Method Development and Validation for the Estimation of Umifenovir in Bulk and Tablet Formulation by UV-Spectrophotometric and RP-HPLC
Komal Somkuwar*, Prafulla Sabale, Vaibhav Sawale, Priya Rahangdale
Department of Pharmaceutical Sciences, R. T. M. Nagpur University,
Mahatma Jyotiba Fuley Educational Campus, Amravati Road, Nagpur - 440033, Maharashtra, India.
*Corresponding Author E-mail: sweetsomkuwar@gmail.com
ABSTRACT:
Umifenovir, an antiviral drug that is used to treat influenza, has recently been used in COVID-19 infection. According to a literature survey, no UV technique for the estimation of umifenovir has yet been established; hence, there is an imperative need for a simple analytical method. Additionally, we developed an alternative, RP-HPLC method for the estimation of umifenovir. UV spectrophotometry was carried out at 223nm absorption maxima using the solvent methanol. A concentration range of 2-12µg/ml was found to obey Beer's Law, with a correlation coefficient (r2) of 0.9995. A C-18 column (250mm 4.6µm, 5µm) was used for chromatographic separation using isocratic mode. The mixture consisted of acetonitrile: 0.1% trimethylamine (pH adjusted to 2.7 by the addition of orthophosphoric acid) 60:40 as the mobile phase with a flow rate of 1.0 ml/min. The temperature was kept at 25ºC, and detection at 223nm was performed using a PDA detector. The estimated percentage of the drug was close to 100%, corresponding to the label claim of the tablet made in the laboratory. The results and statistical study demonstrated the utility of the current methods in the routine evaluation of umifenovir bulk and formulation.
KEYWORDS: Umifenovir, UV Spectroscopy, RP-HPLC, Assay and Validation.
INTRODUCTION:
Umifenovir (UMI) (1-methyl-2-((phenylthio)methyl)-3-carbethoxy-4-((dimethylamino)methyl)-5-hydroxy-6-bromindole hydrochloride. Fig. 1, is an antiviral drug and indole derivative1-2. UMI is marketed under the brand name Arbidol and is a broad-spectrum antiviral used to treat influenza in both China and Russia. UMI is a fusion inhibitor. It works by blocking the fusion of the virus and the target host cells. It also stimulates the immune system. UMI is effective against a variety of enveloped and nonenveloped viruses3. It has also been used to treat influenza virus A and B viruses and has recently been used to treat hepatitis C virus. UMI has also been shown to be effective in inhibiting COVID-19 infection by interfering with the release of SARS-CoV-2 from intracellular vesicles4,5,6.
Figure 1: Chemical structure of UMI
According to a literature survey the drug has been estimated in biological fluids such as human urine, plasma, and rat plasma, Wang et al., identified the metabolites in human urine and studied the fragmentation pathways using HPLC coupled with ion trap mass spectrometry7. Liu X, et al., developed a method based on cloud-point extraction to determine arbidol in rat plasma by HPLC and UV detection8 and used the LC-ESI-MS method for arbidol determination in human plasma9 while Annampura et al., established a stability-indicating LC method for the determination of UMI in tablets10 and quantified arbidol by RP-HPLC with PDA detection11. Surabhi et al. validated the stability-indicating method for the determination of UMI-remdesivir in the presence of its degradation products12. Damle et al. developed a validation HPTLC method for the determination of arbidol from a pharmaceutical formulation13. As far as we know, there is no UV spectrophotometric method for the determination of UMI. Therefore, we believe that developing a UV method is worth exploring, and very few HPLC methods have been reported, to provide an alternative HPLC method that is simple, has high reproducibility, has a good retention time, has more percent recovery and utilizes very less time of analysis, i.e., 6 minutes to the existing methods for estimation of UMI in API and in tablet formulation along with validation of the methods as referred to in ICH guidelines14.
UMI is available as a tablet (Arpeflu; label claim: 50 mg; 100mg) and capsule (Arbidol; label claim: 100mg). Tablets were prepared in the laboratory as formulations were not available in India.
Double distilled water was obtained from the Millipore unit. HPLC grade acetonitrile (ACN), methanol, orthophosphoric acid and triethylamine (TEA) were all purchased from Sisco Research Laboratories Pvt. Ltd. Mumbai, India.
Instrumentation:
A double-beam UV-1800 Shimadzu UV spectrophotometer with a pair of 10mm matched quartz cells was used. The HPLC system was a Shimadzu model no. DGU-20A5R, a Shimadzu ATX224 digital analytical balance and a PCi analytics ultrasonic bath.
