PCCA Algorithm as a fingerprint resolution technique for the analysis of Ciprofloxacin in the presence of its acid induced degradation product

 

Amir Alhaj Sakur*, Reem Hasan Obaydo

Department of Analytical and Food Chemistry, Faculty of Pharmacy, Aleppo University, Syria.

*Corresponding Author E-mail: profsakur@gmail.com

 

ABSTRACT:

Pure component contribution algorithm (PCCA) is an accurate analytical method was developed for extracting the contribution of each component from a mixture where the components of this mixture are completely overlapped. It is based on elimination the signal of interfering components using mean centering as a processing tool finally, the pure contribution of each component is extracted form complex signals and that worked as a finger-print resolution for the component. (PCCA) has been successfully applied to UV data of ciprofloxacin hydrochloride (CIP) in the presence of its acidic degradation product (DCIP). The spectrophotometric technique (PCCA) was successfully applied to the laboratory prepared mixtures and the pharmaceutical dosage form where the purity of the extracted signals were tested by calculating the spectral contrast angle (Ɵ) where the results were compared to show the capability to recover pure spectral profiles and detect the presence of impurities. The proposed method proved that spectrophotometric techniques can be used for identification and separation of signals, similar to chromatographic techniques. The method validated according to the ICH guidelines and the accuracy, precision and repeatability found to be within the acceptable limits. The selectivity of the proposed method was tested using laboratory prepared mixtures and assessed by applying the standard addition technique. So, they can be used for the routine analysis of ciprofloxacin in quality-control laboratories.

 

KEYWORDS: Ciprofloxacin, Stability-indicating, Pure component contribution algorithm, Acidic degradation, Spectral contrast angel, Finger-print.

 

 


INTRODUCTION:

Ciprofloxacin (CIP), [1-cyclopropyl-6-fluoro-1, 4-dihydro-4-oxo-7-(piperazinyl)-quinolone-3-carboxylic acid] Fig. 1 is a widely used second generation fluoroquinolone, useful for the treatment of a number of bacterial infections. It has been used in the treatment of infections including anthrax, biliary-tract, bone, joint, ear, nose and throat infections. It is also used for urinary-tract infections including chronic bacterial prostatitis [1].

 

CIP is an official drug presented in the British Pharmacopoeia [2]. Literature review encountered that several analytical methods were published for the quantitative analysis of CIP in pharmaceutical drugs and in physiological fluids [3]–[7][8][9][9][10].

 

Some methods include few simple and direct UV-spectrophotometric methods for the resolution of Ciprofloxacin hydrochloride in existence of its degradation products through the application of Ratio derivative, Ratio difference, Mean centering, Dual wavelength methods [11], high performance liquid chromatographic methods [5], [12], and TLC method [13]. Stability is considered to be one of the most important criteria in quality control related to pharmaceuticals. Only stable pharmaceutical drugs would guarantee accurate delivery of the drug to the patients. Expiration dating on any drug formulation depends on scientific studies at ordinary and stressed conditions. The International Conference on Harmonization (ICH) [14] approved a stability test which suggests that it is important to investigate the inherent stability properties of the drug product through certain stress studies, for instance, hydrolysis study; therefore, this will lead to the identification and determination of drug products which will support the developed stability-indicating analytical method.

Acid degradation of Ciprofloxacin hydrochloride (CIP) is considered an easy technique as it is a chloro analogue drug. It was reported [12] that it was: 6-chloro-1-cyclopropyl-4-oxo-7-piperazin-1-yl-1,4-dihydroquinoline-3-carboxylic acid. However, the acid degradation products used in this work were prepared using the acid hydrolysis as mentioned in [12] where its condition was 50.0mL of0.1 N methanolic hydrochloric acid added to 25mg of CIP and the resulting solutions were kept for 72 h, a stock solution labeled to contain degradation derived from 0.5mg/mL of CIP.


 

 

Figure 1: the chemical structure of CIP and DCIP

 


The previously mentioned conditions were enough for complete acid degradation. The ultimate target of this study is to achieve and validate stability-indicating UV-spectrophotometric method that is simple, rapid, and selective with less cost and time for the analysis of ciprofloxacin hydrochloride in existence of its acid degradation product in its authentic form and in market samples excluding any separation steps by using Matlab® manipulating program to obtain the spectral profile of CIP so the purity of this resolved spectra could be checked by Spectral contrast angel (Ɵ), this step is very important especially in the case if the UV spectrum of the degrade is very similar to the pure spectrum [15] for the same component in this case finger print resolution technique would be the best choice for the analyzer to avoid any analytical mistake maybe obtained.

