Determination of Rizatriptan Benzoate in Rat Plasma by Liquid Chromatography with Application to Pharmacokinetic Study

 

Priyanka Gawarkar-Patil, Vividha Dhapte-Pawar*

Department of Pharmaceutics, Poona College of Pharmacy,

Bharati Vidyapeeth (Deemed to be University), Pune, Maharashtra 411038, India.

*Corresponding Author E-mail: vividhadhapte@gmail.com

 

ABSTRACT:

Rizatriptan benzoate (RBZ), a member of the triptan class, is frequently prescribed in migraine therapy due to its high selectivity towards 5-hydroxytryptamine (5-HT1B, 5-HT1D) receptors. Despite its efficacy, the therapeutic dose (typically 5–10 mg) results in relatively low plasma concentrations. Such nanogram-level concentrations require highly sensitive analytical methods for accurate detection in biological matrices. A simple, cost-effective and sensitive ‘high-performance liquid chromatography’ (HPLC) method was thus, developed and validated for the quantification of RBZ in rat plasma. The chromatographic separation was accomplished using an isocratic mobile phase of 10 mM anhydrous dipotassium hydrogen orthophosphate (K2HPO4) pH 3.5 and methanol (MeOH) in the ratio of 80:20 (v/v), with ultra-violet (UV) detection at 225 nm. The developed method was validated in accordance with the ‘International Council for Harmonisation’ (ICH) M10 guidelines, which included examining specificity, linearity, recovery, accuracy, precision, limit of detection (LOD) and quantification (LOQ), robustness as well as stability. This validated method was applied in the pharmacokinetic investigation of Rizatriptan benzoate (5 mg/kg) in Sprague-Dawley rats after single oral administration. The method demonstrated excellent linearity (5–1500 ng/mL, R˛ = 0.99), with retention-time (Rt) of 6 minutes. Methanol was identified as the most effective extraction solvent, offering 95.81% recovery. This validated, developed HPLC method was robust, reliable and sensitive, making it apt for routine study of RBZ in preclinical studies emphasizing its translational relevance.

 

KEYWORDS: Rizatriptan benzoate, HPLC, Bioanalytical method validation, Rat plasma, Pharmacokinetics.

 

 


INTRODUCTION: 

The chemical compound Rizatriptan benzoate is termed N, N dimethyl-5-(1H-1, 2, 4-triazol -1-ylmethyl)-1H indole-3-ethanamine monobenzoate (Figure 1), which functions as a selective agonist for the 5-hydroxytryptamine 1B/1D receptor and is employed in the management of acute headaches associated with migraine1.

 

Rizatriptan has a poor affinity towards the 5-HT7 receptor and other 5-HT1 receptor subtypes, including 5-HT1A, 5-HT1E and 5-HT1F, but a substantial affinity for human cloned 5-HT1B as well as 5-HT1D receptors. At 5-HT2, 5-HT3, α- and β-adrenergic, dopaminergic, muscarinic, histaminergic and benzodiazepine receptors, it has minimal or no discernible activity2–5. According to current beliefs, the dilatation of cranial blood vessels and/or release of pro-inflammatory and vasoactive peptides from the sensory nerve endings within a trigeminal system are the causes of headaches2.

 

Figure 1: Chemical structure: Rizatriptan benzoate

 

There are several dosage forms of RBZ in the market, including oral tablets and orally disintegrating tablets. A very low dosage of RBZ (5-10 mg) for treatment is recommended clinically, resulting in a less therapeutic plasma concentration (21-30 ng/mL)6,7. A highly sensitive analytical method is essential for accurately quantifying the low RBZ plasma concentrations in clinical conditions. This is accomplished via a variety of analytical methods, including reverse-phase liquid chromatography (RP-LC)2, reverse-phase high-performance liquid chromatography (RP-HPLC)1,8, liquid chromatography-tandem mass spectrometry (LC-TMS/MS)9 and UV spectrophotometry10 have been employed to quantify RBZ in biological fluids and pharmaceutical formulations. RP-HPLC11, UV12 and LC-MS/MS11,13 methods have been employed to simultaneously quantify RBZ alongside other drug components in combination formulations and biological matrices.

 

Many existing methods for RBZ analysis in plasma are associated with prolonged chromatographic run times and high consumption of organic solvents and buffer solutions, limiting their practicality in certain settings. Consequently, there is a demand for an analytical technique that is straightforward, reliable, sensitive, accurate and economical for the quantification of RBZ in blood plasma.

 

For the measurement of RBZ in rat plasma, accurate, precise, sensitive and economical HPLC technique was created and validated in the current study. The validation technique was carried out in accordance with regulatory standards and assessed system suitability, recovery, accuracy, precision, specificity, limits of detection (LOD) and quantification (LOQ) and linearity. The applicability of the validated approach for in-vivo bioanalytical applications was next assessed by applying it to a pharmacokinetic investigation of RBZ in male Sprague-Dawley rats (SD).

