Development and Validation of RP HPLC Method for Estimation of Ruxolitinib Tablet in Pharmaceutical Dosage Form
Honey Joshi1, Bhakti Shah2*
1Department of Science, Shree Swaminarayan College of Science,
Swaminarayan University, Kalol, Gandhinagar, Gujarat, India - 382725.
2Department of Pharmaceutics, Shree Swaminarayan College of Pharmacy,
Swaminarayan University, Kalol, Gandhinagar, Gujarat, India-382725.
*Corresponding Author E-mail: bhaktishah@gmail.com
ABSTRACT:
HPLC is a separation technique used to conduct chromatography. There are various technologies of chromatography in which RP-HPLC is one of them. Reverse phase HPLC is highly valuable in research technique because it uses water based mobile phase in organic solvent. Ruxolitnib is a JAK ˝ inhibitor used in the treatment of myelofibrosis. Ruxolitinib gives highly separation in RP-HLPC method. So using this method gives marvelous result with help of different ingredients.(RP-HPLC) method was developed for the quantification of Ruxolitinib zorbax SB (Stable bond) C18 phenyl (150×4.6mm 3.5μm) column was used as stationary phase to separate drug. Mobile phase comprised of ACN: MeOH (50:50) having pH 3.5was run isocratically at a flow rate of 1mL/min at 35şC. The method was validated according to ICH guidelines for linearity, precision, accuracy,. The method was found accurate, precise with average retention time 15min with regression coefficient R2 value of 0.9996 as per ICH norms, the developed method was successfully validated and used for the quantification of Ruxolitinib in fast dissolving tablets.
KEYWORDS: RPHPLC technique, Ruxolitinib, ICH guidelines, zorbax SB C18 phenyl column, chromatography.
INTRODUCTION:
To develop and validate a simple, cheap, accurate, the rapid Reverse Phase High Performance Liquid Chromatography (RP-HPLC) method is used as per ICH guidelines for estimation of Ruxolitinib in pharmaceutical dosage forms. Ruxolitinib .is use for secretion of inflammatory cytokines results in myeloproliferation, which is use for increased bone marrowfibrosis (BMF), and extramedullary hematopoiesis.1
The past decade has witnessed considerable progress in the understanding of the cellular and molecular biology of MPNs, and this has recently resulted in the addition of the Janus kinase (JAK) 1 and JAK2 inhibitor ruxolitinib to our therapeutic armamentarium2 Janus kinases (JAKs), consisting of JAK1, JAK2, JAK3,and Tyk2, are intracellular signaling enzymes that actdownstream of key proinflammatory cytokines, and each isknown to contribute to the pathogenesis of atopic dermatitis3-10
Analytical Method validation:
High performance liquid chromatography Definition:11-15
High Performance Liquid Chromatography which is also known as High Pressure Liquid Chromatography. It is a popular analytical technique used for the separation, identification and quantification of each constituent of mixture. HPLC is an advanced technique of column liquid chromatography. The solvent usually flows through column with the help of gravity but in HPLC technique the solvent will be forced under high pressures up to 400 atmospheres so that sample can be separated into different constituents with the help of difference in relative affinities
HPLC Theory and Principle:8,9,10
Chromatography is an analytical method that finds wide application for the separation, identification and determination of chemical components in complex mixtures. This technique is based on the separation of components in a mixture (the solute) due to difference in migration rates of the component through a stationary phase by gaseous or liquid mobile phase. HPLC was derived from classical column Chromatography and has found an important place in analytical technique. Most of drugs in multi component dosage forms can be analyses by HPLC method because of the several advantages like rapidity, specificity, accuracy, precision and ease of automation in this method. The principle of separation in normal phase mode and reverse phase mode is adsorption. When mixtures of components are introduced into HPLC column, they travel according to their relative affinities towards the stationary phase. The component which has more affinity towards the adsorbent travels slower.
MATERALS AND METHOD:
Moble phase preparations:
Diluted Orthophosphoric acid: Accurately transfer 2.0ml of orthophosphoric acid into10ml volumetric flask and make volume up to the mark with water and mixed well.
Buffer preparation:
Accurately weigh 6.8gm of potassium dihydrogen phosphate in 1000ml of water and mixed well. Filtered the buffer solution with a 0.45µ PVDF membrane filter.
Buffer pH 3.5:
1000ml of above buffer was taken and pH was adjusted to 3.50 with a diluted ortho phosphoric acid solution (1.0ml):
Mobile phase A – Buffer pH 3.5(100%) line A
Mobile Phase B – Acetonitrile: MeOH (100%) line B 50:50 (V/V)
Methodology:
Selection of mobile phase:
For optimization of mobile phase, several trials were done to select the best mobile phase for the separation of Ruxolitinib with an effective resolution from its degradant impurity and peak symmetry of Ruxolitinib peak. The trials for the selection of mobile phase are shown in table
Standard solution 1:
Weigh accurately about 66mg of Ruxolitinib phosphate standard and transfer into a 50ml volumetric flask. Add 35ml of Diluent and sonicate to dissolve it. Diluent to volume with diluent and mix well.