Sample preparation for the UV method development:
Standard stock solution (100μg/ml):
5mg of precisely measured pure UMI was poured into a clean dry 50ml volumetric flask, methanol was used for dissolution, and the volume was made up to 50ml.
Standard working solution (10μg/ml):
1ml from the standard stock solution was transferred and dissolved in 10ml of methanol.
Sample solution for the UV assay method:
Ten tablets (containing 100mg of UMI) were precisely weighed and prepared in our laboratory, and the average weight of the tablets was determined. Tablets were reduced to a powdery consistency. The tablet powder equivalent to 5mg of UMI was weighed and further transferred into a 50ml volumetric flask. The content was dissolved in methanol and then sonicated for ten minutes. The volume was brought up to the mark (Conc.100μg/ml) and the solution was filtered with Whatman filter paper No. 41. From the clear sample stock solution, appropriate dilutions were made to obtain a concentration of 10µg/ml sample solution.
Sample preparation for the HPLC method development:
Standard stock solution (A1) (1000 μg/ml):
10mg of precisely measured pure UMI was poured into a clean dry 50ml volumetric flask, acetonitrile was used for dissolution and volume was made up to 50ml.
Standard stock solution (A2) (100μg/ml):
1ml of A1 stock solution was added to a 10ml volume flask, and the volume was made up to the mark with acetonitrile.
Standard working solution (A3) (10μg/ml):
1ml of A2 stock solution was added to a 10ml volume flask, and the volume was made up to the mark with acetonitrile.
Mobile phase preparation:
The mobile phase consisted of 60:40% v/v acetonitrile and 0.1% trimethylamine (with orthophosphoric acid pH adjusted to 2.7).
Sample preparation for HPLC assay:
Ten tablets (containing 100mg of UMI) were accurately weighed and prepared in our laboratory, and the average weight was determined. Tablets were reduced to a powdery consistency. The tablet powder equivalent to 10mg of UMI was weighed and further transferred into a 50ml volumetric flask. The content was dissolved in methanol and then sonicated for ten minutes. The volume was brought up to the mark with methanol (Conc.100μg/ml), and the solution was filtered with Whatman filter paper No. 41. From the clear sample stock solution, appropriate dilutions were made to obtain a concentration of 10µg/ml sample solution.
Validation of the Proposed Method:
System suitability:
It is considered under every validation parameter. Five replicate injections of 100 μg/ml drug standard solution were auto injected. Parameters including the tailing factor and theoretical plate counts were noted.
Linearity and Range:
It is the ability to obtain results that are proportional to the analyte concentration. From a standard stock solution of UV, 0.2, 0.4, 0.6, 0.8, 1.0, and 1.2ml aliquots were added to a series of 10 volumetric flasks, and the volume was brought to the mark by methanol, resulting in concentrations of 2, 4, 6, 8, 10 and 12µg/ml respectively. At 223nm their absorbance was measured against a blank. Similarly, 5-30µg/ml concentrations from the working standard solution of HPLC were prepared and auto injected with optimized chromatographic parameters.
Accuracy:
It is the assessment of the difference between the measured value and the real value. Accuracy was measured at 80%, 100%, and 120% by spiking a standard solution of UMI (0.8, 1.0, 1.2μg/ml) into the sample solution. The percentage recoveries were calculated.
Precision:
It is the degree of agreement under specified conditions between a set of measurements acquired from multiple samplings of homogeneous samples. This was demonstrated by intraday and interday variation studies. Six sample solutions were prepared, and %RSD was calculated.
Intraday and Interday variation:
The absorbance of the standard working solution (10 µg/ml) was recorded by UV at 0hours, 6hours, the 2nd day and the 3rdday. Similarly, for the HPLC method, a 10μg/ml concentration of six standard solutions of UMI was injected and chromatograms were recorded.
RESULTS AND DISCUSSION:
Determination of wavelength of maximum absorbance (λmax) for UMI:
The working standard solution of UMI (10μg/ml) was scanned in the range of 200-400nm in a 1.0cm matched quartz cell against ablank (methanol) and the spectrum was recorded. From the UV spectra, λmax was found to be 223nm, which was selected as the analytical wavelength for further analysis. The spectrum was recorded as shown in Fig. 2
Figure 2: UV spectrum of 10μg/ml UMI standard solution.
