Compatibility of Complete Blood Count Values between HumaCount 5D CRP and Sysmex XN - 1000
Bagus Dwi Irianto Widodo1, Yetti Hernaningsih2,3*, Hartono Kahar2,3
1Clinical Pathology Residency Program, Department of Clinical Pathology,
Faculty of Medicine, Universitas Airlangga - Dr. Soetomo General Academic Hospital, Surabaya, Indonesia.
2Department of Clinical Pathology, Faculty of Medicine, Universitas Airlangga, Surabaya, Indonesia.
3Dr. Soetomo General Academic Hospital, Surabaya, Indonesia.
*Corresponding Author E-mail: yetti-h@fk.unair.ac.id
ABSTRACT:
HumaCount 5D CRP is an instrument for hematology examination that comprises hemoglobin, leukocytes, erythrocytes, hematocrit, and platelet examinations. Sysmex XN-1000 is one of the current established instruments for hematology examination. Comparison of the results between new and established instruments can be done by comparing each parameter in all result categories (low, normal, high). This study aims to analyze and compare the results of hematology examinations between the HumaCount 5D CRP and one of the current established instruments (the Sysmex XN-1000) on 5 parameters of hemoglobin, leukocytes, erythrocytes, hematocrit, and platelets. Comparative tests were conducted on 172 samples using HumaCount 5D CRP and Sysmex XN-1000. The result of each examined parameter was categorized into low, normal, and high. SPSS was used to analyze the data on the entire result, and each group of parameters using the normality test, and further was adjusted using parametric and non-parametric tests. The examination of the whole samples using the One-Way ANOVA test showed that the results were hemoglobin (p=0.818), leukocytes (p=0.325), erythrocytes (p=0.626), hematocrit (p=0.000), and platelets (p=0.394). Testing each parameter group, parametric and non-parametric tests were used to adjust the results of the normality test and the results were leukocytes (low p=0.109, normal p=0.010, high p=0.020), erythrocytes (low p=0.015, normal p=0.825), hemoglobin (low p=0.167, normal p=0.898), hematocrit (low p=0.000, normal p=0.000), and platelets (low p=0.472, normal p=0.088, high p=0.695). The overall test results showed that the HumaCount 5D CRP instrument has quite good results. The HumaCount 5D CRP is a fairly good hematology analyzer instrument and can be used for clinical services. However, it is necessary to be aware of the hematocrit result if there is a discordance with the patient's clinical condition.
KEYWORDS: Hematology, Hematology Analyzer, HumaCount 5D CRP, Sysmex XN-1000, Compatibility.
INTRODUCTION:
A complete blood count (CBC) is a routine test performed by clinical laboratories as a supporting component frequently used in the diagnostic process of a disease1-3. The CBC is an important tool in routine hematological studies of patients. The detection of anomalies in leukocytes, erythrocytes, and peripheral blood platelets aids in the clinical interpretation of patients' signs and symptoms. Modern automated analyzers enable for the objective and precise measurement of several hematological parameters, as opposed to manual approaches. Every day, innovative technologies are created to produce more dependable and accurate results4. Tests with haematology analyzers are becoming more efficient. In clinical laboratories, hematology testing frequently use an instrument called a hematology analyzer. A hematology analyzer provides blood test results quickly and accurately. The haematology analyser (Sysmex XN-1000) was one of the most common used. With the rapid development of technology, many hematology analyzers have been produced. Therefore, quality control of the latest instruments needs to be assessed to maintain the accuracy of the test results5-7.
Dr. Soetomo General Academic Hospital is one of the tertiary hospitals in Indonesia, and the complete blood count examination in this hospital was performed using Sysmex XN-1000 instrument in the clinical pathology laboratory. The examination principle of this instrument used the flow cytometry method for leukocyte testing, impedance for RBC and platelet testing, and the Sodium Lauryl Sulfate (SLS) method for hemoglobin testing. The Sysmex XN-1000 in Indonesia is one of the hematology instruments that serves as the standard for complete blood count testing8. The laboratory test has three stages: pre-analytical, analytical, and post-analytical. The pre-analytical step is referred to as the initial preparation stage, as it impacts the quality of the sample generated and influences the subsequent work process. The pre-analytical stage includes the specimen collection process, where the use of the right tube for blood samples is one of the most influential pre-analytical factors, because each tube contains anticoagulants, such as Ethylenediaminetetraacetic Acid (EDTA), which is the recommended standard for routine blood tests9-11.
