Method Development and Validation for Estimation of Istradefylline in Tablet Dosage form by RP-HPLC
Krishna Kishore Adireddy*, Srinivasa Rao Baratam, Nagarjuna Hari Pratap S
Department of Pharmaceutical Analysis, St. Ann's College of Pharmacy, Vizianagaram,
Andhra Pradesh-535003, India.
*Corresponding Author E-mail: adireddi.kk@gmail.com
ABSTRACT:
A simple, rapid, accurate and precise RP-HPLC method was developed and validated for the determination of Istradefylline in table dosage form. Chromatographic analysis of the drug was achieved on Shimadzu HPLC comprising of LC- 20 AD binary gradient pump, a variable wavelength programmable SPD-20A detector and SCL system controller. C18G column (250 mm x 4.6 mm, 5 μ) as stationary phase with mobile phase consisting of 0.1 % orthophosphoric acid and acetonitrile in the ratio of 30: 70 v/v. The method showed a good linear response in the concentration range of 10-90 μg/ml with correlation coefficient of 0.9993. The flow rate was maintained at 1.0 ml/min and detection was carried out at 246 nm. The retention time was 3.125 min. The method was statistically validated for accuracy, precision, linearity, ruggedness, robustness, solution stability, selectivity and sensitivity. The results obtained in the study were within the limits of ICH guidelines and hence this method can be used for the determination of istradefylline in tablet formulation.
KEYWORDS: Istradefylline, RP-HPLC, Tablet dosage form.
INTRODUCTION:
Istradefylline, chemically 8-[(E)-2-(3,4-dimethoxyphenyl) ethenyl]-1,3-diethyl-7-methyl-2,3,6,7-tetrahydro-1H-purine-2,6-dione having molecular formula C20H24N4O4 and molecular weight 384.43 g/mol. Istradefylline is a light yellow-green crystalline powder. Istradefylline has a dissociation constant (pKa) of 0.78. The aqueous solubility of istradefylline is ~0.5 μg/mL across the physiological pH range and 0.6 μg/mL in water. Istradefylline is a kind of selective adenosine A2A receptor antagonists, by the exploitation of Japan Kyowa Hakko Kirin company. Istradefylline tablets (trade name Nouriast, specification 20 mg) obtains Japan PMDA approval listing in March, 2013. Istradefylline was granted FDA approval on 27 August 2019. It was clinically used to improve dyskinesia in the early stages of PD and prolong the half life period of L-dopa and improve the wearing off phenomenon and other side effects produced by L-dopa, further prolong the recurrence time of patients. Istradefylline is considered as a suitable drug in the treatment of Parkinson’s disease1, unlike standard dopaminergic therapies for Parkinson's, Istradefylline targets adenosine A2A receptors in the basal ganglia2,3. This region of the brain is highly involved in motor control. Istradefylline is indicated as an adjunct treatment to levodopa and carbidopa for Parkinson's disease. Literature survey reveals some analytical methods reported related to istradefylline which estimation of istradefylline and carbidopa by LC-MS4, Photostability studies by HPLC5, related substances and impurity studies of istradefylline6. However there is no analytical method reported yet for estimation of istradefylline in dosage form. The spectrophotophotometric methods7 have a lack of selectivity and sensitivity, so chromatographic method is a method of choice for simultaneous estimation of istradefyllin. Out of many chromatographic method , HPLC8-18 is a versatile analytical technique with good sensitivity and selectivity. The aim of this work was to develop an HPLC method for estimation of istradefylline in tablet dosage form.
Fig. 1: Structure of Istradefylline
MATERIALS AND METHODS:
Instrumentation
Chromatographic separation was performed on a Shimadzu LC-20AD HPLC system equipped with a C18 G column (250 × 4.6 mm i.d, 5 μm particle), binary pumps, degasser, variable wave length detector and Rheodyne injector with 20 μl loop volume. ‘LC solution’ software was used to collect and process the data.