 

The aim of the presented study was to developa new finger-print UV technique as a stability-indicating method providing assay of the Active Pharmaceutical Ingredient (API) together with tracking of the acid impurity already present or as a result of forced degradation studies in film-coated tablets of ciprofloxacin HCl.

 

Experimental:

Apparatus and software:

JASCO V-650 dual beam UV-VIS spectrophotometer, Quartz cuvettes were used in measurement. PCCA Algorithm was done with our own written code in Matlab®, The t-test and F-test were performed using Microsoft Excel 2016.

 

Standard Solutions:

Preparation of CIP Standard Solution:

·       CIP standard stock solutions; 100μg/mL in 0.1 N methanol.

·       CIP standard working solutions; 50μg/mL in 0.1 N methanol.

Preparation of the Degradation Product (DCIP)

 

25 mg of pure ciprofloxacin hydrochloride powder was transferred to a 50-mL flask and the volume was completed with 0.1 N methanolic hydrochloric acid. The solution was kept in a dark place under lab-temperature for 72 h[12]. After that, the solution was neutralized with 0.1N sodium hydroxide and a stock solution labeled to contain degradate derived from 0.5 mg/mL of CIP. Working solution of degradate (50 μg/mL) was obtained by further dilution of the stock solution with the solvent (methanol). Aliquots of different concentrations of ciprofloxacin degradation product (DCIP) were accurately transferred into series of 10-mL volumetric flasks and the volumes were completed to the mark with methanol. These solutions were scanned in range 200–400 nm and stored in the Computer.

 

Materials and Reagents:

·       Zhejiang Langhua Pharmaceutical Company, Linhai, China, provided Ciprofloxacin hydrochloride (CIP) with a purity proportion of 99.95±0.40 according to official BP criteria [2].

·       Pharmaceuticals tablets of CEPROZ®, Batch No. 671 CC, marked to contain 500 mg of (CIP) per tablet, produced by (ElSaad pharma, Syria)

·       Concentrated hydrochloric acid solution (Merck) used to prepare 0.1 N HCl.

·       Sodium hydroxide pellets (Merck) used to prepare 0.1 N NaOH.

·       De-ionized water.

·       Methanol (Panreac, Spain).

·       Methanolic acid 0.1 N.

 

Theoretical background:

PCCA Algorithm:

In order to explain the PCCA technique, an expression of the mean centering technique must be done first, let us consider a three-dimensional vector[16]:

 

 

We center or mean center (MC) this column by subtracting the mean of three numbers

 

 

It could be proved that if the vector y is multiplied by n (a constant number), the mean centered vector is also multiplied by n and also if a constant number is added to the vector Y, the mean center of this vector is not changed. Consider a mixture of tow compounds CIP, and DCIP. If Beer’s low is obeyed for each compound, it can be written:

 

 

where Am is the vector of the absorbance of the mixture, αX and αY are the molar absorptivity vectors of CIP and DCIP and CCIP and CDCIP are the concentrations of CIP and DCIP, respectively. If Eq. (1) is divided by αDCIP corresponding to the spectrum of a standard solution of DCIP in binary mixture, the first ratio spectrum is obtained in the form of Eq. (2) (for possibility of dividing operation, the zero values of αDCIP should not be used in the divisor)

 

 

If Eq. (2) is mean centered (MC), since the mean centering of a constant (CDCIP) is zero, Eq. (3) obtained:

 

 

By dividing Eq. (3) by MC () constant value corresponding to CCIP is obtained, Eq. (4)

 

By multiplying Eq. (4) by aCIP corresponding to the spectrum of a standard solution of , the absorption spectrum of  in the mixture is obtained as Eq. (5):

 

Eq. (5) is the mathematical foundation of each component in the laboratory prepared mixtures and pharmaceutical formulations[17]. This permits the determination of each component in the mixture (CIP in this equation) without interfering from the other components of the binary system (DCIP in these equations). As Eq. (5) shows, the obtained spectra permits the determination of component CIP by direct measurement of the estimated absorbance value at its λmax using the corresponding regression equation obtained by plotting the absorbance of the pure spectra of CIP at its λmax versus its corresponding concentration and that acts as a finger print profile for CIP component . [18]–[27].

