.
27. Torul H, Tamer U. Determination of enantiomers of atenolol and propranolol in pharmaceutical formulations by HPLC. Journal of AOAC International. 2011; 94: 833-838.
28. Imam SS, Ahad A, Aqil M, Sultana Y, Ali A. A validated RP-HPLC method for simultaneous determination of propranolol and valsartan in bulk drug and gel formulation. Journal of Pharmacy and Bioallied Sciences. 2013; 5: 61-65.
29. Modamio P, Lastra CF, Marino EL. Error structure for the HPLC analysis for atenolol, metoprolol, and propranolol: a useful weighting method in parameter estimation. Journal of Pharmaceutical and Biomedical Analysis. 1998; 17: 507-513. https://doi.org/10.1016/S0731-7085(97)00236-7.
30. Salman SA, Sulaiman SA, Ismail Z, Gan SH. Quantitative determination of propranolol by ultraviolet HPLC in human plasma. Toxicology Mechanisms and Methods. 2010; 20: 137-142. doi.org/10.3109/15376511003602112.
31. Venkatesh G, Ramanathan S, Mansor SM, Nair NK, Sattar MA, Croft SL, Navaratnam V. Development and validation of RP-HPLC-UV method for simultaneous determination of buparvaquone, atenolol, propranolol, quinidine and verapamil: a tool for the standardization of rat in situ intestinal permeability studies. Journal of Pharmaceutical and Biomedical analysis. 2007; 43: 1546-1551. doi.org/10.1016/j.jpba.2006.11.013.
32. El‐Sherbiny DT, Eid MI, El‐Wasseef DR, Al‐Ashan RM, Belal F. Analysis of flunarizine in the presence of some of its degradation products using micellar liquid chromatography (MLC) or microemulsion liquid chromatography (MELC)–application to dosage forms. Journal of Separation Science. 2005; 28: 197-202. doi.org/10.1002/jssc.200401830.
33. Wahbi AA, El-Walily AF, Hassan EM, Saliman FG, El-Gendi A. Liquid chromatographic determination of flunarizine dihydrochloride in the presence of its degradation product. Journal of Pharmaceutical and Biomedical Analysis. 1995; 13: 777-784. doi.org/10.1016/0731-7085(95)01266-N
34. Chabala OR, Haque Md S, Thirumoorthy DA. Stability-Indicating Liquid Chromatographic Method Development for the Simultaneous Determination of Amitriptyline Hydrochloride and Propranolol Hydrochloride in Tablet Dosage Form. Journal of Chromatographic Science 2025; 63: bmae060.https://doi.org/10.1093/chromsci/bmae060.
35. Bhavar G, Chatpalliwar VA. Quantitative analysis of propranolol hydrochloride by high-performance thin-layer chromatography. Indian J Pharm Sci. 2008; 70: 395-398. doi:10.4103/0250-474X.43016.
36. Borman P. Elder D. Q2 (R1) Validation of Analytical Procedures: Text and Methodology. ICH Qual Guidel an Implement Guid. 2017; 127 - 166.
|
Received on 09.09.2025 Revised on 03.01.2026 Accepted on 12.03.2026 Published on 08.08.2026 Available online from August 12, 2026 Research J. Pharmacy and Technology. 2026;19(8):3547-3552. DOI: 10.52711/0974-360X.2026.00502 © RJPT All right reserved
|
|
|
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. Creative Commons License. |
|