 

MATERIALS AND METHODS:

Rizatriptan benzoate was kindly provided as a gift sample from Cipla Ltd. (Kurkumbh, Pune. India). HPLC-grade methanol (MeOH) and acetonitrile (ACN) were sourced from Merck Chemicals. Anhydrous dipotassium hydrogen phosphate was purchased from Loba Chemie Pvt. Ltd. Ultrapure water produced through Milli-Q purifying system was used in all experiments.

 

Methods:

Equipment and Chromatographic Methodology:

A Jasco HPLC system was employed for chromatographic separations. This setup included a PU-2080 pump and a UV/Vis detector (model UV-2075). Data acquisition and interpretation were performed using Borwin version 1.5 software. RBZ was measured using Thermo Scientific C18 column (250 x 4.6 mm, 5 μm). To ensure phase purity, mobile phases were filtered using 0.22 µm Ultipor® nylon membrane filters (Pall Life Sciences, India) and degassed via an ATS-2 LED ultrasonic bath sonicator (Athena Technology). A digital pH meter (Systronics Model 802, India) was used to adjust the aqueous phase.

 

Preparation of Standard Solution and Selection of Analytical Wavelength:

Various quantities of RBZ, (5, 10, 50, 100, 500, 1000, 1500 ng/mL) were prepared in rat plasma by diluting a methanolic stock solution of RBZ i.e., (1 mg/mL). To access the procedure for accuracy and precision, quality-control (QC) samples were generated at three various concentration levels: low level (LQC, 5 ng/mL), medium level (MQC, 100 ng/mL) and high level (HQC, 1500 ng/mL). A separate working solution of RBZ (10 µg/mL) was made from the primary stock. To determine the maximum absorbance wavelength (λmax) for RBZ, a UV-Visible spectrophotometer (Jasco V-730, Japan) was used to scan the solution from 200-400 nm.

 

Preparation of Plasma Sample:

All animal-related procedures were authorized by the ‘Institutional Animal Ethics Committee’ (IAEC) of Poona College of Pharmacy, Pune, under the protocol number PCP/IAEC/2024/3-29. Male Sprague-Dawley (SD) rats (230–250 g) were sourced from Lacsmi Biofarms Pvt. Ltd., Pune and acclimatized for one week. The animals were fed a regular pellet diet and had unlimited access to water. Following single oral administration of RBZ (5 mg/kg), blood was obtained at specified time intervals of 0.5, 1, 2, 4 and 6 hours via EDTA-coated tubes. Centrifugation at 7000 rpm for 10 minutes was done to separate the plasma, which was then kept at -20°C for further analysis14.

 

Solvent Selection for Extraction:

To maximize drug recovery from biological matrices, the extraction solvent choice is crucial. After vortexing the frozen plasma, it was allowed to thaw at normal temperature. To extract RBZ from plasma, a number of solvents were explored, including MeOH, ACN and the mobile phase. Once at room temperature, 2mL Eppendorf tubes holding 100 µL of plasma were filled with 200 µL of the RBZ solution and 700 µL of the chosen extraction solvent. To facilitate protein precipitation, the resultant mixture was vortexed for 2-3 minutes. The samples were then centrifuged for 20 minutes at 4°C and 15,000 rpm. Before being injected into the HPLC system for analysis, 20 µL of the final mixture which consisted of 100 µL of the recovered supernatant and 900 µL of the extraction solvent went through a 0.22 µm syringe filter15,16.

 

Method Validation:

HPLC technique was validated following the ICH M10 guidelines for bioanalytical method validation, evaluating important factors such as linearity, specificity, recovery, accuracy, precision, robustness, along with LOD and LOQ to ensure it is appropriate for the intended use17,18.

 

System Suitability and Specificity:

A drug combination at a concentration of 100 ng/mL was introduced into the HPLC system six times to determine the suitability of the analytical method for its intended application. To evaluate the system's ability to adequately separate the analytes, three replicates of the diluent, blank plasma and spiked plasma were also analyzed.

 

Linearity:

A calibration curve (CC) for RBZ was created by graphing the peak areas against the related drug concentrations. Suitable dilutions were performed from a starting stock solution of 5000 ng/mL to establish RBZ concentration of 5, 10, 50, 100, 500, 1000 as well as 1500 ng/mL.

 

Precision:

Precision measurements were used to assess the reproducibility of the established bioanalytical approach. By adding known concentrations of RBZ to blank plasma matrices, it was assessed at three QC levels: LQC(5 ng/mL), MQC (100 ng/mL) as well as HQC (1500 ng/mL). Analysis was done at each QC level to look at intraday, inter-analyst and interday precision.