 

Spectral contrast angle (Ɵ):

This is applied for accurate testing of similarity and purity of the resolved spectra of both X and Y present in the mixtures. Upon applying finger-print resolution to get the zero spectra D0 of X and Y separately, the absorbance values at the entire wavelength range are compared to those of the pure standards X and Y of the same concentration. This is done by calculating the spectral contrast angle (Ɵ) as follows [28]:

 

Where ai and bi are the absorbance values of the resolved and reference spectra of the same component at a given wavelength (i). The average value of cos Ɵ was calculated at the selected wavelength range Σi for X and Y. An angle Ɵ of zero between two vectors means that there is no distinguishable difference between the resolved and reference spectra of the pure drug. Accordingly, the value of cos Ɵ will be equal to one for the exact similar resolved and reference spectra which indicates 100% purity. The previously spectrophotometric PCCA Algorithm was successfully applied to the laboratory prepared mixtures and the combined dosage form where the purity of the extracted signals were tested by calculating the spectral contrast angle (Ɵ) where the results were compared to show the capability to recover pure spectral profiles and detect the presence of impurities. The proposed method proved that spectrophotometric techniques can used for identification and separation of signals, similar to chromatographic techniques. This work describes the application of the finger-print technique[29] for the quantification of the CIP and DCIP where the two compounds are simultaneously determined in laboratory prepared mixtures and in tablets, without prior separation, using absorptivity at their maxima and then the obtained results of the resolved spectra were compared with those obtained by the pure drugs. In addition, spectral contrast angle was calculated to estimate the similarity and purity of the resolved spectra.

 

Procedure:

Linearity and Construction of Calibration Curve of CIP:

Aliquots from CIP working standard solution accurately measured transferred into a set of 10-mL volumetric flasks and completedto volume with methanol to give (1.0–15.0μg/mL). The zero-order absorption spectrum of each solution was recorded versus methanol as a blank at λmax=278 nm, and plotted against the corresponding concentrations.

 

Application to Laboratory Prepared Mixtures:

Accurate aliquots of CIP and DCIP were transferred from their working solutions into a series of 10-mL volumetric flasks to prepare mixtures containing different ratios of both. The volumes were completed with methanol. The spectra of the prepared series from 200 to 400 nm were recorded and stored in PC. The concentrations of CIP were calculated by dividing the previously zero order absorption spectrum of each solution by the normalized spectrum of DCIP (1 μg/mL) used as a divisor concentration, (normalized spectrum is prepared mathematically by using sum of different spectra of DCIP and divided them by the total concentrations of them). The PCCA Algorithm to each ratio spectrum was done, in the wavelength range from (220.0 -340.0nm) in order to absence of zero absorptivity value for CIP, then the zero-order absorption spectrum of CIP at λ278 nm were obtained, and the corresponding concentrations were calculated.

 

Application to Pharmaceutical Preparation:

Ten tablets were weighed and finely powdered. Appropriate weight of powder equivalent to 20mg CIP was accurately transferred to 100-mL volumetric flask and the volume was made up to 75mL with methanol. The solution was shaken vigorously for 15 min then sonicated for 30 min. The volume was completed to 100 mL with solvent then filtered through Whatman filter paper no.41. Necessary dilutions of the filtrate were made with methanol to obtain different concentrations of CIP samples as stated under linearity. To assess the accuracy of the method, standard addition technique was applied.

 

Purity testing of recovered spectra:

For accurate purity testing of the resolved spectra of CIP present in the mixtures with DCIP, the absorbance values at the selected wavelength range were compared to that of the pure CIP standard of the same concentration. This was done by calculating the spectral contrast angle (Ɵ) as follows:

 

 

Where ai and bi are the absorbance values of the resolved and standard pure spectra of CIP at a given wavelength (i). The average value of cos Ɵ was calculated at the selected wavelength range Σi (220-350 nm for CIP). The resolved spectra of CIP by using the PCCA Algorithm method was compared with the pure stander spectra CIP in terms of purity and the values of cos Ɵ was calculated.

 

RESULTS AND DISCUSSION:

The zero-order absorption spectra of ciprofloxacin hydrochloride and its acidic degradate, as shown in Fig. 2, show severe overlap, which does not permit direct determination of ciprofloxacin hydrochloride in the presence of its degradation product. To overcome this interference PCCA method was done allowing the determination of CIP in the presence of its acidic degradation product. This method comprises two critical steps, first is the selection of the divisor and this has been discussed by Hegazy and she has proved that using normalized divisor was the best chosen in order to eliminate the effect of the divisor [24]. The second critical step is the choice of the wavelengths range does not contain zero absorptivity at which measurements recorded. Several wavelength ranges have been tested and the range from 220 to 340nm showed the best results. The ratio spectra obtained by normalized divisor of DCIP (the concentration of normalized DCIP´ is 1.0 μg/mL) these normalized divisor gave minimum noise in ratio spectra and maximum sensitivity. The ratio spectra were mean centered in the range (220–340nm) for CIP, then the PCCA technique followed until the D0 spectra of CIP were obtained, then the concentrations of CIP were calculated by using the regression equation representing the linear relationship between (A values) at 278nm and the corresponding concentrations.as shown in Fig. 3.