Preliminary optimization of the mobile phase and other chromatographic conditions:
To obtain the best possible chromatographic results, one or more parameters were adjusted in each trial. The trial runs are shown in Table 1.
Table 1: Method optimization (Trial runs)
|
Mobile Phase |
Flow rate (ml/min) |
pH |
Injection Vol. (µl) |
Retention time (min) |
Comment |
|
ACN: Methanol (50:50) |
0.6 |
- |
10 |
- |
Peak did not appear |
|
ACN: Methanol (70:30) |
0.6 |
- |
10 |
- |
Peak did not appear |
|
ACN: TEA (1%) (50:50) |
0.6 |
3.2 |
10 |
9.712 |
Peak show Tailing |
|
ACN: TEA(1%) (55:45) |
0.8 |
3.2 |
10 |
8.501 |
Peak show Tailing |
|
ACN: TEA (1%) (55:45) |
0.8 |
3.0 |
10 |
4.666 |
Peak show Tailing |
|
ACN: TEA (1%) (60:40) |
1.0 |
3.0 |
20 |
3.650 |
Peak show Tailing |
|
ACN: TEA (1%) (60:40) |
1.0 |
2.7 |
20 |
3.515 |
Sharp peak appear with no tailing |
Based on the trial, the following chromatographic optimized parameters were chosen for further study.
Table 2: Optimized chromatographic parameters
|
Column |
Shimatzu C18 (250mm × 4.6mm i.d., 5µm particle size) |
|
Mobile Phase |
Acetonitrile:0.1% trimethylamine (60:40) (pH 2.7 adjusted with orthophosphoric acid) on isocratic mode |
|
Flow rate |
1 ml/min |
|
Wavelength |
223 nm |
|
Injection volume |
20µl |
|
Column oven temperature |
25 ˚c |
|
Retention time |
3.515 min |
|
Total run time |
6 min |
A 20µL aliquot of solution was autoinjected into the HPLC system, and the chromatogram was recorded as shown in Fig. 3 and its 3D image in Fig. 4.
Figure 3: A model chromatogram showing the separation of UMI.
Study of system suitability:
After equilibrating the column with the mobile phase, the solution (conc. 10µg/ml) was auto-injected five times and the chromatograms were recorded. The results of the system suitability are reported in Table 3.
Fig. 4: 3D image of UMI chromatogram.
Table 3: Observations of system suitability parameters
|
Sr. No. |
Weight of Std. drug taken (mg) |
Peak area (mV) |
Tailing factor |
Theoretical plates |
|
1 |
10.0 |
798006 |
1.291 |
3313 |
|
2 |
791841 |
1.293 |
3318 |
|
|
3 |
791959 |
1.265 |
3387 |
|
|
4 |
797809 |
1.312 |
3287 |
|
|
5 |
798511 |
1.267 |
3369 |
|
|
% RSD |
0.88 |
0.92 |
0.95 |
|
|
Limit |
NMT 2% |
NMT 2 |
NLT 2000 |
|
Linearity and range:
The UV and HPLC methods confirmed linearity in the ranges of 2-12µg/ml and 5-30µg/ml, and the linearity equations were y = 0.064x+0.0065 and y = 7523x+ 63074 with correlation coefficients of 0.9995 and 0.9997 respectively. The results are tabulated in Table 4. The calibration plots for UV and HPLC are depicted in Figs. 5 and 6, respectively.
Table 4: Calibration curve data of the UV and HPLC methods
|
UV method |
HPLC method |
||
|
Conc. (µg/mL) |
Absorbance |
Conc. (µg/mL) |
Area |
|
2 |
0.138 |
5 |
424472 |
|
4 |
0.266 |
10 |
837809 |
|
6 |
0.385 |
15 |
1183954 |
|
8 |
0.513 |
20 |
1570907 |
|
10 |
0.644 |
25 |
1944492 |
|
12 |
0.782 |
30 |
2316201 |
|
r2 |
0.9995 |
r2 |
0.9997 |
Figure5: The calibration curve of UMI for UV
Figure 6: The calibration curve of UMI for HPLC
Accuracy:
The percentage recovery of UMI at different spiked levels (i.e., 80%, 100%, 120%) was ascertained to be in the range of 98.50 – 101.6% and 99.97– 102% for UV and HPLC respectively. The % RSD was less than 2 as shown in Table 5. This means that there is no interference of the excipients when determining umifenovir. Therefore, the method is accurate. Tables 13 and 14, show the percentage recovery results that are statistically compared for the F test and the T test respectively.