Chemicals and Reagents:
Istradefylline pure drug (purity 98.65 %) was purchased online. Marketed formulation Nourianz™ tablets (Kyowa Kirin, Inc) labeled to contain 20 mg of Istradefyllin procured online. HPLC grade water and acetonitrile was from MERCK India Ltd. HPLC grade methanol was from standard reagent Pvt Ltd Hyderabad. Analytical grade orthophosphoric acid was from SD Fine chemicals Mumbai, India. Nylon membrane filters 0.2 µm and 0.45 µm were from PALL life sciences Mumbai, India.
Chromatographic conditions:
The chromatographic system used for the method development and validation consisted of Shimadzu HPLC comprising of LC-20AD binary gradient pump, a variable wavelength programmable SPD-20A detector and an SCL 20A system controller. A Rheodyne injector 7725i fitted with a 20 μL loop was used and data were recorded and evaluated by use of LC solutions software version 5.0. Separation was performed on a Enable C18G (250 × 4.6 mm i.d.,5µ) at ambient temperature. A mixture of 0.1% ortho-phosphoric acid and acetonitrile in the ratio of 30: 70 v/v was found to be the most suitable mobile phase for ideal chromatographic analysis of istradefylline. The solvent mixture was filtered through 0.45 μ membrane filter and sonicated before use. It was pumped through the column at a flow rate of 1.0 mL/min. Injection volume was 20 μL and the column was maintained at room temperature. The column was equilibrated by pumping the mobile phase through the column for at least 30 minutes prior to the injection of the drug solution. The detection of the drug was monitored at 246 nm. The run time was set at 10 min. The optimized chromatographic conditions are shown in Table 1.
Table 1: Optimized chromatographic conditions
|
Parameters |
Conditions |
|
Stationary Phase (Column) |
C18 (250 × 4.6 mm i.d.,5µ) |
|
Mobile Phase |
Acetonitrile: 0.1% ortho-phosphoric acid (70:30,v/v) |
|
Flow rate (ml/min) |
1.0 mL/min |
|
Run time (min) |
5 min |
|
Column temperature (°C) |
Ambient |
|
Volume of injection loop (μL) |
20 |
|
Detection wavelength (nm) |
246 nm |
|
Retention time (min) |
3.125 |
Fig. 2: Chromatogram of standard solution of Istradefylline
Preparation of mobile phase:
Mobile phase was prepared by mixing 700 mL of HPLC grade acetonitrile with 300 mL of 0.1% ortho-phosphoric acid (prepared by mixing 0.1 ml of ortho-phosphoric acid in 100 ml of HPLC grade water. The mobile phase was sonicated for 10 min and filtered through the 0.45 μm membrane filter.
Preparation of standard stock solutions:
The standard stock solutions of 100 μg/mL of the drug were prepared by dissolving 50 mg of pure drug in the mobile phase in a 50 mL volumetric flask and the volume was made up to the mark. Resulting solutions were further diluted with mobile phase to obtain a final concentration of 100 μg/mL and stored under refrigeration. Aliquots of standard stock solutions were put in a 10 mL volumetric flask and diluted up to the mark with mobile phase. In such a way, the final concentrations of the drug were in the range of 10–90 μg/mL.
Preparation of sample solution:
To determine the content of istradefylline in tablet dosage form (Label claim: 20 mg/tablet) ten tablets were accurately weighed and triturated to fine powder. The powder weight equivalent to 10 mg of istradefylline was taken and dissolved in 50 ml of mobile phase. The resulting solution (2.5 ml) was transferred to a 10 ml volumetric flask and diluted up to the mark with mobile phase. The final solution was filtered through 0.45μ membrane filter using injection filter. A 20 µL of the filtrate was injected into chromatographic system. The peak area of the istradefylline was determined and concentration was found using linear regression equation obtained from calibration curve.
Method Validation:
The developed method was validated as per ICH guidelines19 by evaluating linearity, accuracy, precision, robustness, ruggedness, detection limit, quantification limit and stability. Coefficients of variation and relative errors of less than 2 % were considered acceptable.