 

Accuracy:

Three QC levels—LQC (5 ng/mL), MQC (100 ng/mL), along with HQC (1500 ng/mL) were employed to evaluate the enhanced procedure's accuracy. Each sample was injected into the ‘HPLC’. In order to back-calculate the actual concentrations, RBZ peak areas were examined using linear regression algorithms. By contrasting the computed concentrations with their corresponding theoretical values, accuracy was ascertained.

 

LOD, LOQ and Recovery:

The slope of the CC and the standard deviation (SD) of the ‘y’ intercepts from the regression lines were employed to compute the LOD as well as LOQ. RBZ-spiked quality control (QC) samples were made at LQC (5 ng/mL), MQC (100 ng/mL) with HQC (1500 ng/mL) levels in order to assess the accuracy of the procedure using recovery tests. Following the replication of the HPLC analysis, the amount of RBZ in each QC sample was quantified. By contrasting the measured concentrations with the initial spike levels, recovery was ascertained.

 

Robustness:

Deliberate changes were made to some parameters in order to test the consistency of the outcomes and determine how resilient the optimized approach was. While maintaining all other parameters constant, individual factors were changed one at a time, such as the detecting wavelength (225 ± 2 nm) and flow rate (0.8 ± 0.2 mL/min). Response metrics such as Rt, asymmetry factor, peak area and theoretical plates (N) were observed to track changes in method performance19.

 

Stability Study:

Samples were put through three freeze-thaw cycles with QC levels LQC, MQC and HQC assessed in triplicate to evaluate the stability of RBZ in plasma. Three portions of each concentration were kept for a full day at -20°C. After that, the samples were allowed to naturally thaw at normal temperature. They were frozen for an additional 24 hours after fully thawing. Further, two more freeze-thaw cycles were conducted to assess the drug's stability17.

 

Pharmacokinetic Study in Sprague-Dawley Rats:

Healthy male Sprague-Dawley (SD) rats weighing 230-250 g and 6-8 weeks of age were procured from Lacsmi Biofarms Pvt. Ltd., Pune. One week prior to the trials, the animals were acclimated to the lab environment. They were given free access to water and a typical pellet food (Nutrivet, Pune). All animal procedures were approved by the Institutional Animal Ethics Committee (IAEC) of Poona College of Pharmacy, Pune (Protocol No. PCP/IAEC/2024/3-29) and conducted in compliance with CPCSEA guidelines for the care and use of laboratory animals.

 

Study Design and Sample Collection:

A group of nine (n=9) male SD rats received single oral dose of pure RBZ (5 mg/kg body weight) suspended in distilled water. The dose was determined based on prior pharmacokinetic studies conducted using RBZ in different animal models. Rats under isoflurane anaesthesia had their blood samples (0.5-1.0 mL) drawn into 4 mL EDTA-coated tubes (Henso, China) at prearranged intervals of 0.5, 1, 2, 4 and 6 hours. Samples were collected and then centrifuged using a Cryocentrifuge-2810R (Eppendorf, USA) at 10,000 rpm for 20 minutes at 4°C. Before undergoing additional examination, the separated plasma was kept at -20°C. Each plasma sample was processed and analyzed for RBZ content using a validated bioanalytical method, followed by HPLC quantification. Analysis of pharmacokinetic parameter was performed by PKSolver, Microsoft Excel add-in (version 2.0). The pharmacokinetic profiles of tmax, Cmax and AUC0-t were determined15,20.

 

RESULTS AND DISCUSSION:

The bioanalytical method for quantifying RBZ in rat plasma was developed in accordance with the guidelines set by the ICH (M10). An initial investigation was conducted based on methods reported in the literature and pharmacopoeial standards. To enhance RBZ recovery while minimizing peak tailing and achieving optimal peak resolution, multiple trials were carried out using different extraction solvents.

 

HPLC Method Development:

The working standard solution of RBZ produced in the mobile phase was scanned throughout a wavelength range of 200-400 nm using a UV-visible spectroscopy method. The UV-visible spectra displayed a peak absorption at 225 nm when there was no interference from the mobile phase, baseline noise or endogenous compounds (Figure 2). Furthermore, RBZ demonstrated enhanced signal intensity and better spectrum response at this wavelength.

 

To optimize the composition of the mobile phase for efficient analyte separation, preliminary studies were carried out utilizing various ratios of solvents, including water, MeOH, acetonitrile (ACN) and buffer. In order to attain the best chromatographic performance, different concentrations of the mobile phase components were assessed. Mobile phase, 10 mM of anhydrous K2HPO4: MeOH (80:20, v/v), adjusted to a pH 3.5 with flow rate of 0.8 mL/min was optimized as it allowed RBZ to respond better without interference from endogenous compounds within a short Rt.

 

With a total run time of 10 minutes, RBZ demonstrated a retention time of 6 minutes. Based on RBZ recovery from various solvents, MeOH was selected as the extraction solvent as it showed highest recovery of RBZ (Table 1). This method exhibited a well-resolved peak with reduced retention time, high theoretical plate count, minimal peak tailing and optimal detection at 225 nm.