 

Figure 2: Zero order spectra of (10.0 μg/mL) CIP, (10.0 μg/mL) DCIP and a Mixture of (5 μg/mL) CIP and (5 μg/mL) DCIP

 

For testing the purity of the recovered spectra of CIP using the fingerprint PCCA technique, the spectral contrast angle (Ɵ) was calculated for the resolved D0 spectra of CIP. By applying the (PCCA), the D0 spectra of CIP were resolved from the synthetic mixtures, then, the purity of the resolved spectra was tested by measuring vector (a), which was represented by the absorbance signals of resolved spectra of CIP at (220-350 nm) at each nm with interval (0.1 nm), and vector (b) which was represented by the absorbance signals of reference pure CIP spectra of the same concentration in the same wavelength range and interval. The sum of (a) and (b) vectors, and their squares were used to calculate the cos Ɵ for each resolved spectrum in comparison to its corresponding reference one. The values of cos Ɵ for CIP spectra in each synthetic mixture were calculated and it was found that the calculated values approached unity which indicated high purity of the resolved peaks, as shown in Table 2. It was observed that the resolved peaks coincide with the standard peaks especially at the peak.


 

 

Figure 3 (a) The zero-order spectra of mixtures containing different ratios of CIP and DCIP, (b) MC of the ratio spectra, (c) Constants corresponding to the concentration of CIP in the mixtures and (d) The pure contribution of Recovered CIP in the prepared mixtures.

 


 

Figure 4: the cos(Ɵ) value calculating between the recovered zero-order of 5 μg/mL from CEPROZ® and pure 5 μg/mL of pure CIP

 

Validation of the Methods:

Validation of the proposed methods was assessed as per the ICH guidelines [14]

 

Linearity range:

Linearity of the methods examined by making six different calibration curves on three consecutive days. The calibration curves are set up within concentration ranges that were picked on the basis of the expected drug concentration throughout the assay of the dosage form. Each concentration repeated three times. All the validation parameters listed in Table 1.

 

Accuracy:

The accuracy of the presented methods examined by analyzing three samples of CIP standard solutions. The recovery percentages listed in Table 1, and the results showed high accuracy of the suggested methods.

 

Precision:

Repeatability and intermediate precision (intra-day and inter-day precision) are expressed as RSD and can be obtained by analyzing three different concentrations of the target drug in the linearity range on single day and on three following days as shown in Table 1.

 

Specificity:

Specificity checked by analyzing different mixtures consisting of the drug and its degradation products in several ratios within the linearity range. Good percentage recoveries with minimal standard deviation among the PCCA method obtained as presented in Table 2. The results of the analysis of the pharmaceutical formulations illustrated in Table 3. According to ICH, the accuracy of pharmaceutical formulation using the proposed method were checked in case of CEPROZ® 500mg tablets through standard addition technique, which is a recovery attained when spiking the sample solution of the pharmaceutical formulation with known concentration of the intact drug as illustrated in Table 3. A comparative study with a reported method [12]was done as illustrated in Table 4. As well as spectral contrast angel for the resolved spectra of CIP from CEPROZ® was obtained in table 4. Figure 4

 

System Suitability:

System Suitability was determined by replicating the measurement of standard preparation absorbance for six times. RSD percentage for those six results was calculated and it was less than 2%, which ascertain the precision of the used system as listed in table 1.

 

Table 1: Regression and validation parameters of ciprofloxacin (CIP) by applying PCCA method

Analyte

CIP

Wavelength (nm)

278

Spectral region

220-340

Intercept

0.004

Slop

0.1211

aRange (μg/mL)

1.0-15.0

Correlation coefficient

0.9999

bAccuracy ± SD

99.95 ± 0.54

cRepeatability (RSD %)

0.75

dIntermediate Precision (RSD %)

1.32

fSystem Suitability (RSD %)

1.19

LOD (μg/mL)

0.10

LOQ (μg/mL)

0.31

bAccuracy was checked using concentrations (6.0, 10.0, 14.0 μg/mL) for CIP.

cand dare the intraday and interday respectively relative standard deviation of three concentrations (4.0, 8.0, 12.0 μg/mL) of CIP, each concentration repeated for three times.

f Relative standard deviation of (6.0 μg/mL) of CIP, repeated for six times.

