Stability Indicating RP-HPLC Method Development and Validation for Estimation of Diroximel Fumarate in bulk and its dosage forms
Ramachandrapuram Kiranjyothi1*, Mahalingam Balakrishnan2, Kothapalli Bannoth Chandrasekhar3
1Department of Pharmaceutical Analysis and Quality Assurance, Research Scholar,
Jawaharlal Nehru Technological University, Ananthapuramu, Andhra Pradesh., India.
2Department of Pharmacognosy, Sri Seshachala College of Pharmacy, Puttur Tirupati, Andhra Pradesh India.
3Vice Chancellor, Krishna University, Machillipatnam. Andhra Pradesh., India.
*Corresponding Author E-mail: jyothikiran460@gmail.com
ABSTRACT:
Aim and objective: The aim of present research work is to develop validated RP-HPLC stability indicating method for the quantification of Diroximel Fumarate in bulk and its pharmaceutical dosage forms. Materials and methods: Chromatographic method was carried on C18 column (Waters 250mm x 4.6mm, 5m). Mobile phase was prepared by mixing water: Acetonitrile: 85% OPA: in the ratio of 70:30. The flow rate was 0.1 mL/min and the injection volume was 20μL. The absorbance maxima of Diroximel Fumarate was measured at 215nm. The retention time was found to be 2.390 min. Result: The method was proved to be specific and linear in the range of 50-150μg/mL with correlation coefficient of 0.999. The % RSD for precision was found to be less than 2% and the mean percentage recovery was 100.15%. All the validation parameters were statistically validated according to ICH guidelines and were found to be within acceptance criteria. Conclusion: The developed method was simple, specific, precise, accurate and robust. The described HPLC method can be successfully employed for the analysis of Diroximel Fumarate.
KEYWORDS: Diroximel Fumarate, RP-HPLC method, 85% OPA, Acetonitrile, Validation, ICH guidelines.
INTRODUCTION:
Diroximelfumarate is a new drug from the fumarate class formulated to treat various relapsing forms of Multiple sclerosis. Diroximelfumarate is indicated for the treatment of relapsing forms of multiple sclerosis (MS) in adults; specifically active secondary progressive disease and clinically isolated syndrome, as well as relapsing-remitting MS. Diroximelfumarate relieves the neurological symptoms of relapsing MS with less gastrointestinal effects than its bioequivalent counterpart, dimethyl fumarate. It is important to note that diroximel fumarate can cause angioedema, anaphylaxis, hepatotoxicity, flushing, lymphopenia, and Progressive Multifocal Leukoencephalopathy (PML).
Fig 1. Structure of Diroximel Fumarate
Literature survey revealed that only a few analytical methods were reported for the estimation Diroximel Fumarate. Bio analytical methods (LC-MS/MS) have been reported for the quantification of Diroximel Fumarate in biological fluids.9 There is no RP-HPLC method for the analysis of Diroximel Fumarate.
MATERIALS AND METHODS:
Materials:
Diroximel Fumarate pure drug was obtained as gift sample from Biophore pharma, Hyderabad. Diroximel Fumarate Capsules (VUMERITY) were purchased from local market. HPLC grade acetonitrile, and distilled water were purchased from Merck, Mumbai. Analytical grade phosphate buffer, ortho-phosphoric acid were obtained from Rankem laboratories, Hyderabad.
Instrument:
The liquid chromatographic system was performed on waters HPLC 2965 system equipped with auto injector and PDA Detector. Empower software was used for data acquisition, processing and reporting.
Chromatographic conditions:
The method development was achieved on Waters C18 column (250 x 4.6mm, 5µm). Mobile phase was freshly prepared by mixing water (85% OPA) and acetonitrile in the ratio of 70:30 v/v. The flow rate was maintained at 1.0mL/min. Detector wavelength was monitored at 215 nm, and the injection volume was 20μL and run time was kept 5 min.