Standard solution:
Pipette 5.0ml of sample solution 1 and transfer into a 50ml volumetric flask. Dilute to volume with diluent and mix well.
Sample Solution 1:
Carefully weigh 20 Tablets and Determine average weight Take 20 Tablets and crush in mortal pestal and prepared fine powder fine powder. Mix the content homogeneously prior to weighing. Weigh accurately equivalent to about 66mg of Ruxolitinib and transfer into a 50ml volumetric flask (with swirling) containing about 35ml of Diluent, and sonicate for about 30minutes with intermittent shaking. Maintain the temperature of ultrasonic bath at room temperature. Cool to room temperature and diluent to volume with Diluent, mix well
Sample solution:
Pipette 5.0ml of sample solution-1 and transfer into a 50ml volumetric flask. Dilute to volume with diluent and mic well. Filter the solution through 0.45μm PVDF membrane syringe filter. Discard the first 5ml of filtrate.
Chromatographic Conditions:
Column: zorbax SB(Stable bond) phenyl (150×4.6) mm 3.5µm
Wavelength – : 310nm
Column temperature - : 35 şC
sample cooler temperature : 4ş C
injection volume - : 10µL
Rinsing line : Methanol (100%)
Elution type : Gradient
Flow rate – : 1.0ml/ min
Run time - : 15min
Gradient Program :
Table no 1: Mobile Phase gradient Ratio
Time |
Mobile phase A |
Mobile phase B |
0.0 |
100 |
0 |
2.0 |
30 |
70 |
5.0 |
30 |
70 |
8.0 |
30 |
70 |
10.0 |
100 |
0 |
15.0 |
100 |
0 |
EVALUTION PARAMETERS16;17
Method Validation:
Authentification of the investigative method is the process that starts by laboratory studies in which the requirements of the performance properties of method are met for the intended analytical application. To the validation of analytical procedures RP-HPLC method developed was validated according to International Conference on Harmonization (ICH) and USP guidelines. Various parameters or criteria are used for the method of validation, such as linearity, accuracy, precision, limit of Quantification (LOQ) and limit of Detection (LOD).
Linearity:
Linearity the method was tested from 80-120% of the targeted level of the assay concentration for analyte. Standard solutions contained 2-12μg/mL Ruxolitinib
Linearity solutions were injected in triplicate. The equations of the calibration curves for Montelukast sodium obtained were y = 02E+06X in the Ruxolitinib determination, the calibration graphs were found to be linear in the aforementioned concentrations with correlation coefficients 0.9998.
Accuracy:
The precision was determined with the help of recovery experiments, by the verification with percentage mean recovery of sample at three different levels (50-150%). 20 blank tablets were powdered and mixed. This powder was then spiked with a quantity of Ruxolitinib corresponding to 50%, 100% and 150% of the labeled claim.
Precision:
System precision:
The value for %RSD (Relative Standard Deviation) was obtained less than 2 by six replicate injections of the standard solution at working concentrations, concerning the peak area for the drug. It designates the adequate reproducibility and hence the accuracy of the system.
Method precision:
The accuracy of procedure was determined by conducting assay of sample with the tests of (I) Repeatability (Intraday precision) and (II) Intermediate precision or ruggedness (Interday precision) completed within 3 successive daysbythree different analysts, at working concentration.
Repeatability (Intraday precision):
The value for %RSD (Relative Standard Deviation) less than 2 for six successive injections of the sample solution from the same homogenous mixture at working concentrations, concerning % assay for understood that the method gives consistently reproducible results the drug which indicate that the method developed is method precise by the test of repeatability.
LIMIT OF DETECTION (LOD):
The detection limit of an specify analytical procedure is the lowest amount of analyte in a sample which can be detected but not necessarily quantitated as an exact value. The limit of detection may be expressed as:
DL = 3.3×σ/S
Where,
DL = The standard deviation of the response
S = The slope of the calibration curve
The estimation of LOD may be carried out in multiple ways. LOD can calculate from the linearity correlation equation. The slope S may be estimated by calibrated curve.
LIMIT OF QULIFICATION (LOQ):
Several approaches for determining the quantitation limit are possible, depending on whether the procedure is non-instrumental or instrumental. LOQ may be expressed as:
QL=10×σ/s
Where,
σ = Response of standard observation
S = The slope of the calibration curve
The slope S may be estimated from the calibration curve of the analyte.The estimate of σ may be carried out in a variety of ways. LOD can calculate from the linearity correlation equation.