 

 

Figure 2: UV-visible spectra of RBZ

 

Validation of the HPLC Method with Emphasis on System Suitability and Specificity:

Table 2 displays the outcomes of the validated and optimized bioanalytical technique. System suitability criteria, such as the theoretical plate count (> 2000), asymmetry factor (0.7–1.5) and relative standard deviation (RSD) of peak area (< 2%), were all within acceptable range, demonstrating the method's dependability and applicability. HPLC chromatograms of MeOH, pure RBZ, blank plasma and RBZ-spiked plasma are superimposed in Figure 3. At the RBZ retention time in the blank sample, no interference was detected. These outcomes demonstrate the selectivity and suitability of the devised isocratic approach for RBZ measurement.


 

 

Table 1: Mean area, Rt, % RSD and % recovery of RBZ in various solvents for extraction

Extraction Solvent

Mean Area

Rt

% RSD

% Recovery

MeOH

129031.33 ± 730.12

6.08

0.69

95.81 ± 0.66

ACN

34527 ± 392.94

6.19

1.39

25.63 ± 0.35

Mobile Phase

76616 ± 471.53

6.18

0.75

56.89 ± 0.42

Data are represented as mean SD (n=3), Rt: Retention Time; % RSD: % Relative standard deviation; MeOH: Methanol; ACN: Acetonitrile

 

 

Table 2: RBZ in rat plasma: Optimization of bioanalytical chromatographic conditions

Validation Parameter

Method Responses

Acceptance Standards

Observed Values

System Suitability

RSD of peak area (n=6)

Asymmetry factor

Theoretical plates (N)

RSD < 2.0%

0.7-1.5

> 2000

0.71

1.09

9853.67

Linear Regression Data

Linearity (ng/mL)

Slope

Y-intercept when X=0

R2

-

5-1500

129.19

3166.5

0.9997

Accuracy

Initial concentration (ng/mL)

 

LQC

MQC

HQC

Observed mean concentration (ng/mL)

4.73 ± 2.15

98.77 ± 1.28

1490.81 ± 15.22

Recovery (%)

 

94.63 ± 2.85

99.06 ± 1.25

99.39 ± 8.86

Recovery

 

LQC

MQC

HQC

% RSD < 2.0%

% RSD

0.44

1.08

0.66

Robustness

Flow rate (mL/min)

Detection Wavelength

% RSD < 2.0%

0.55

0.74

Data expressed as mean SD (n=3), % RSD: % Relative standard deviation, LQC: Low-quality control, MQC: Medium-quality control, HQC: High-quality control

 

Table 3: Rat plasma RBZ intraday, interanalyst and interday precision results

Precision Parameters

QC Levels

Intraday Precision (Peak Area)

% RSD

Inter-analyst Precision

(Peak Area)

% RSD

Interday Precision

(Peak Area)

% RSD

LQC

3860 ± 42.6

1.20

3883 ± 46.77

1.22

3871 ± 43.89

1.13

MQC

15520.5 ± 133.72

0.94

15536 ± 145.46

0.93

15543.16 ± 150.65

0.96

HQC

195702 ± 1044.94

0.58

195628.5 ± 1167.5

0.59

195641.17 ± 1194.8

0.61

Data expressed as mean SD (n=3), % RSD: % Relative standard deviation

 


Linearity:

The method demonstrated a strong linear relationship between analyte concentration and detector response over the range of 5 to 1500 ng/mL. The calibration curve showed a correlation-coefficient (R˛) of 0.9997 and the regression model followed the equation y = 129.19x+ 3166.5, indicating precise quantification capabilities (Figure 4).

 

Figure 3: HPLC chromatographs of (A) MeOH, (B) Blank Plasma, (C) Pure RBZ and (D) Plasma with RBZ

 

Precision:

Three QC levels-LQC, MQC as well as HQC were used to evaluate the accuracy of the bioanalytical method that was established for RBZ in rat plasma. Three duplicates of intraday, inter-analyst and interday precision investigations were conducted21. The method's strong repeatability was demonstrated by the % RSD values being below 2% across all levels and situations. The approach is accurate, reliable and appropriate for routine analysis of RBZ in biological matrices, as demonstrated by Table 3, which shows that these results are within the acceptable limits specified in bioanalytical method validation recommendations.

 

Figure 4: Calibration curve of RBZ in rat plasma

 

Accuracy:

Accuracy is an essential parameter that reflects how closely the measured values align with the theoretical value. In this study, the mean recovery of RBZ at three spike concentrations (50%, 100% and 150%) ranged from 94.63 ± 2.85% to 99.39 ± 8.86%. The method's accuracy is confirmed by these results, which fall within the acceptable range of 90–110% since there is little difference between the actual and expected numbers22.