Preparation of standard stock solution:
Accurately weighed 10mg of Diroximel Fumarate was dissolved in 10ml mobile phase (1000µg/ml of Diroximel Fumarate standard stock solution). From the above standard stock solution 0.125ml was transferred into a 10ml volumetric flask and made up to the mark with mobile phase. (125µg/ml of Diroximel Fumarate).
Preparation of Sample stock solutions:
Average weighed amount equivalent to each tablet was calculated. Weight equivalent to 18.49mg was transferred into a 100ml volumetric flask and made up to the mark with mobile phase (125µg/ml of Diroximel Fumarate). The solution was sonicated for 25 min and filtered by HPLC filters. Further pipette out 0.125ml of filtered sample stock solution was transferred to 10ml volumetric flask and made up with mobile phase. (125µg/ml of Diroximel Fumarate).
Method validation:
System suitability parameters:
System suitability was evaluated in terms peak tailing and USP plate count by injecting 6 replicates of 10 μg/mL Diroximel Fumarate standard drug concentration. The calculated % RSD for the area of six standard injections results should not be more than 2%.
Specificity:
The specificity of developed method was evaluated by injecting blank sample (mobile phase) to demonstrate the absence of interference with elution of Diroximel Fumarate standard solution. (125μg/mL).
Linearity:
Six different concentrations ranging from 62.5-187.5 μg/mL of Diroximel Fumarate standard stock solution were prepared and injected. Calibration curve was constructed by plotting mean response factor against the respective concentration. The method was evaluated by determination of the correlation coefficient and intercept value.
Accuracy:
Recovery assessment was obtained by using standard addition technique which was performed by adding known quantities of pure standards at three different levels in 50%, 100% and 150% to the pre analysed sample formulation. From the amount of drug found and amount of drug recovered, percentage recovery was calculated.
Precision:
Diroximel Fumarate standard concentrations (125 μg/mL) were analyzed for 6 times in consecutive days (inter day precision) and 6 times during the same day (intra-day precision). The precision of proposed method was obtained by calculating the relative standard deviation (RSD) values for intra-day and inter-day analysis with acceptance criteria of ≤ 2% RSD.
Robustness:
Robustness of the method was evaluated by small but deliberate changes in method like Flow rate at ±0.2 ml/min and temperature by ±5°C.
Assay:
Assay of the marketed formulation was carried out by injecting sample corresponding to equivalent weight into HPLC system to calculate the percentage purity.
Forced Degradation studies:
Degradation studies is carried out on the sample by using Acid, Alkaline, Oxidative, Thermal studies, Photolytic and Thermal degradations. The sample was stressed to these conditions and peak was evaluated for peak purity indicating the above method can be used successfully separate the degradation products from pure active ingredients.
Degradation in acidic condition:
Degradation in acidic medium is carried out by weighing 125 micro grams of Diroximel Fumarate in 10ml Volumetric flask. and the contents was dissolved in 3ml of 0.5N HCL and heated on a water bath at 85°C Neutralise the sample with 0.5N NaOH and injected into HPLC system at different time intervals of 0, 2, 4, 6, 10, 12, 24 hrs.
Degradation in Basic condition:
Degradation in Alkaline medium was Performed by 0.5N NaOH by weighing 125 micrograms in 10ml volumetric flask.it was dissolved in 3ml of 0.5N NaOH and it was kept aside. Neutralise the sample with 0.5N HCL and injected into HPLC system at different time intervals of 0, 2, 4, 6, 10, 12, 24 hrs.
Oxidative degradation:
Oxidative Degradation is performed by Weighing 125 micrograms of Diroximel Fumarate which was transferred into volumetric flask. The contents are dissolved in 5ml of 3% H2O2 and kept aside for 1 day and injected into HPLC System at different time intervals of 0, 2, 4, 6, 10, 12, 24 hrs.