System Suitability Test:
Before performing validation experiments, system suitability test (SST) has to be applied to indicate that HPLC system and method are capable of providing data with admissible quality. SST was performed by investigating capacity factor, tailing factor, theoretical plates number, and also relative standard deviation (RSD) of the peak areas.
Stability:
Stability was assessed by analyzing QC standard solutions after keeping them at room temperature for 48 h. Obtained results were investigated as recovery values and compared to the freshly prepared solutions.
Linearity:
A stock solution of istradefylline of 1000 μg/mL was prepared with mobile phase. From it, various working standard solutions were prepared in the range of 10 to 120 μg/ml and injected into HPLC. It was shown that the selected drug had linearity in the range of 10–90 μg/mL. The calibration plot (peak area of istradefylline versus istradefylline concentration) was generated by replicate analysis (n=9) at all concentration levels and the linear relationship was evaluated using the least square method within Microsoft Excel® program.
Accuracy
The accuracy of the method was carried out using one set of different standard addition methods at different concentration levels, 80 %, 100 % and 120 %, and then comparing the difference between the spiked value (theoretical value) and actual found value.
Precision:
The precision of the method was ascertained from the peak area obtained by actual determination of six replicates of a fixed amount of the drug (50 μg/mL). The precision of the assay was also determined in terms of intra- and inter-day variation in the peak areas of a set of drug solutions on three different days. The intra- and inter-day variation in the peak area of the drug solution was calculated in terms of relative standard deviation (RSD).
Robustness:
Robustness of the proposed method for istradefylline was carried out by the slight variation in flow rate, analytical wavelength and mobile phase ratio. The percentage recovery and RSD were noted for istradefylline.
Ruggedness:
The test solutions were prepared as per test method and injected under variable conditions. Ruggedness of the method was studied by different analysts.
Detection limit and quantification limit:
The limit of detection (LOD) and limit of quantification (LOQ) were established based on the calibration curve parameters, according to the following formulas:
LOD=3.3SD/slope
LOQ=10SD/slope
or detection limit=3.3σ/s, quantification limit=10σ/s, where σ is the standard deviation of y-intercept of regression line, and s is the slope of the calibration curve.
Specificity:
The specificity of the proposed method was determined against blank and placebo applications. Here mobile phase was used as blank and excipients like starch, lactose, magnesium stearate were used as placebo.
RESULTS AND DISCUSSION:
Method validation
System Suitability Test
After setting the optimum conditions, system suitability parameters for the developed method were determined and compared with recommended limits. To determine the parameters, the study was performed with standard solution of 50 µg/ml concentration and the results were acquired from six injections. System suitability parameters of the method were demonstrated in Table 2. According to the results, all of the system suitability parameters were within the recommended limits and the method was found to be suitable for the analysis.
Table 2: Results of system suitability test (n = 6)
|
Parameter |
Criteria |
Result |
|
Capacity factor(k’) |
k’> 2 |
3.824 |
|
Tailing factor (T) |
T < 2 |
1.156 |
|
Theoretical plates (N) |
N> 2000 |
4320 |
|
% RSD (peak area) |
% RSD ≤ 1 |
0.74 |
Stability:
The sample solution stability was analyzed by injecting the same solution at 0, 12, 24, and 48 h. Identical change was not observed in the developed method. Also, results were found within acceptable limits (% RSD < 2), which are summarized in Table 3.
Table 3: Stability data of Istradefylline (standard solutions)
|
Time (hr) |
Assay (%) |
% Difference |
|
Initial |
100.08 |
---- |
|
After 12 hr |
100.02 |
0.05 |
|
After 24 hr |
99.87 |
0.21 |
|
After 36 hr |
99.16 |
0.92 |
|
After 48 hr |
98.32 |
1.76 |
Linearity and sensitivity:
Linearity study was performed with calibration standards with 10, 20, 30, 40, 50,60,70, 80 and 90 µg/ml concentrations. The standards were injected in triplicate. Calibration curves were obtained by plotting the peak areas against the given concentrations. The calibration curve was evaluated by the determination coefficient. The determination coefficient (R2) of the calibration curves was 0.9993. Therefore, the calibration curve for istradefylline was found to be linear within the range of 10–90 µg/ml concentrations as shown in Fig.3. The regression equations were calculated from the calibration graphs. The sensitivity of the analytical method was evaluated by determining the limits of detection (LOD) and quantitation (LOQ). The values of LOD and LOQ are given in Table 4. The low values of LOD and LOQ indicates the sensitivity of method.