HumaCount 5D CRP is an innovative complete hematology testing instrument that includes the examination of hemoglobin, leukocytes, erythrocytes, hematocrit, and platelets, along with the rapid and accurate measurement of C-reactive protein (CRP). CRP is an important biomarker for assessing inflammation in patients. The principle of hematology testing on this instrument uses the same methods as the Sysmex XN-1000 called flow cytometry for leukocyte testing, impedance for RBC and platelet testing, and colorimetry for hemoglobin testing12. Many studies have been conducted to compare various hematological analyzer instruments, but the HumaCount 5D CRP has not been studied yet. This study aimed to analyze and compare the results of complete blood count values between the HumaCount 5D CRP and Sysmex XN-1000.
METHODS:
This study used primary data from patients who underwent a complete blood count examination in the clinical pathology laboratory of Dr. Soetomo General Academic Hospital from July 20, 2023, to August 31, 2023. The population in this study consisted of 172 patient samples of all ages, both men and women, who had a complete blood count test in the clinical pathology laboratory of Dr. Soetomo General Academic Hospital. This was a quantitative descriptive study conducted to determine the values of each variable. The patients’ blood samples were collected in EDTA tubes, after checked the patient's age according to the inclusion criteria, the volume of blood sample from each subject was 3ml, and checking the condition of the sample (whether there are clots or fibrin threads). A complete blood test was performed using both different instruments (HumaCount 5D CRP and Sysmex XN-1000). SPSS 24.0 was used to analyze the data on the entire result, and each group of parameters using the normality test, and further was adjusted using parametric and non-parametric tests using One-Way ANOVA. The data was presented with bland altman graphs.
RESULTS:
Figure 1 below showed the data distribution of each sample which showed that for the leukocyte parameter, 26 samples showed low results (<4,000/µL), 60 samples showed normal results (4,000-10,000/µL), and 86 samples showed high results (>10,000/µL). While for the erythrocyte parameter, 81 samples showed low results (<3,800,000/µL), 91 samples showed normal results (3,800,000-6,000,000/µL), and no samples showed high results (>6,000,000/µL). For the hemoglobin parameter, 72 samples showed low results (<11.6 g/dL), 100 samples showed normal results (11.6-16.6 g/dL), and no samples showed high results (>16.6 g/dL). For the hematocrit parameter, 114 samples showed low results (<36%), 58 samples showed normal results (36-54%), and no samples showed high results (>54%). For the platelets parameter, 32 samples showed low results (<150,000/µL), 116 samples showed normal results (150,000-400,000/µL), and 24 samples showed high results (>400,000/µL).
Figure 1. Data distribution chart.
The data presented in table 1 showed that the data obtained for the erythrocyte, hemoglobin, hematocrit, and platelet parameters were consistent (accordance) in both testing instruments, while for the leukocyte parameter, the data from both testing instruments showed inconsistency (discordance).
Table 1. Overall data distribution.
|
Parameter |
Examination Instruments |
N |
Mean |
SD |
CV% |
|
Leukocyte |
SYSMEX XN-1000 |
172 |
13.27 |
11.92 |
89.83 |
|
HUMACOUNT 5D CRP |
172 |
12.04 |
11.20 |
93.02 |
|
|
Erythrocyte |
SYSMEX XN-1000 |
172 |
3.88 |
0.72 |
18.56 |
|
HUMACOUNT 5D CRP |
172 |
3,92 |
0.71 |
18.11 |
|
|
Hemoglobin |
SYSMEX XN-1000 |
172 |
11.03 |
2.09 |
18.95 |
|
HUMACOUNT 5D CRP |
172 |
11.08 |
2.11 |
19.04 |
|
|
Hematocrit |
SYSMEX XN-1000 |
172 |
32.89 |
6.36 |
19.34 |
|
HUMACOUNT 5D CRP |
172 |
35.68 |
6.63 |
18.58 |
|
|
Platelet |
SYSMEX XN-1000 |
172 |
261.20 |
139.65 |
53.46 |
|
HUMACOUNT 5D CRP |
172 |
274.13 |
141.55 |
51.64 |
Table 2. Data distribution of each variable.