Table 5: Results of Accuracy (Recovery) Studies
|
Sr. No. |
Conc.(%) |
Tablet (μg/ml) |
Pure drug (μg/ml) |
Total amount (μg/ml) |
UV Method |
HPLC Method |
||||
|
Actual amount found (μg/mL) |
% Recovery |
% RSD |
Actual amount found (μg/mL) |
% Recovery |
% RSD |
|||||
|
1 |
80 |
10 |
8 |
18 |
17.98 |
99.85 |
0.44 |
18.09 |
101.2 |
0.60 |
|
2 |
80 |
10 |
8 |
18 |
18.05 |
100.71 |
18.06 |
100.75 |
||
|
3 |
80 |
10 |
8 |
18 |
18.10 |
100.12 |
18.0 |
100 |
||
|
4 |
100 |
10 |
10 |
20 |
19.85 |
98.50 |
1.59 |
20.1 |
101 |
0.60 |
|
5 |
100 |
10 |
10 |
20 |
20.16 |
101.60 |
20.2 |
102 |
||
|
6 |
100 |
10 |
10 |
20 |
19.94 |
99.42 |
20.09 |
100.9 |
||
|
7 |
120 |
10 |
12 |
22 |
22.08 |
100.60 |
0.27 |
21.99 |
99.97 |
0.19 |
|
8 |
120 |
10 |
12 |
22 |
21.95 |
100.09 |
22.04 |
100.33 |
||
|
9 |
120 |
10 |
12 |
22 |
22.02 |
100.19 |
22.0 |
100 |
||
%RSD = percentage relative standard deviation
Precision:
It was studied by analyzing homogenous samples intraday and interday. The results are shown in Table. 6. The %RSD for UMI of the standard solution was less than 2, which significantly assures the precision of the proposed method.
Table 6: Intraday andInterday Precision Result
|
UV Method |
||||||
|
Concentration (10 μg/ml) |
Intraday Study (n=6) |
Interday Study (n=6) |
||||
|
0 Hour |
6 Hours |
2nd Day |
3rd Day |
|||
|
*Mean Absorbance |
0.685 |
0.637 |
0.566 |
0.536 |
||
|
Standard deviation |
0.006 |
0.0054 |
0.0047 |
0.0045 |
||
|
% RSD |
0.88 |
0.85 |
0.83 |
0.84 |
||
|
HPLC Method |
||||||
|
Concentration (10 μg/ml) |
Intraday Study (n=6) |
Interday Study (n=6) |
||||
|
0 Hour |
6 Hours |
2nd Day |
3rd Day |
|||
|
*Mean Area |
794781 |
784506 |
772668 |
764276 |
||
|
Standard deviation |
7104.50
|
6048.82
|
6698.09 |
6281.27 |
||
|
% RSD |
0.89 |
0.77 |
0.87 |
0.82 |
||
n = number of measurements, %RSD = percentage relative standard deviation, *Mean of six observations
Ruggedness:
It was determined by performing the same procedure by two different analysts and two different UV spectrophotometers were also used for the study. The results are reported in Tables 7 and 8.
Table 7: Ruggedness Study Result (Analysis Performed by Different Analysts)
|
UV Method |
||
|
Conc. (10 μg/ml) |
Analyst 1 |
Analyst 2 |
|
*Mean Absorbance |
0.653 |
0.662 |
|
Standard deviation |
0.0060 |
0.0059 |
|
% RSD |
0.92 |
0.89 |
|
HPLC Method |
||
|
Conc. (10 μg/ml) |
Analyst 1 |
Analyst 2 |
|
*Mean Area |
794455 |
794788 |
|
Standard deviation |
7576.24 |
7101.41 |
|
% RSD |
0.95 |
0.85 |
RSD= Relative standard deviation, *Mean of six observations
Table 8: Ruggedness Study (Analysis Performed on Different Instruments)
|
Conc. (10 μg/ml) |
Instrument Model & Make |
|
|
UV-1800 Shimadzu UV Spectrometer |
Perkin Elmer UV-Visible Spectrometer |
|
|
*Mean Absorbance |
0.653 |
0.638 |
|
Standard deviation |
0.0064 |
0.0059 |
|
% RSD |
0.98 |
0.92 |
RSD = Relative standard deviation,*Mean of six observations
Robustness:
In the proposed UV method, to validate the robustness parameter, slight variation was employed in wavelength (±2nm), and HPLC robustness was assessed by introducing small, intentional variations in the chromatographic conditions which comprised the percent of acetonitrile in the mobile phase (±5%), the flow rate of mobile (±0.2ml/min) and pH (±1). The data are represented in Tables 9 and 10. The %RSD was less than 2, which significantly represents the robustness of the method.