Table 4: Spectral and statistical data for determination of Istradefyllin by proposed RP-HPLC method.
|
Parameter |
Result |
|
Detection wavelength (nm) |
246 |
|
Linearity range (µg/ml) |
10-90 |
|
Coefficient of determination (r2) |
0.9993 |
|
Regression equation (Ya) |
Y= 123586x-29037 |
|
Slope (m) |
123586 |
|
Intercept (c) |
-29037 |
|
Limit of detection, LOD (µg/ml) |
0.02 |
|
Limit of quantitation, LOQ (µg/ml) |
0.06 |
aY = mx + c, where x is the concentration (µg/ml).
Fig. 3: Calibration curve for Istradeylline
Accuracy:
To study the reliability, the suitability, and the accuracy of the method, recovery experiments were carried out. Known quantities of the pure drug were added to the preanalyzed sample to make samples at the levels of 80 %, 100 %, and 120 %, and were assayed by the proposed method. Accuracy was calculated as the percentage of recovery. The recovery and relative standard deviation for each of the analytes are given Table 5. From the recovery studies it is evidence that the method is highly accurate and can give excellent results.
Table 5: Accuracy results
|
% Level |
Concentration(µg/mL) |
Recovery (%) |
Statistical Results |
|||
|
Formulation |
Pure drug |
Mean |
SD |
%RSD |
||
|
80 80 80 |
50 50 50 |
40 40 40 |
98.4 98.1 100.8 |
99.1 |
0.87 |
0.88 |
|
100 100 100 |
50 50 50 |
50 50 50 |
100.5 101.9 98.8 |
101.3 |
0.12 |
0.10 |
|
120 120 120 |
50 50 50 |
60 60 60 |
99.4 99.7 100.4 |
99.8 |
0.51 |
0.52 |
Table 6: Precision results
|
Precision |
Results |
||
|
Concentration (µg/mL) |
% RSD of Peak area |
% RSD of Retention Time |
|
|
Repeatability |
40 50 60 |
0.89 1.21 1.11 |
0.02 0.08 0.12 |
|
Intermediate precision |
40 50 60 |
1.42 0.75 0.67 |
0.08 0.06 0.06 |
|
Reproducibility |
40 50 60 |
1.64 0.78 0.85 |
0.11 0.17 0.09 |
Precision:
The precision was demonstrated at three levels: repeatability, intermediate precision, and reproducibility (between laboratories’ precision). Each level of precision was investigated by 3 sequential replicate of injections of three concentrations of 40, 50 and 60 µg/mL. The precision was expressed as relative standard deviation (RSD) or coefficient of variation (CV). The results of three levels of precision are shown in Table 6. The developed method was found to be precise as the RSD values for repeatability, intermediate precision and reproducibility studies were <2 %, respectively as recommended by ICH guidelines (ICH Q2 (R1), 2005).
Robustness and ruggedness:
Robustness of the method was studied by deliberate variations of the analytical parameters such as flow rate (1.0±0.1 mL/min), mobile phase composition (± 5 % organic phase) and analytical wavelength (±2 nm). The results are given in Tables 7. The result shown that have the negligible effect on retention time, recoveries and peak area of istradefylline indicating the developed method is robust. Ruggedness of the method was carried out by different analysts. The results are displayed in Table 8. There is no variation in peak areas and retention time of istradefylline from studies carried out by two analysts as indicated by % RSD < 2 gives the method rugged.