 

LOD, LOQ and Recovery:

As evidence of the high sensitivity of the devised analytical approach, the LOD and LOQ for RBZ were determined to be 4.66 ng/mL and 14.12 ng/mL, respectively23. Peak areas of the prepared QC samples at low, medium and high concentrations were compared to those of the corresponding standard solutions in order to determine absolute recovery. At the LOQ, MOQ and HOQ levels, the mean extraction recoveries of RBZ from plasma were 98.36 ± 0.44%, 98.77 ± 1.08% and 98.68 ± 0.66%, respectively. These values were all well within the acceptable range of 85% to 115%. These outcomes validate how well the approach recovers RBZ. Table 2 displays the recovery percentage RSD data.

 

Robustness:

A method is considered robust if it remains effective despite minor changes in experimental conditions24. To assess this, deliberate alterations were made to the experimental setup and the resulting chromatographic separation was examined. Table 4 shows that the devised approach remained stable even when the mobile phase wavelength (± 2nm) and flow rate (± 0.2mL) varied slightly. These modifications had no significant impact on the % RSD or system suitability parameters, confirming the reliability and robustness of the analytical method.

 

Table 4: Robustness of the developed method

Variable Parameters

Peak Area

Asymmetry Factor

Theoretical Plates (N)

Rt

(min)

Changes in flow rate

0.6 mL/min

5014.67 ± 30.66

1.20 ±

0.008

8531.3 ± 20.20

8.1

1.0 mL/min

5119.33 ± 25.66

1.18 ±

0.006

8233 ±

53.84

4.9

Changes in wavelength

223 nm

5067 ± 35.0

1.19 ±

0.006

8345.7 ± 31.26

6.1

227 nm

5052 ± 40.5

1.18 ±

0.002

8413 ±

33.40

6.1

Data expressed as mean SD (n=3), Rt: Retention Time

 

Stability Study:

The stability of the spiked plasma samples is evaluated following three thaw cycles. Three independent evaluations of low, medium and high-quality control samples were carried out for each freeze-thaw cycle26. As depicted from Table 5, recovery of RBZ after three freeze-thaw cycles is almost identical to the original, ranging from 98.30 ± 1.75% to 101.47 ± 0.75 %.

 

Table 5: Freeze-thaw stability of RBZ in rat plasma

QC Samples

% Recovery

at -20şC

% Recovery at

Room Temperature

LQC (5 ng/mL)

98.98 ± 0.64

99.27 ± 1.08

MQC (100 ng/mL)

99.03 ± 0.74

101.47 ± 0.75

HQC (1500 ng/mL)

98.67 ± 0.62

98.30 ± 1.75

Data expressed as mean SD (n=3), LQC: Low-quality control, MQC: Medium-quality control, HQC: High-quality control

 

Pharmacokinetic Study in Sprague-Dawley Rats:

After receiving a single oral dosage of RBZ (5 mg/kg body weight), male SD rats were subjected to pharmacokinetic investigation using the validated HPLC method. Plasma samples collected at defined time points were processed using the optimized extraction protocol and analyzed under the established chromatographic conditions25.

 

The concentration–time profile of rizatriptan revealed a rapid onset of systemic absorption, with peak plasma concentration (Cmax) recorded at 16.07 ± 3.86 ng/mL. The time to reach maximum concentration (Tmax) was observed at approximately 1 hour post-administration (Figure 5), suggesting moderate absorption kinetics for the formulation used. Elimination of the compound followed a biphasic pattern, with a terminal half-life (t˝) calculated at 3.53 ± 0.86hours.

 

Table 6: Pharmacokinetic parameters of RBZ after single oral dose (5 mg/kg body weight) administration in SD rats

Parameter

Unit

Value

Cmax

ng/mL

16.07 ± 3.86

Tmax

hr

1

Half life

hr

3.53 ± 0.86

AUC0-t

ng·h/mL

52.80 ± 18.85

MRT0-inf_obs

hr

5.54 ± 1.25

The mean ± SD is represented by each value. AUC: area under the plasma concentration-time curve from 0 hours to time or t; Cmax: maximum concentration in plasma; Tmax: time to reach maximum concentration; MRT0-inf_obs: average retention duration

 

Total systemic exposure, represented by the area under the plasma concentration–time curve from 0 to 6 hours (AUC0-t), was found to be 52.80 ± 18.85 ng·h/mL. Additionally, the mean residence time (MRT) was estimated at 5.54 ± 1.25 hours (Table 6). These pharmacokinetic parameters collectively reflect the suitability of the dosage form for delivering therapeutic levels of rizatriptan in-vivo and confirm the analytical method’s reliability in tracking its temporal disposition in biological matrices.

 

 

Figure 5: Mean plasma concentration-time profiles of RBZ in rats following a single oral dose of 5 mg/kg body weight

 

DISCUSSION:

The developed HPLC method demonstrated excellent performance in terms of reproducibility, linearity and sensitivity, enabling the reliable detection of Rizatriptan benzoate at clinically relevant plasma concentrations. Its successful application to a rat pharmacokinetic model underscores the method’s robustness and translational utility.