UV-degradation:
UV Degradation is performed by weighing 125 micrograms of Diroximel fumarate accurately and kept in UV chamber at 275nm and injected into HPLC system at different time intervals of 0, 2, 4, 6, 10, 12, 24 hrs.
RESULTS AND DISCUSSION:
System Suitability:
Standard solutions of Diroximel Fumarate working standard was prepared as per procedure and were injected six times into the HPLC system. The system suitability parameters were evaluated from standard Chromatograms obtained by calculating the %RSD of retention time, tailing factor, theoretical plates and peak areas from five replicate injections are within range and results were shown in table 1.
Table No. 1 System suitability parameters
|
S. No |
STD Area |
Theoretical Plates |
Tailing Factor |
RT |
|
1 |
10709826 |
3815 |
1.77 |
2.390 |
|
2 |
10516939 |
4246 |
1.77 |
2.389 |
|
3 |
10556998 |
4094 |
1.76 |
2.390 |
|
4 |
10622063 |
3887 |
1.75 |
2.389 |
|
5 |
10883852 |
3947 |
1.71 |
2.389 |
|
6 |
10333888 |
4008 |
1.70 |
2.391 |
|
Avg |
10603928 |
|
1.74 |
2.39 |
|
SD |
185791 |
|
0.03 |
0.001 |
|
%RSD |
1.752 |
|
1.765 |
0.034 |
Fig.2 System suitability Chromatogram
Linearity:
To demonstrate the linearity of assay method, 5 standard solutions with concentrations of about 50μg/mL to 150 μg/mL of Diroximel Fumarate were injected. Calibration curve plotted between peak area versus respective concentration which was shown in figure 3 Slope obtained was 835393 Y-Intercept was 60000 and Correlation Co-efficient was found to be 0.999.
Table No.2 Results of Linearity
|
Linearity Level (%) |
Concentration (μg/mL) |
Area |
|
50 |
62.5 |
5281217 |
|
75 |
93.75 |
7921825 |
|
100 |
125 |
10562433 |
|
125 |
156.25 |
12903041 |
|
150 |
187.5 |
15843650 |
Fig .3 Calibration curve of Diroximel Fumarate
Accuracy:
Diroximel Fumarate spiked standard concentrations (50 μg/mL, 100μg/mL, 150μg/mL) at all the three levels were analyzed for percentage recoveries and the results were presented in table 3. The mean percentage recoveries of three levels (3 samples from each concentration were injected) was found to be 100.08%. The accuracy results were within the accepted limits from 98.0% to 102.0% which proves that the method was found to be accurate. Good recovery results obtained for the developed method indicates that this method can be used for regular quality control assay test for Diroximel Fumarate.
Precision:
The precision of a method determines the closeness of agreement between a series of measurements of the same sample. The intraday and interday precisions were carried out 6 times at concentration of 10μg/mL and the %RSD were found to be 0.7 and 0.7, respectively. The precision result (table 4) was within the accepted limits of ≤ 2 % RSD which proves that the method was precise.
LOD and LOQ:
LOD is a limit test parameter and it is a test to determine whether the analyte concentration was present within the specification limit or not. LOQ is a parameter for quantitative assay used particularly for determination of impurities or degradation products as it used for minimum concentrations of analytes in sample. The LOD and LOQ were found to be 6.67 and 22.24 respectively which proves the method was sensitive.