|
Parameter |
Group |
Examination Instruments |
N |
Mean |
Standard Deviation |
CV% |
|
Leukocyte |
Low |
SYSMEX XN-1000 |
26 |
2.68 |
0.89 |
33.21 |
|
Normal |
60 |
7.92 |
1.76 |
22.22 |
||
|
High |
86 |
20.20 |
13.44 |
66.53 |
||
|
Low |
HUMACOUNT 5D CRP |
26 |
2.41 |
0.79 |
32.78 |
|
|
Normal |
60 |
7.11 |
1.66 |
23.35 |
||
|
High |
86 |
18.38 |
12.81 |
69.70 |
||
|
Erythrocyte |
Low |
SYSMEX XN-1000 |
81 |
3.24 |
0.40 |
12.35 |
|
Normal |
91 |
4.44 |
0.41 |
9.23 |
||
|
Low |
HUMACOUNT 5D CRP |
81 |
3.36 |
0.41 |
12.20 |
|
|
Normal |
91 |
4.41 |
0.53 |
12.02 |
||
|
Hemoglobin |
Low |
SYSMEX XN-1000 |
72 |
9.06 |
1.19 |
13.13 |
|
Normal |
100 |
12.45 |
1.31 |
10.52 |
||
|
Low |
HUMACOUNT 5D CRP |
72 |
9.30 |
1.26 |
13.55 |
|
|
Normal |
100 |
12.37 |
1.62 |
13.10 |
||
|
Hematocrit |
Low |
SYSMEX XN-1000 |
114 |
29.33 |
4.36 |
14.87 |
|
Normal |
58 |
39.89 |
2.93 |
7.35 |
||
|
Low |
HUMACOUNT 5D CRP |
114 |
32.32 |
4.97 |
15.38 |
|
|
Normal |
58 |
42.29 |
4.02 |
9.51 |
||
|
Platelet |
Low |
SYSMEX XN-1000 |
32 |
76.12 |
43.68 |
57.38 |
|
Normal |
116 |
263.43 |
66.01 |
25.06 |
||
|
High |
24 |
497.20 |
121.79 |
24.50 |
||
|
Low |
HUMACOUNT 5D CRP |
32 |
85.00 |
50.48 |
59.39 |
|
|
Normal |
116 |
278.93 |
68.07 |
24.40 |
||
|
High |
24 |
503.12 |
133.33 |
26.50 |
The data presented in table 3 showed that for all variables, p>0.05, means that the data is normally distributed. Subsequently, a parametric test, called One Way ANOVA was conducted.
Table 3. The results of the normality test for the entire data.
|
Examination Instruments |
p-value |
|
|
Leukocyte |
SYSMEX XN-1000 |
0.087 |
|
HUMACOUNT 5D CRP |
0.085 |
|
|
Erythrocyte |
SYSMEX XN-1000 |
0.200 |
|
HUMACOUNT 5D CRP |
0.200 |
|
|
Hemoglobin |
SYSMEX XN-1000 |
0.200 |
|
HUMACOUNT 5D CRP |
0.200 |
|
|
Hematocrit |
SYSMEX XN-1000 |
0.200 |
|
HUMACOUNT 5D CRP |
0.200 |
|
|
Platelet |
SYSMEX XN-1000 |
0.071 |
|
HUMACOUNT 5D CRP |
0.061 |
The data presented in table 2 showed that the data obtained for the overall dataset were consistent for the low and normal leukocyte parameters, erythrocytes, hemoglobin, and hematocrit in both testing instruments. In contrast, the high leukocyte parameter showed inconsistent data. After sampling, a normality test was performed for the entire dataset and for each parameter using the Kolmogorov-Smirnov and Shapiro-Wilk tests.