Table 7: Comparison of developed RBZ HPLC bioanalytical method with peer-reviewed literature

Chromatographic Condition

Limitations

Applicability

References

Column: C18 ACE, (150 × 4.6 mm) 3 mm, Mobile phase: 10 mM ammonium acetate 0.1% acetic acid: ACN (10:90), Flow rate: 1.0 mL/min, Run time: 10 min, Rt: 2.1 min

Mobile phase contains ACN, Robustness, LOD and LOQ was not performed.

Human Plasma

9

Column: C18 (Hibar, 250 x 4.6 mm i.d., 5μ), Mobile phase: MeOH:10 mM di-potassium hydrogen orthophosphate pH 3.2 (23:77), Flow rate: 1.1 mL/min, Run time: 10 min, Rt: 7.7 min

Higher retention time and flow rate at 231 nm. Robustness data was not mentioned. Pharmacokinetic study was not performed.

Rabbit Plasma

27

Column: C18 (SB-Phenyl, 250 x 4.6 mm i.d., 5μ), Mobile phase: Water: ACN: Trifluoroacetic acid (84:16:0.1), Flow rate: 2.0 mL/min, Run time: 10 min

Higher flow rate at 280 nm. Retention time was not mentioned.

Beagle Dogs Plasma

28

Column: C18 (Symmetry, 250 x 4.6 mm i.d., 5μ), Mobile phase: 0.05 M potassium dihydrogen orthophosphate (pH adjusted to 6.3 with the addition of 0.15% v/v triethylamine): ACN (50:50), Flow rate: 1.0 mL/min, Run time: 06 min

Mobile phase contains ACN, retention time was not mentioned.

SLNs in Male Wistar Rats

29

Mobile phase: ACN: Phosphate buffer (pH 3.5) (80:20)

Mobile phase contains ACN, retention time was not mentioned.

NLCs in Male Wistar Rats

30

 


Comparing the current findings to prior literature, the observed Tmax and Cmax are consistent with previously published data on orally administered RBZ in animal models (Table 7), although the exact values may vary depending on formulation type and physiological condition. The relatively shorter half-life observed in this study supports the drug’s classification as a fast-acting antimigraine agent, aligning with its therapeutic use during acute episodes.

 

What distinguishes the present method from previously reported techniques is the balance it strikes between analytical performance and operational efficiency. With a 6-minute run time, minimal buffer requirements, and a single-step protein precipitation protocol, the method provides a practical advantage in resource-limited settings. This positions it as a suitable platform not only for pharmacokinetic screening but also for routine bioavailability and drug interaction studies involving rizatriptan and its analogues.

 

CONCLUSION:

This investigation successfully established a validated, HPLC method tailored for the quantification of RBZ in rat plasma. Through careful optimization of mobile phase composition, detection parameters and extraction procedures, the method achieved desirable analytical outcomes; demonstrating strong linearity, high sensitivity, excellent recovery and reproducibility under varied operational conditions.

 

Besides its technical merits, this method was validated according to current regulatory guidelines and applied to an in-vivo pharmacokinetic study, where it proved effective in characterizing the drug's absorption, distribution, and elimination. The findings from the rat model revealed a pharmacokinetic profile compliant with rizatriptan’s known rapid onset and short duration of systemic action.

Importantly, the simplicity and cost-efficiency of this method-achieved without compromising analytical performance-make it highly adaptable for routine application in preclinical settings. It offers a reliable tool not only for pharmacokinetic assessments but also for future work involving drug interaction studies, bioequivalence investigations and formulation screening. Indeed, this developed, validated method holds promise as a standardized approach for the plasma-level evaluation of rizatriptan and potentially other structurally related triptans.

 

CONFLICT OF INTEREST:

The authors have no conflicts of interest regarding this investigation.

 

ACKNOWLEDGMENTS:

This work was funded by Bharati Vidyapeeth (Deemed to be University) Seed Money; Minor Research Project BVDU/A10/2024-2025/5839-2.

 

REFERENCES:

1.      Haarika B, Reddy VP. A RP-HPLC method development for the estimation of rizatriptan benzoate from formulated fast disintegrating sublingual tablets. International Journal of Pharmaceutical Sciences and Research. 2014; 5(9): 3784-3789. DOI: 10.13040/IJPSR.0975-8232.5(9).3784-89

2.      Kannappan N, Madhukar A, Ganesh A, Naveen Kumar CH, Mannavalan R. Analytical method development and validation of rizatriptan benzoate tablets by RP-LC. International Journal of Pharm Tech Research. 2009; 1(4): 1704-1708. http://www.sphinxsai.com/PTVOL4/pdf_vol4/PT=120%20(1704-1708).pdf