Table No.3 Accuracy data
|
Sample No. |
Spiked Level |
Sample Weight (mg) |
Sample Area |
µg/ml added |
µg/ml found |
% Recovery |
% Mean Recovery |
|
1 |
50 |
9.245 |
5301055 |
6.23 |
6.25 |
100.30 |
100.15 |
|
2 |
50 |
9.245 |
5288206 |
6.23 |
6.23 |
100.05 |
|
|
3 |
50 |
9.245 |
5182224 |
6.23 |
6.11 |
98.05 |
|
|
4 |
50 |
9.245 |
5343124 |
6.23 |
6.30 |
101.09 |
|
|
5 |
50 |
9.245 |
5272227 |
6.23 |
6.21 |
99.75 |
|
|
6 |
50 |
9.245 |
5372227 |
6.23 |
6.33 |
101.64 |
|
|
7 |
100 |
18.49 |
10608765 |
12.46 |
12.51 |
100.36 |
99.82 |
|
8 |
100 |
18.49 |
10556161 |
12.46 |
12.44 |
99.86 |
|
|
9 |
100 |
18.49 |
10490284 |
12.46 |
12.37 |
99.24 |
|
|
10 |
150 |
27.735 |
15950992 |
18.69 |
18.80 |
100.60 |
99.94 |
|
11 |
150 |
27.735 |
16073603 |
18.69 |
18.95 |
101.37 |
|
|
12 |
150 |
27.735 |
15855434 |
18.69 |
18.69 |
100.00 |
|
|
13 |
150 |
27.735 |
15801278 |
18.69 |
18.63 |
99.65 |
|
|
14 |
150 |
27.735 |
15853267 |
18.69 |
18.69 |
99.98 |
|
|
15 |
150 |
27.735 |
15542577 |
18.69 |
18.32 |
98.02 |
Table No.4 Precision da
Intraday precision
|
S. No |
Sample Weight |
Sample Area |
% Assay |
|
1 |
18.49 |
10537857 |
99.39 |
|
2 |
18.49 |
10492882 |
98.96 |
|
3 |
18.49 |
10494555 |
98.98 |
|
4 |
18.49 |
10676745 |
100.70 |
|
5 |
18.49 |
10648596 |
100.43 |
|
6 |
18.49 |
10479237 |
98.83 |
|
Avarage Assay: |
99.55 |
||
|
STD |
0.81 |
||
|
% RSD |
0.82 |
||
Interday precision
|
S. No |
Sample Weight |
Sample Area |
% Assay |
|
1 |
18.49 |
10561216 |
99.61 |
|
2 |
18.49 |
10581125 |
99.80 |
|
3 |
18.49 |
10569306 |
99.68 |
|
4 |
18.49 |
10682469 |
100.75 |
|
5 |
18.49 |
10672469 |
100.66 |
|
6 |
18.49 |
10452469 |
98.58 |
|
Avarage Assay: |
99.85 |
||
|
STD |
0.80 |
||
|
% RSD |
0.80 |
||
Robustness:
Robustness of the method was performed by changing flow rate (± 0.2mL/min) and change in temperature (± 50 C). The results were summarized in table 5. It was observed that even in minor changes of method conditions there was no marked changes in the results demonstrate that the HPLC method developed was robust. The robustness results were within the accepted limits of ≤ 2 % RSD.
Table No.5 Robustness Data
|
S. No |
Condition |
Peak area |
% Assay |
|
1 |
0.8 ml/min |
10630112 |
100.26 |
|
2 |
1 ml/min |
10591131 |
99.89 |
|
3 |
1.2 ml/min |
10652150 |
100.47 |
|
4 |
25 °C |
10679545 |
100.72 |
|
5 |
30°C |
10531211 |
99.33 |
|
6 |
35 °C |
10582877 |
99.81 |
Assay of Marketed Formulation:
Sample solution was injected separately 6 times from the same sample individually into the system and chromatograms were recorded and %RSD was reported from the calculated percentage purity values.