Table 4 below showed that the results of the normality test mostly indicated that the data was not normally distributed. Therefore, a non-parametric test, called the Mann-Whitney test was conducted. For the normal leukocyte parameter, the data showed a normal distribution; therefore, a parametric test called the Independent T-Test was then conducted.
Table 4. Normality test results for each parameter.
|
Group |
Examination Instruments |
p-value |
|
|
Leukocyte |
Low |
SYSMEX XN-1000 |
0.014 |
|
Normal |
0.080 |
||
|
High |
0.000 |
||
|
Low |
HUMACOUNT 5D CRP |
0.005 |
|
|
Normal |
0.443 |
||
|
High |
0.000 |
||
|
Erythrocyte |
Low |
SYSMEX XN-1000 |
0.000 |
|
Normal |
0.003 |
||
|
Low |
HUMACOUNT 5D CRP |
0.000 |
|
|
Normal |
0.003 |
||
|
Hemoglobin |
Low |
SYSMEX XN-1000 |
0.000 |
|
Normal |
0.001 |
||
|
Low |
HUMACOUNT 5D CRP |
0.001 |
|
|
Normal |
0.000 |
||
|
Hematocrit |
Low |
SYSMEX XN-1000 |
0.012 |
|
Normal |
0.002 |
||
|
Low |
HUMACOUNT 5D CRP |
0.049 |
|
|
Normal |
0.006 |
||
|
Platelet |
Low |
SYSMEX XN-1000 |
0.041 |
|
Normal |
0.042 |
||
|
High |
0.000 |
||
|
Low |
HUMACOUNT 5D CRP |
0.162 |
|
|
Normal |
0.036 |
||
|
High |
0.012 |
The results of the parametric and non-parametric tests of each parameter were presented in table 5. The results of the parametric statistical test of all the parameters showed that most of the parameters showed no significant difference between the HumaCount 5D CRP hematology test instrument and the Sysmex XN-1000. These results were confirmed in the blant-altman test which showed the results within the standard deviation range. However, the hematocrit parameter results showed that there was a difference between the HumaCount 5D CRP hematology test instrument and the Sysmex XN-1000.
Table 5. Overall statistical test results.
|
p-value |
|
|
Leukocyte |
0.325 |
|
Erythrocyte |
0.626 |
|
Hemoglobin |
0.818 |
|
Hematocrit |
0.000 |
|
Platelet |
0.394 |
Figure 2. Graph of Bland Altman on leukocyte, erythrocyte, hemoglobin, and platelet parameters
Figure 2 above showed the graph of Bland Altman on leukocyte, erythrocyte, hemoglobin, and platelet parameters. The results of parametric and non-parametric statistical tests for each group (low, normal, high) presented in table 6 showed that the results of most of the parameter groups showed no significant difference between the HumaCount 5D CRP hematology and the Sysmex XN-1000 test instruments. These results were confirmed in the blant-altman test which showed the results included in the standard deviation range. However, the results of the leukocyte parameters (normal and high), erythrocytes (low), and hematocrit (low and normal) showed that there was a difference between the HumaCount 5D CRP hematology and the Sysmex XN-1000 test instruments.
Table 6. Statistical test results for each parameter.
|
Group |
p-value |
|
|
Leukocyte |
Low |
0.109 |
|
Normal |
0.010 |
|
|
High |
0.020 |
|
|
Erythrocyte |
Low |
0.015 |
|
Normal |
0.825 |
|
|
Hemoglobin |
Low |
0.167 |
|
Normal |
0.898 |
|
|
Hematocrit |
Low |
0.000 |
|
Normal |
0.000 |
|
|
Platelet |
Low |
0.472 |
|
Normal |
0.088 |
|
|
High |
0.695 |
Figure 3. Graph of Bland Altman on parameters of leukocytes (low), erythrocytes (normal), hemoglobin (low, normal), and platelets (low, normal, high).
Figure 3 above showed the graph of Bland Altman on the parameters of leukocytes (low), erythrocytes (normal), hemoglobin (low, normal), and platelets (low, normal, high).