RESULT AND DISCUSSION:
Figure no. 1: Diluent Chromatogram of Ruxolitinib Contractions (132μg/ml
Figure no. 2: Standard Chromatogram of Ruxolitinib Contractions (132μg/ml)
Figure no. 3: Sample Chromatogram of Ruxolitinib Contractions (132μg/ml)
Table no. 2: Result of Sample Chromatogram of Ruxolitimb Contractions
Solution |
Mean peak area |
%RSD |
Standard |
2349764 |
0.11 |
Sample |
2292161 |
0.49 |
Table No 3: Linearity Data of Ruxolitinib
Concentration of RUXOLITINIB (µg/ml) (x) |
Area (y) |
25.415 |
560868 |
50.829 |
1122142 |
81.327 |
1796699 |
101.659 |
2276371 |
152.488 |
3338381 |
Table No 4: Data of regression analysis of Ruxolitinib
Drug
|
Straight line equation of calibration curve |
Correlation coefficient (r) |
Ruxolitinib |
Y=21936.2900+12577.8888 |
0.9996 |
Acceptance Criteria:
The correlation co-efficient (r) value should not be less than 0.9996 over the working range.
Table No 5: Precision Repeatability:
Set No. |
Area |
% Assay |
Set-1 |
2253624 |
100.9 |
Set-2 |
2252698 |
100.8 |
Set-3 |
2251469 |
100.7 |
Set-4 |
2250469 |
100.8 |
Set-5 |
2235625 |
100.3 |
Set-6 |
2226548 |
99.8 |
Mean |
2245072.16 |
100.55 |
% Standard Deviation |
11258.11 |
0.4230 |
%RSD |
0.5 |
0.4 |
Acceptance Criteria: % RSD of values of six sample preparation should not be more than 2.0%
Table No. 6: Inter day Precision
Sr No. |
Area |
%Assay |
Set -1 |
2256452 |
101 |
Set-2 |
2254896 |
101 |
Set-3 |
2247692 |
101.2 |
Set-4 |
2246859 |
100.6 |
Mean |
2251474 |
100.94 |
% Standard Deviation |
4902.11 |
0.2516 |
% RSD |
0.22 |
0.25 |
Acceptance Criteria:
% RSD of values of sample preparation should not be more than 2.0%. Cumulative RSD value of precision and Interday precision should not more than 2.0%
CONCLUSION:
The %RSD value for Ruxolitinib is Well within the limit of acceptance criteria and % cumulative RSD value of precision and interday precision is well within the limit of acceptance criteria that means the method was precise for intended use for tiopronin tablets assay in market formulation
Accuracy:
Table no 7: Accuracy Data
Accuracy level |
Test amount |
Drug taken |
Spiked standard amoount |
Total amount recover in |
% recovery |
50% |
132 |
66 |
198 |
197.490 |
99.74 |
100% |
132 |
132 |
264 |
263.612 |
99.85 |
150% |
132 |
264 |
396 |
395.164 |
99.78 |
Calculations:
% Assay of Ruxolitinib=
AuAs×w150×550×50w2×505×P100×100
96.924% |
Au= Average peak area of Ruxolitinib in sample solution
As= Average peak area of Ruxolitinib in standard solution
W1=Standard weight of Ruxolitinib
W2=Sample weight of Ruxolitinib
P= Potency of Ruxolitinib
Acceptance criteria: NLT 90% and NMT 105%
SUMMARY:
Table No. 8: Result Parameters
Sr No |
Parameter |
Result |
1 |
Specificity |
All the peaks were separate and identical |
2 |
Regression equation |
Y=21936.2900x+12577.8888 |
3 |
correlation coefficients (R)2 |
0.9996 |
4 |
Repeatability (%RSD |
0.25% |
5 |
Inter day Precision (%RSD) |
0.4% |
6 |
Range (µg/ml) |
25.415-152.488 µg/m |
7 |
Assay |
96.924% |
8 |
Accuracy (%Recovery) |
99.74 -99.85% |
CONCLUSION:
Based on the empirical evidence, the present method was strongly claimed about the novelty of the developed method over the reported methods. This is the first stability-indicating method, which is ‘rapid’ because it significantly reduced the total analysis time which is the lowest analysis time required. The method justifies “easy” because the proposed method does not involve use of dual-wavelength, gradient techniques. The present method is “stability indicating” as this has been shown less degradation pattern in stressed conditions and good separation of ruxolitinib among the other degraded peaks. The method considers “validated” because all the results of validation parameters were found within the limits as per the ICH Q2B guidelines. Hence the present developed method can be designate as reliable, validated and highly useful for the routine analytical and quality control study of the ruxolitinib in the tablet dosage form.