3.      Bou J, Gras J, Cortijo J, Morcillo EJ, Llenas J, Palacios JM. Vascular effects of the new anti-migraine agent almotriptan on human cranial and peripheral arteries. Cephalalgia. 2001; 21(8): 804-812. DOI: 10.1046/j.1468-2982.2001. 218233.x

4.      Bou J, Domčnech T, Puig J, Heredia A, Gras J, Fernández-Forner D, Beleta J, Palacios JM. Pharmacological characterization of almotriptan: An indolic 5-HT receptor agonist for the treatment of migraine. European Journal of Pharmacology. 2000; 410(1): 33-41. https://doi.org/10.1016/S0014-2999(00)00876-1

5.      Ahrens SP, Farmer MV, Williams DL, Willoughby E, Jiang K, Block GA, Visser WH. Rizatriptan wafer protocol 049 study group. Efficacy and safety of rizatriptan wafer for the acute treatment of migraine. 1999; Cephalalgia 19(5): 525-530. https://doi.org/10.1046/j.1468-2982.1999.019005525.x

6.      United States Food and Drug Administration (USFDA). 2019. https://www.accessdata.fda.gov/drugsatfda_docs/label/2019/020864s023,020865s024lbl.pdf. Accessed October 2019.

7.      DrugBank. 2005. https://go.drugbank.com/drugs/DB00953. Accessed 13 Jun 2005.

8.      Sultana MA, Roy PK, Begum N. Method development and validation of rizatriptan in dosage form by RP-HPLC. International Journal of Pharmacy and Analytica Research. 2018; 7(4): 602-609. https://doi.org/10.61096/ijpar.v7.iss4.2018.602-609

9.      Velusamy S, Masimukku VM, Chereddy S, Jadapalli JK, Palur K, Archakam SC, Kumarachari RK. Bioanalytical method development and validation of rizatriptan in human plasma using LC-MS/MS method. International Journal of Chemical and Analytical Science. 2023; 4(2): 108-114. https://doi.org/10.1016/j.ijcas.2013.03.009

10.   Prasanta S, Padma MS. Estimation of rizatriptan benzoate tablet by using UV spectophotometric methods. International Journal of Pharmaceutical Sciences Review and Research. 2013; 19(2): 97-100. https://www.globalresearchonline.net/journalcontents/v19-2/19.pdf

11.   Sagar PV, Kumar D, Dey S, Samal HB. Simultaneous estimation of rizatriptan, sumatriptan and zolmitriptan by RP-HPLC method in bulk. Journal of Pharmacy Research. 2010; 3(12): 2930-2933. https://www.researchgate.net/publication/284421062_Simultaneous_estimation_of_rizatriptan_sumatriptan_and_zolmitriptan_by_RP-HPLC_method_in_bulk

12.   Bahram M, Mohammadzadeh N. Multivariate curve resolution-alternative least squares for simultaneous kinetic–spectrophotometric determination of furosemide and rizatriptan in real samples based on their degradation study. Analytical and Bioanalytical Chemistry Research. 2019; 6(2): 441-448. https://doi.org/10.22036/abcr.2019.155529.1273

13.   Vishwanathan K, Bartlett MG, Stewart JT. Determination of antimigraine compounds rizatriptan, zolmitriptan, naratriptan and sumatriptan in human serum by liquid chromatography/electrospray tandem mass spectrometry. Rapid Communications in Mass Spectrometry. 2000; 14(3): 168-72. https://doi.org/10.1002/(sici)1097-0231(20000215)14:3%3C168::aid-rcm861%3E3.0.co;2-9

14.   Tamilselvi N, Kanagapriya K. Bioequivalence study and bioanalytical method development of remogliflozin etabonate tablets in wistar rat plasma using RP-HPLC method. Research Journal of Pharmacy and Technology. 2024; 17(2): 789-4. DOI: 10.52711/0974-360X.2024.00122

15.   Bharati S, Gaikwad V, Chellampillai B. Investigation of a quantitative bioanalytical HPLC method for the assessment of famotidine in pre-clinical samples: Application for pharmacokinetic studies. Analytical Chemistry Letters. 2023; 13(2): 202-212. https://doi.org/10.1080/22297928.2023.2217176

16.   Boraste S, Derle D. Bioanalytical method development and its validation by using RP-HPLC for determination of trimetazidine dihydrochloride in human plasma. Research Journal of Pharmacy and Technology. 2025; 18(10): 4790-4. DOI: 10.52711/0974-360X.2025.00690

17.   Tamilselvi N, Kanagapriya K. Bioequivalence study and bioanalytical method development of remogliflozin etabonate tablets in wistar rat plasma using RP-HPLC method. Research Journal of Pharmacy and Technology. 2024; 17(2): 789-4. DOI: 10.52711/0974-360X.2024.00122

18.   Telugu G, Suresh PV. Bioanalytical method development and validation of eprosartan mesylate and hydrochlorthiazide using RP-HPLC in human plasma. Research Journal of Pharmacy and Technology. 2023; 16(3): 1095-9. DOI: 10.52711/0974-360X.2023.00182