Table No.6 Assay of Formulation
|
Sample No |
%Assay |
|
1 |
100.68 |
|
2 |
100.40 |
|
3. |
99.63 |
|
4. |
99.08 |
|
5. |
98.94 |
|
6. |
99.95 |
|
AVG |
99.78 |
|
StdDev |
0.70 |
|
%RSD |
0.7 |
Forced Degradation studies
Acid Degradation studies
Table No 7:Percentage of various degradation studies
|
S. No |
Nature of the Sample |
Sample Weight |
Sample Area |
% Assay |
% of Degradation |
|
1 |
Acid |
18.49 |
9524683 |
89.83 |
10.17 |
|
2 |
Base |
18.49 |
9629742 |
90.82 |
9.18 |
|
3 |
Peroxide |
18.49 |
9734801 |
91.81 |
8.19 |
|
4 |
Heat |
18.49 |
9555185 |
90.12 |
9.88 |
|
5 |
UV |
18.49 |
9661216 |
91.12 |
8.88 |
CONCLUSION:
The developed method was validated as per ICH guidelines. All the validation parameters were found to be well within the acceptance criteria. We concluded that the method is accurate, precise, linear and robust. The developed method can be successfully applied for the analysis of Diroximel fumarate in bulk and pharmaceutical dosage form in quality control laboratories.
CONFLICT OF INTEREST:
None.
REFERENCES:
1. Fox, EJ, Vasquez, A, Grainger, W, . Gastrointestinal tolerability of delayed-release dimethyl fumarate in a multicenter, open-label study of patients with relapsing forms of multiple sclerosis. Int J MS Care 2016; 18(1): 9–18
2. K, Spinelli, KJ, Stuchiner, T, . Three-year clinical outcomes of relapsing multiple sclerosis patients treated with dimethyl fumarate in a United States community health center. Mult Scler 2018; 24(7): 942–950.
3. Gold, R, Kappos, L, Arnold, DL, . Placebo-controlled phase 3 study of oral BG-12 for relapsing multiple sclerosis. N Engl J Med 2012; 367: 1098–1107.
4. Gold, R, Arnold, DL, Bar-Or, A, . Long-term effects of delayed-release dimethyl fumarate in multiple sclerosis: Interim analysis of ENDORSE, a randomized extension study. Mult Scler 2017; 23(2): 253–265.
5. Polman, CH, Reingold, SC, Banwell, B, . Diagnostic criteria for multiple sclerosis: 2010 revisions to the McDonald criteria. Ann Neurol 2011; 69(2): 292–302.
6. Giles, K, Hanna, J, Wu, F, . Efficacy of delayed-release dimethyl fumarate in newly diagnosed and other early MS patients, and patients switching from interferon or glatiramer acetate, in routine medical practice: Interim results from Esteem. Mult Scler J 2017; 23(Suppl 3): 612.
7. Mehta, D, Miller, C, Arnold, DL, . Effects of dimethyl fumarate on lymphocytes in RRMS: Implications for clinical practice. Neurology 2019; 92(15): e1724–e1738.
8. Giles, K, Hanna, J, Wu, F, . Efficacy of delayed-release dimethyl fumarate in newly diagnosed and other early MS patients, and patients switching from interferon or glatiramer acetate, in routine medical practice: Interim results from Esteem. Mult Scler J 2017; 23(Suppl 3): 612.
9. Mehta, D, Miller, C, Arnold, DL, . Effects of dimethyl fumarate on lymphocytes in RRMS: Implications for clinical practice. Neurology 2019; 92(15): e1724–e1738.
10. Mirabella, M, Prosperini, L, Lucchini, M, . Safety and efficacy of dimethyl fumarate in multiple sclerosis: An Italian, multicenter, real-world study. CNS Drugs 2018; 32(10): 963–970.
11. International Conference on Harmonization of Technical Requirement for Registration of Pharmaceuticals for Human use. ICH harmonized tripartite guideline. Validation of analytical procedures text and methodology Q2 (R1).
12. ICH Harmonized Tripartite Guidelines Validation Of Analytical Procedures : Methodology, (1996), pages1-8
Received on 14.01.2020 Modified on 04.05.2020
Accepted on 16.07.2020 © RJPT All right reserved
Research J. Pharm. and Tech. 2021; 14(5):2603-2607.
DOI: 10.52711/0974-360X.2021.00458