Mobile phase stability:
The stability of the mobile phase was evaluated, so the mobile phase was stored at 4–8 °C for 1 week. The aged mobile phase was compared using a freshly prepared one. The mobile phase was stable up to 1 week at 4–8 °C.
Specificity:
Specificity is the ability to unequivocally assess the analyte in the presence of components that may be expected to be present. Typically, these might include impurities, degradants or matrix. Specificity of an analytical method is its ability to accurately and specifically measure the analyte of interest without interference from blank or placebo. The peak purity of Istradefyllin was assessed by comparing the retention times of standard istradefylline and the sample, and good correlation was obtained between the retention time of the standard and sample. Placebo and blank were injected and there were no peaks. There is no interference of blank and placebo on drug peaks hence, the method is specific. The chromatogram for placebo and blank was shown in Fig 4 and 5 respectively.
Table 7: Robustness studies
|
Parameter |
Variation |
Observed value |
|||
|
% RSD of area |
% RSD of R.T |
Tailing factor |
Theoretical plates (N) |
||
|
Flow rate
|
0.9 1.1 |
0.47 0.65 |
0.09 0.07 |
1.14 1.15 |
4112 4225 |
|
M.Phase Composition |
75% acetonitrile 65 % acetonitrile |
0.79 0.81 |
0.04 0.13 |
1.14 1.14 |
4187 4302 |
|
Wavelength
|
248 nm 244 nm |
0.66 0.92 |
0.07 0.02 |
1.13 1.14 |
4221 4107 |
Table 8: Ruggedness studies
|
Analyst |
Observed value |
|||
|
% RSD of area |
% RSD of R.T |
Tailing factor(T) |
Theoretical plates(N) |
|
|
Analyst I |
0.45 |
0.07 |
1.15 |
4286 |
|
Analyst II |
0.52 |
0.06 |
1.15 |
4311 |
Fig .4: Chromatogram of placebo solution
Fig. 5: Chromatogram of blank solution
Sample Analysis:
The developed and validated method was applied for analysis of tablet formulation contain istradefylline. The sample was analyzed in triplicate. Analysis results were evaluated using a calibration curve. The amount of istradefylline in the samples was calculated from calibration curve equation and recovery and RSD values were determined. The results of analysis are given in Table 9. The recoveries were in good agreement with the label claims. The chromatogram obtained were clear as shown in Fig. 6. It was concluded that the method can be applied successfully for the analysis of istradefylline in tablet dosage form.
Fig. 6: Chromatogram of sample solution of istradefyllin
Table 9: Assay results from commercial formulation
|
Sample |
Labelled amount (mg) |
Amount obtained* (mg) |
Percentage Recovery*±SD |
|
NOURIANZ™ tablets (Istradefylline) |
20 |
19.92 |
99.62 ± 0.92 |
*Average of five determinations
CONCLUSION:
The proposed method for the estimation of Isrtradefyllin was validated as per the ICH guidelines and it is simple, specific and reliable. Furthermore, this simple and rapid RP-HPLC method can also be used successfully for the determination of istradefylline in pharmaceutical formulations without any interference from the excipient.
REFERENCES:
1. Fernandez HH et al. Istradefylline as monotherapy for Parkinson disease: results of the 6002-US-051 trial. Parkinsonism related Disorders.16(1); 2010:16–20.
2. Nagayama H et al. Effect of istradefylline on mood disorders in Parkinson's disease. Journal of Neurological. Sciences.396; 2019:78 —83.
3. Y. Mizuno, T. Kondo, Adenosine A2A receptor antagonist istradefylline reduces daily OFF time in Parkinson's disease Movement Disorders. 28(8);2013: 1138–1141.
4. Wang L et al. A LC-MS/MS method for simultaneous quantification of levodopa and istradefylline for parkinson’s disease and application to a pharmacokinetic study. Latin American Journal of Pharmacy.36;2017:1081-1087.
5. Xu H et al. Separation and identification of an impurity from the istradefylline intermediate. RCS Advances.10; 2020:14493: 14499. Xu H et al. Investigation of Photostability of Istradefylline Aqueous Solution. Chemistry Select. 5;2020:2337-2341.