DISCUSSION:
The clinical laboratory is an essential component of health services because it plays a vital role in in-vitro diagnostics. The primary reason is that laboratory tests are required for screening, diagnosis, illness monitoring, and monitoring treatment. Given the significance of these laboratory tests, every laboratory is obliged to give exact, timely, and reliable findings13-15. The complete blood count is very essential. A previous study found that a routine serial blood count determination during the full course of palliative radiotherapy had a substantial influence on RBC count16. This current study was conducted to compare the results of hematology examinations between the HumaCount 5D CRP and one of the current established instruments (the Sysmex XN-1000) on 5 parameters of hemoglobin, leukocytes, erythrocytes, hematocrit, and platelets.
The results of this study showed that the distribution of data on all parameters generally showed similar coefficient of variation (CV%) results, but the leukocyte parameters showed differences in CV% between the Sysmex XN-1000 leukocyte parameters (89.83%) and the HumaCount 5D CRP instrument (93.02%). This showed that the leukocyte parameters on the HumaCount 5D CRP instrument have more varied results compared to the Sysmex XN-1000 instrument.
The distribution results of each parameter level showed that most of the results have similar CV% results but there were differences in the high leukocyte parameters. The Sysmex XN-1000 instrument showed a high leukocyte CV% (66.53%), while the HumaCount 5D CRP instrument showed a high leukocyte CV% (69.70%). This result showed that the high leukocyte parameters according to the HumaCount 5D CRP instrument was more varied, compared to the Sysmex XN-1000 instrument.
Furthermore, the overall normality test on both instruments showed that all data parameters were normally distributed, therefore the parametric Independent T Test was carried out. However, the results of the normality test for each parameter level were different from the overall normality test of the data, showed that almost all levels of data parameters were not normally distributed, only the normal leukocyte parameter showed that the data was normally distributed. Therefore, all parameter levels used the Mann-Whitney non-parametric statistical test, except for the normal leukocyte parameter, the parametric Independent T Test was carried out.
The parametric test results on the overall data showed that the parameters of leukocytes, erythrocytes, hemoglobin, and platelets showed no significant difference between the two instruments (p>0.05). However, the results of hematocrit parameter showed a significant difference between the two instruments (p<0.05). Furthermore, the parametric and non-parametric tests from each group showed that the parameters of low leukocyte, normal erythrocytes, low and normal hemoglobin, low platelets, normal platelets, and high platelets have good results with p>0.05 which means that the data between the two instruments showed no significant difference. A previous study using two advanced hematology analyzers also reported that the results obtained from different automated CBC analyzers were similar4. Another study that compared the two hematological analyzers also reported that the two instruments have good correlation and consistency, and the two instruments can replace each other17. However, our current study found that the results of the parameters of normal and high leukocyte, low erythrocytes, low and normal hematocrit have significant differences between the two instruments with p values of <0.05.
LIMITATION:
The sample was obtained by random sampling (not depicting a homogeneous population) where there were patients with various diseases and conditions which could cause the varies in distribution of the sample population.
CONCLUSION:
Compliance testing by comparing the latest innovative hematology instrument HumaCount 5D CRP with the established instrument Sysmex XN-1000 showed quite good results in overall results but there were still significant differences in several parameter levels. It is necessary to be aware of hematocrit parameter results if there is a discordance with the patient's clinical condition. Further study is needed using the prevalent data distribution at each level.
CONFLICT OF INTEREST:
The authors have no conflicts of interest regarding this study.
REFERENCES:
1. Indonesian Ministry of Health. Indonesia Health Profile 2021. 1st ed.Vol. 1. Indonesian Ministry of Health, Jakarta. 2022.
2. Putra AAMS. Hernaningsih Y. Comparison of Complete Blood Count Parameters using EDTA, Sodium Citrate, and Heparin Anticoagulants. Research J. Pharm. and Tech. 2022; 15: 4687-1. doi: 10.52711/0974-360X.2022.00786.
3. Nima RS. Ablaa HA. Mohammed HJ. The Relationship Between Blood Counts and Serum Levels of zinc and Copper in Children with β-thalassemia Living in Najaf Governorate, in Iraq. Research J. Pharm. and Tech. 2018; 11: 4640-4644. doi: 10.5958/0974-360X.2018.00848.X.