19.   Marakatham S, Shanmugapandiyan P. Bioanalytical method development and validation of doravirine, lamavudine and tenofovir disoproxil fumarate using HPLC in human plasma. Research Journal of Pharmacy and Technology. 2021; 14(8): 4087-1. DOI: 10.52711/0974-360X.2021.00708

20.   Vyas KP, Halpin RA, Geer LA, Ellis JD, Liu L, Cheng H, Chavez-Eng C, Matuszewski BK, Varga SL, Guiblin AR, Rogers JD. Disposition and pharmacokinetics of the antimigraine drug, rizatriptan, in humans. Drug Metabolism and Disposition. 2000; 28(1): 89-95. https://doi.org/10.1016/S0090-9556(24)15007-6

21.   Chitlange SS, Imran M, Bagri K, Sakarkar DM. A stability-indicating reverse phase high performance liquid chromatography method for the simultaneous determination of ramipril and valsartan in pharmaceutical dosage form. Research Journal of Pharmacy and Technology. 2008; 1(3): 215-217. https://rjptonline.org/HTMLPaper.aspx?Journal=Research%20Journal%20of%20Pharmacy%20and%20Technology;PID=2008-1-3-57

22.   Shinde SR, Bhoir SI, Pawar NS, Bhagwat AM. Quantitation of valsartan in human plasma by high performance liquid chromatography with fluorescence detection and its application to bioequivalence study. Research Journal of Pharmacy and Technology. 2009; 2(3): 487-490. https://rjptonline.org/HTMLPaper.aspx?Journal=Research%20Journal%20of%20Pharmacy%20and%20Technology;PID=2009-2-3-56

23.   Remi SL, Varkey J, Maheshwari RK, Nair AJ. A novel ecofriendly, cost effective mobile phase for HPLC- Simultaneous estimation and validation of paracetamol and diclofenac sodium in bulk and pharmaceutical formulation by RP-HPLC using hydrotropic solution as mobile phase. Asian J. Pharm. Res. 2020; 10(3): 163-170. DOI: 10.5958/2231-5691.2020.00029.5

24.   Jadhav V, Pawara PI, Pawara H. Development and validation of a new RP-HPLC method for nirmatrelvir in tablet formulations and bulk pharmaceutical applications. Asian Journal of Research in Pharmaceutical Sciences. 2025; 15(3): 232-6. DOI: 10.52711/2231-5659.2025.00035

25.   Abdel-Hamid ME, Sharaf LH, Phillips OA, Kehinde EO, Babu A. A rapid liquid chromatography-tandem mass spectrometry method for the determination of ciprofloxacin in rabbits serum following intra-gastric dosing: Application to pharmacokinetics study. Research Journal of Pharmacy and Technology. 2012; 5(5): 624-631. https://rjptonline.org/HTMLPaper.aspx?Journal=Research%20Journal%20of%20Pharmacy%20and%20Technology;PID=2012-5-5-20

26.   Rahade P, Sonawane S, Bhalerao A, Kshirsagar S. Development of a validated RP-HPLC method for estimation of ethionamide in spiked human plasma with UV detection. Asian Journal of Research in Pharmaceutical Sciences. 2016; 6(4): 230-234. https://ajpsonline.com/HTMLPaper.aspx?Journal=Asian+Journal+of+Research+in+Pharmaceutical+Sciences%3bPID%3d2016-6-4-4

27.   Awari VP, Meyyanathan SN, Karthik Y, Jawahar N. HPLC method development and validation of rizatriptan in rabbit plasma. Journal of Pharmaceutical Sciences and Research. 2014; 6(1): 24.

28.   Chokshi A, Vaishya R, Inavolu R, Potta T. Intranasal spray formulation containing rizatriptan benzoate for the treatment of migraine. International Journal of Pharmaceutics. 2019; 571: 118702. DOI: 10.1016/j.ijpharm.2019.118702

29.   Girotra P, Singh SK. Multivariate optimization of rizatriptan benzoate-loaded solid lipid nanoparticles for brain targeting and migraine management. AAPS PharmSciTech. 2016, 18(2): 517-528. DOI: 10.1208/s12249-016-0532-0

30.   Mathure D, Ranpise H, Awasthi R, Pawar A. Formulation and characterization of nanostructured lipid carriers of rizatriptan benzoate-loaded in situ nasal gel for brain targeting. Assay and Drug Development Technologies. 2022, 20(5): 211-224. DOI: 10.1089/adt.2022.044.

 

 

 

 

Received on 05.08.2025      Revised on 03.12.2025

Accepted on 08.02.2026      Published on 08.08.2026

Available online from August 12, 2026

Research J. Pharmacy and Technology. 2026;19(8):3497-3504.

DOI: 10.52711/0974-360X.2026.00496

© RJPT All right reserved

 

This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. Creative Commons License.