Table 9: Robustness Study for UV
|
Conc. (10 μg/ml) |
At 221 nm |
At 225 nm |
|
*Mean Absorbance |
0.687 |
0.658 |
|
Standard deviation |
0.0062 |
0.0059 |
|
% RSD |
0.90 |
0.89 |
*Mean of six observations, RSD = Relative standard deviation
Table 10: Robustness Study for HPLC
|
Parameters |
*Mean Peak Area (mV) |
Standard deviation |
% RSD |
|
ACN: TEA (55:45) |
798005 |
7144.50 |
0.86 |
|
ACN: TEA (65:35) |
799733 |
9424.45 |
0.95 |
|
Flow rate (0.8ml/min) |
791844 |
7576.23 |
0.99 |
|
Flow rate (1.2ml/min) |
792364 |
7860.87 |
0.98 |
|
pH (2.6) |
789065 |
7112.30 |
0.99 |
|
pH (2.8) |
792011 |
7781.20 |
0.95 |
Sensitivity:
The LOD and LOQ values computed from the stated formulae are reported in Table 11.
Table 11: Sensitivity result
|
Sensitivity |
UV Method |
HPLC Method |
|
LOD (μg/ml) |
0.58 |
0.29 |
|
LOQ (μg/ml) |
1.75 |
0.88 |
Assay procedure of UMI by UV and HPLC methods:
A standard working solution of UMI (conc. 10µg/ml) and sample solution (conc. 10µg/ml) were used, and the absorbance of both solutions was measured at 223nm. Similarly, both solutions were autoinjected into HPLC. The content of UMI present in the laboratory prepared tablet was evaluated by UV and HPLC. The obtained results of the % label claim by UV and HPLC are shown in Table 12. The % contents of UMI obtained by UV and HPLC were statistically compared for the F test and T test, as shown in Tables 13 and 14, respectively.
Table 12: Results of Analysis of Tablet Formulation
|
Methods |
Sample |
Label Claim (100mg/tablet) |
% Label claim estimated (% content) |
Mean |
% RSD |
|
UV |
Laboratory prepared Tablet |
100 |
99.24 |
99.54 |
0.23 |
|
99.57 |
|||||
|
99.45 |
|||||
|
99.51 |
|||||
|
99.93 |
|||||
|
99.53 |
|||||
|
HPLC |
Laboratory prepared Tablet |
100 |
99.97 |
99.90 |
0.83 |
|
98.79 |
|||||
|
100.87 |
|||||
|
99.16 |
|||||
|
100.77 |
|||||
|
99.85 |
%RSD = percentage relative standard deviation
Statistical comparison between % contents and % recovery study of assay samples by UV and HPLC methods:
F Test:
Table 13: Observations and results of the Ftest for % contents and % recovery study
|
Parameter |
Variance |
F Value |
P (F<=f) one tail |
F critical one tail |
|
|
UV Method |
HPLC Method |
||||
|
% Content |
0.0504 |
0.6970 |
13.8274 |
0.0059 |
5.0503 |
|
% Recovery |
0.7453 |
0.4650 |
0.6239 |
0.2599 |
0.2909 |
From the above F test results, it could be inferred that the % content and % recovery F values were higher than the F-critical values; hence, the null hypothesis is rejected.
T Test:
Table 14: Observations and results of T test for % contents and % recovery study
|
Parameter |
t Stat |
P(T<=t) one tail |
t Critical one tail |
P(T<=t) two-tail |
t Critical two-tail |
|
%Content |
1.0954 |
0.1616 |
2.0150 |
0.3233 |
2.5706 |
|
%Recovery |
1.6823 |
0.06551 |
1.8595 |
0.1310 |
2.3060 |
From the above T test results, it could be inferred that the P value is higher than the alpha value (P = 0.05). Hence the null hypothesis is accepted, i.e., there is a statistically significant difference between the mean % content and % recovery determined by UV and HPLC.