4. Bhola RK. Fudaly C. Rastogi S. A Comparative Evaluation of Performance of Sysmex XN 3000 and Horiba Yumizen H2500 Automated Complete Blood Count Analysers. Indian J Hematol Blood Transfus. 2024; 40: 303–314. https://doi.org/10.1007/s12288-023-01687-6.
5. Permatasari CA. Hajat A. Sedana MP. Diagnostic Testing of Hematology Analyzer Equipped with WPC Channel Scattergram in Determining Strains of Hematological Malignancies. Research J. Pharm. and Tech. 2022; 15: 4005-1. doi: 10.52711/0974-360X.2022.00671.
6. Putra IGPAFS. Prihatiningsih D. Subhaktiyasa PG. Analysis of Internal Quality Assurance of Platelet Examination in the Clinical Laboratory of Abiansemal I Primary Health. Jurnal Analis Laboratorium Medik. 2020; 5(2): 28–34. https://doi.org/10.51544/jalm.v5i2.2152.
7. Bain BJ, Bates I, Laffan MA. Dacie and Lewis Practical Haematology. 12th ed. Vol. 1. Elsevier, London. 2017.
8. Sysmex Company. PPT: Sample Analysis. Sysmex Asia Pasific Pte. Ltd, 2011.
9. Yasmeen R. Reddy NK. Souza MGD. Swathi B. Waheed AA. Chinmayee K. Reddy JV. Study of Anticoagulants low molecular weight heparin and Unfractionated heparin in the management of Non-St elevation Myocardial Infarction. Research J. Pharm. and Tech. 2020; 13: 3151-3155. DOI: 10.5958/0974-360X.2020.00557.0
10. Hussein TA. Saihood AS. Rayshan AR. Molecular Identification and Characterization in blood samples of patients with progressive multifocal leukoencephalopathy and human kidney cell line culture of John Cunningham Virus (JCV) in Iraq. Research J. Pharm. and Tech. 2018; 11: 3392-3398. DOI: 10.5958/0974-360X.2018.00625.X
11. Victor ON. Tufon EN. Osmond EC. Effect of Vitamin E-Stored Experimental Blood Samples on Malondialdehyde and Protein Concentration. Research J. Pharm. and Tech. 2011; 4: 1785-1786.
12. Human Gesellschaft für Biochemica und Diagnostica mbH. HumaCount 5DCRP: User Manual. Rev. 04/2023-02. Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden, Germany. 2023.
13. Varanita SV. Kahar H. Fitriah M. Compatibility of Two EDTA Tubes for Hemoglobin (Hb) Hematocrit (Hct) Tests with Aptus Equipment. Research J. Pharm. and Tech. 2022; 15: 4447-0. doi: 10.52711/0974-360X.2022.00745.
14. Subbalakshmi NK. Sunandha S. Influence of Perceived Stress on Hemoglobin Concentration. Research J. Pharm. and Tech. 2017; 10: 61-64. DOI: 10.5958/0974-360X.2017.00015.4.
15. Ahmad RI. Al-Barry HA. Salloum SI. Fetal Hemoglobin (HbF) levels in Sickle cell anemia patients in Lattakia, Syria. Research J. Pharm. and Tech. 2020; 13: 3246-3248. DOI: 10.5958/0974-360X.2020.00575.2.
16. Brundha MP. Pathmashri VP. Sundari S. Quantitative Changes of Red Blood cells in Cancer Patients under Palliative Radiotherapy-A Retrospective Study. Research J. Pharm. and Tech. 2019; 12(2):687-692. DOI: 10.5958/0974-360X.2019.00122.7.
17. Wang J. Zhao S. Su Z. Liu X. Analytical comparison between two hematological analyzer systems: Mindray BC‐5180 vs Sysmex XN‐1000. J Clin Lab Anal. 2019; 33: e22955. https://doi.org/10.1002/jcla.22955.
|
Received on 04.12.2024 Revised on 13.04.2025 Accepted on 17.06.2025 Published on 01.12.2025 Available online from December 06, 2025 Research J. Pharmacy and Technology. 2025;18(12):6084-6090. DOI: 10.52711/0974-360X.2025.00880 © RJPT All right reserved
|
|
|
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License. Creative Commons License. |
|