6. Leela MP, Hemant TK, Srinivasa YR and Rao KVP. UV Spectrophotometric Method for Estimation of Sacubitril in Synthetic Mixture. Asian Journal of Research in Chemistry. 12(1); 2019: 07-10.
7. Bhavana V, Hemant TK, Srinivasa YR, Rao KVP. RP-HPLC method for estimation of Solifenacin Succinate in API and Tablet dosage form. Asian Journal of Pharmaceutical Analysis. 9(3); 2019:118-122.
8. Mounika PS, Hemant TK, Srinivasa YR, Rao KVP. RP-HPLC Method for Quantification of Empagliflozin In Pharmaceutical Formulation. Asian Journal of Pharmacy and Technology. 9(3); 2019:208-211.
9. Dhalape VM, Santosh TK and Pinjari VR. RP-HPLC Method Validation for Quantitative Analysis of Pemetrexed Disodium Hemipentahydrate. Research Journal of Pharmacy and Technology. 13(8); 2020:3685-3689.
10. Sabarathinam S and Vijayakumar TM. RP-HPLC Method Development and Validation of Midazolam: A CYP3A4 Probe Drug. Research Journal of Pharmacy and Technology.13(8); 2020: 3565-3568.
11. Vitthal MD, Santosh T and Rahul VP. RP-HPLC Method Validation for Quantitative Analysis of Pemetrexed Disodium Hemipentahydrate. Research Journal of Pharmacy and Technology, 13(8); 2020:3685-3689.
12. Deep Shikha Sharma, Sheetu Wadhwa, Sachin Kumar Singh, Arya Ramanunny, Rajan Kumar. Estimation of 5-fluorouracil by high-performance liquid Chromatography Reversed-phase Validated method. Research Journal of Pharmacy and Technology, 2020; 13(9):4249-4254.
13. Gupta S. K., Sachan N., Chandra P., Sharma A. K. A Robustustic Eco-Friendly Analytical Method to Quantify Topotecan in Pharmaceuticals by RP-HPLC. Research Journal of Pharmacy and Technology, 2020; 13(9):4387-4390.
14. Sarvesh Sabarathinam, T M Vijayakumar. RP-HPLC Method Development and Validation of Midazolam: A CYP3A4 Probe Drug. Research Journal of Pharmacy and Technology, 2020; 13(8):3565-3568.
15. Khairi M. S. Fahelelbom, Moawia M. M. Al-Tabakha, Nermin A. M. Eissa, Dana Emad Eddin Obaid, Sadik Sayed. Development and Validation of an RP-HPLC Analytical Method for Determination of Lisinopril in Full and Split Tablets. Research Journal of Pharmacy and Technology, 2020; 13(6):2647-2652.
16. Kallol Jana, Beduin Mahanti. Development and Validation of a new improved RP-HPLC Method for estimation of Rosuvastatin calcium in pharmaceutical dosage form. Research Journal of Pharmacy and Technology, 2020; 13(6): 2886-2892.
17. Survi Mishra, S. T. Narenderan, B. Babu, Kuntal Mukherjee, S. N. Meyyanathan. Validated Analytical Method for the Estimation of Gemcitabine from its Pharmaceutical Formulation by RP-HPLC. Research Journal of Pharmacy and Technology, 2019; 12(11): 5407-5412.
18. ICH-Guidelines Q2A, Validation of Analytical Procedures: Definition and Terminology (CPMP III/5626/94) March (1995). Geneva, Switzerland: ICH. ICH - Guidelines Q2B, Validation of Analytical Procedures: Methodology, (CPMP/ICH/281/95) November (1996). Geneva, Switzerland: ICH
Received on 31.08.2020 Modified on 28.09.2020
Accepted on 19.10.2020 © RJPT All right reserved
Research J. Pharm. and Tech. 2021; 14(9):4767-4772.
DOI: 10.52711/0974-360X.2021.00829