REFERENCES:
1. Vainchenker W, Kralovics R. Genetic basisand molecular pathophysiology of classical myeloproliferative neoplasms. Blood. 2017; 129(6): 667-679.
2. Jakafi® (Ruxolitinib) tablets, for oral use [prescribing information]. Wilmington, DE: Incyte Corporation. Available from: http://www.incyte.com/products/uspi_jakafi.pdf. Accessed September 9, 2013.
3. Rodrigues MA, Torres T. JAK/STAT inhibitors for the treatment of atopic dermatitis. J Dermatolog Treat. 2020; 31: 33–40. https://doi. org/ 10. 1080/ 09546 634. 2019. 15775 49.
4. Solimani F, Meier K, Ghoreschi K. Emerging topical and systemicJAK inhibitors in dermatology. FrontImmunol. 2019; 10: 2847. https:// doi. org/ 10. 3389/ fimmu. 2019. 02847.
5. Howell MD, Kuo FI, Smith PA. Targeting the Janus kinase family in autoimmune skin diseases. Front Immunol. 2019; 10: 2342. https:// doi. org/ 10. 3389/ fimmu. 2019. 02342.
6. Welsch K, Holstein J, Laurence A, Ghoreschi K. Targeting JAK/STAT signalling in inflammatory skin diseases with small moleculeinhibitors. Eur J Immunol. 2017; 47: 1096–107. https:// doi.org/ 10. 1002/ eji. 20164 6680.
7. Rerknimitr P, Otsuka A, Nakashima C, Kabashima K. The etiopathogenesis of atopic dermatitis: barrier disruption, immunological derangement, and pruritus. Inflamm Regen. 2017; 37: 14. https:// doi. org/ 10. 1186/ s41232- 017- 0044-7.
8. Brunner PM, Guttman-Yassky E, Leung DY. The immunology of atopic dermatitis and its reversibility with broad-spectrum and targeted therapies. J Allergy Clin Immunol. 2017; 139: S65-s76. https:// doi. org/ 10. 1016/j. jaci. 2017. 01. 011.
9. Gandhi NA, Bennett BL, Graham NM, Pirozzi G, Stahl N, Yancopoulos GD. Targeting key proximal drivers of type 2 inflammation in disease. Nat Rev Drug Discov. 2016; 15: 35–50. https:// doi. org/10. 1038/ nrd46 24.
10. Rogatsky E. Modern high performance liquid chromatography and HPLC 2016 International Symposium. J Chromatogr Sep Tech. 2016; 7: e135.
11. Mulubwa M, et al. Development and validation of high performance liquid chromatography tandem mass spectrometry (HPLC-MS/MS) method for determination of tenofovir in small volumes of human plasma. J Chromatogr Sep Tech. 2015; 6: 300.
12. Santini DA, et al. Development of a high performance liquid chromatography method for the determination of tedizolid in human plasma, human serum, saline and mouse plasma. J Chromatogr Sep Tech. 2015; 6: 270. 14. Lin G, et al. Determination of sodium tanshinone iia sulfonate in rat plasma by high performance liquid chromatography and its application to pharmacokinetics studies. Pharm Anal Acta. 2015; 6: 383.
13. AL-Jammal MKH, et al. Development and validation of micro emulsion high performance liquid chromatography(MELC) method for the determination of nifedipine in pharmaceutical preparation. Pharm Anal Acta. 2015; 6: 347.
14. Myron P, et al. Tributylamine facilitated separations of fucosylated chondroitin sulfate (fucs) by high performance liquid chromatography (HPLC) into its component using 1-phenyl- 3-methyl-5-pyrazolone (pmp) derivatization. J Chromatogr Sep Tech. 2015; 6: 256.
15. Tang M, et al. HPLC analysis of monomer release from conventionally and high temperature high-pressure polymerised urethane dimethacrylate intended for biomedical applications. J Chromatograph Separat Techniq. 2014; 5: 227.
16. Yuri Kazakevich, Rosario Lobrutto. HPLC for pharmaceutical scientists. United States of America. A John Wiley and Sons, Inc, Publication; 2007: 3.
17. International Conference on Harmonization of Technical Requirements for Registration of Pharmaceuticals for Human use. Validation of Analytical Procedures: Text and Methodology ICH Q2 (R1). 2005.
18. Ewing GW. Instrumental methods of chemical analysis, 5th ed., McGraw Hill International Publication. Chemistry Series. 1985: 1-4.
Received on 01.09.2023 Modified on 11.01.2024
Accepted on 01.03.2024 © RJPT All right reserved
Research J. Pharm. and Tech 2024; 17(7):3040-3044.
DOI: 10.52711/0974-360X.2024.00475