The proposed UV and HPLC methods were found to be simple, precise, fast, robust and rugged. The % RSD was less than 2%. Our studies revealed a recovery percentage ranging from 98 to 102%, indicating that the developed methods were accurate. The method did not interfere with excipients while determining umifenovir. Its chromatographic run time of 6 minutes allows the analysis of samples in a short period. The simple UV method could be implemented in laboratories that lack high-tech analytical instruments, which are complicated, expensive and time consuming. Based on the statistical tests, the F test and the T test, it can be concluded that the HPLC method is better than the UV method for drug determination. Therefore, the developed UV method and HPLC method can be used for routine quality control of UMI in bulk and tablet formulation.
ACKNOWLEDGMENT:
The authors are grateful to Mylan Labs Ltd. (India) for providing the gift samples of UMI and to colleague Pranita Kanojiya for helping in the preparation of tablets in the laboratory.
REFERENCES:
1. Balzarini J, Ruchko EA, Zakharova EK, Kameneva IY, Nawrozkij MB. Structural analogs of umifenovir.
2. Blaising J, Polyak SJ, Pécheur EI. Arbidol as a broad-spectrum antiviral: An update. Vol. 107, Antiviral Research. Elsevier B.V.; 2014. p. 84–94.http://dx.doi.org/10.1016/j.antiviral.2014.04.006.
3. Deng P, Zhong D, Yu K, Zhang Y, Wang T, Chen X. Pharmacokinetics, metabolism, and excretion of the antiviral drug arbidol in humans. Antimicrob Agents Chemother. 2013; 57(4): 1743–55.https://doi:10.1128/AAC.02282-12
4. Nojomi M, Yassin Z, Keyvani H, Makiani MJ, Roham M, Laali A, et al. Effect of Arbidol (Umifenovir) on COVID-19: a randomized controlled trial. BMC Infect Dis. 2020; 20(1): 1–10. https://doi.org/10.1186/s12879-020-05698-w
5. Wright ZVF, Wu NC, Kadam RU, Wilson IA, Wolan DW. Structure-based optimization and synthesis of antiviral drug Arbidol analogues with significantly improved affinity to influenza hemagglutinin. Bioorganic Med Chem Lett. 2017; 27(16): 3744–8.https://doi:10.1016/j.bmcl.2017.06.074.
6. Yuya Wang, Xiaoyan Chen QL and DZ. Metabolite identification of arbidol in human urine by the study of CID fragmentation pathways using HPLC coupled with ion trap mass spectrometry. J mass Spectrom. 2008;43:1099–1109.https://doi: 10.1002/jms.1394
7. XiaoLiuXiao-HuiChenYuan-YuanZhangWen-TaoLiuKai-ShunBi. Determination of arbidol in rat plasma by HPLC–UV using cloud-point extraction. J Chromatogr B. 2007; 856(1–2): 273–7. https://doi:10.1016/j.jchromb.2007.06.009
8. Liu X, Huang YW, Li J, Li XB, Bi KS, Chen XH. Determination of arbidol in human plasma by LC-ESI-MS. J Pharm Biomed Anal. 2007; 43(1):371–5.https://doi:10.1016/j.jpba.2006.06.039
9. Annapurna MM. New stability indicating ultrafast liquid chromatographic method for the determination of umifenovir in tablets. International Journal of Green Pharmacy. 2018; 11(9): 3851–6. https://doi.org/10.22377/ijgp.v12i01.1619
10. Narendra A, Annapurna MM. Quantification of arbidol by RP-HPLC with photo diode array detection. Asian J Pharm. 2018;12(2):S553–7. https://doi.org/10.22377/ajp.v12i02.2391
11. Surabhi SR. Validated Stability Indicating Method for Determination of Umifenovir- Remdesivir in Presence of Its Degradation Products. Int J Dev Res. 2021; 11(model 2695): 46227–32.https://doi.org/10.37118/ijdr.21666.04.2021
12. Damle MC, Phadtare SA, Raskar HD, Gadge KG, Mehendre R, Bothara KG. A Validated HPTLC Method for Determination of Arbidol from Pharmaceutical Formulation. Vol. 2, International Journal of ChemTech Research CODEN( USA).
13. National Center for Biotechnology Information. PubChem Compound Summary for CID 131411, Arbidol. https://pubchem.ncbi.nlm.nih.gov/compound/Arbidol. Accessed May 6, 2023.
14. ICH Validation of Analytical Procedures: Text and Methodology Q2 (R1), International Conference on Harmonization; 2005.
Received on 19.07.2023 Modified on 05.10.2023
Accepted on 14.11.2023 © RJPT All right reserved
Research J. Pharm. and Tech 2024; 17(3):1308-1313.
DOI: 10.52711/0974-360X.2024.00205