 


Table 2: Determination of ciprofloxacin (CIP) and (DCIP) in their laboratory prepared mixtures by applying PCCA method:

Mixture No

CIP (µg/mL)

DCIP (µg/mL)

Recovery (CIP)% Mean*

COS(Ɵ) of Recovered D0 (CIP)

1

5.00

3.00

100.40

0.9998

2

5.00

5.00

100.20

0.9998

3

5.00

7.00

100.60

0.9999

4

5.00

10.00

100.00

0.9999

5

5.00

12.00

99.60

0.9999

Mean ±SD

---

---

100.16± 0.38

0.9999 ± 0.0055

*Average of three experiments.

 


Table 3: Standard addition technique for CEPROZ® 500mg tablets by the proposed PCCA spectrophotometric method

Taken

(µg/mL)

Pure Add

(µg/mL)

Found

(µg/mL)

Recovery %

of Added

5.00

4.00

4.02

100.50

5.00

4.00

4.01

100.25

5.00

4.00

4.03

100.75

5.00

5.00

5.00

100.00

5.00

5.00

4.98

99.60

5.00

5.00

4.88

97.60

5.00

6.00

6.02

100.33

5.00

6.00

6.01

100.17

5.00

6.00

5.98

99.67

Mean ±SD

---

---

99.87±0.93

 

Statistical analysis:

The suggested method and the reported HPLC method for determination of CIP [12], are statistically compared with each other, regarding the t test and F values and revealed that there was no considerable difference between the reported HPLC method and the experimental values, which were obtained in the pure sample analysis by the proposed methods, indicating identical accuracy and precision where the results are illustrated in Table 4.

 

Table 3 Comparison between the results for determination of the*dosage form by the proposed PCCA method and reported HPLC [12] method:

Parameters

CIP

CIP

PCCA

**HPLC method

Mean

100.58

100.20

n

6

6

SD

1.04

1.15

Variance

1.0868

1.3218

F test (5.05)

1.22

---

t test (2.23)

0.61

---

#COS(Ɵ)

0.9999

---

The values between parentheses are the critical values of F and t at p equals 0.05.

*CEPROZ®, Batch No. 671 CC, marked to contain 500 mg of (CIP) per one tablet.

**HPLC reported method for CIP was done as mentioned in reference [12].

#Average of six recovered spectra.

 

CONCLUSION:

This work discussed the application of finger-print spectrophotometric resolution technique “Pure component contribution algorithm” (PCCA), this method offered a new feature for spectrophotometric analysis, which is the identification of the purity of the spectral profiles through the calculation of spectral contrast angle (Ɵ) in addition to the resolution of absorbance signals using percentages of bias and recovery. The proposed method was timesaving and economic since no toxic organic solvents were used when compared to the chromatographic methods. The developed method does not require exhaustive treatment for resolving the complex mixtures in quality control laboratories. In addition, the developed method is direct, simple, and do not need separation, specific detector, or pretreatment steps. Validation of the proposed methods were performed in reference to the ICH guidelines regarding the linearity, range, accuracy, precision, and specificity, and all the obtained results were within the acceptable range. The results were satisfactory compared to that of the reported method [12] of pure powdered form of CIP using Student’s t test, variance ratio F test at 95%confidence interval, where the calculated results were lower than that of the tabulated results which indicates that there is no influential difference with regard to accuracy and precision. The results confirm the applicability of the suggested method for the routine determination of the drug of interest CIP in quality control laboratories.

 

REFERENCES:

1.      C. Wilson, O. Gisvold, and R. F. Doerge, Organic Medicinal and Pharmaceutical Chemistry. 2004.

2.      British Pharmacopoeia Commission., British pharmacopoeia 2013. London: Stationery Office, 2012.

3.      H. Mandil, A. A. Sakur, and B. Nasser, “New ion selective electrode for potentiometric determination of gatifloxacin in pure form and pharmaceutical formulations,” Int. J. Pharm. Pharm. Sci., vol. 5, no. 2, pp. 423–428, 2012.

4.      A. A. Sakur, H. A. Dabbeet, and I. Noureldin, “Novel Drug Selective Sensors for Simultaneous Potentiometric Determination of both Ciprofloxacin and Metronidazole in Pure form and Pharmaceutical Formulations,” Res. J. Pharm. Technol., vol. 12, no. 7, pp. 3377–3384, 2019.

5.      G. Kawas, M. Marouf, O. Mansour, and A. A. Sakur, “Analytical Methods of Ciprofloxacin and its Combinations Review Analytical Methods of Ciprofloxacin and its Combinations Review,” Res. J. Pharm. Technol., vol. 11, no. 5, pp. 2139–2148, 2018.

6.      R. H. Obaydo and A. A. Sakur, “Spectrophotometric strategies for the analysis of binary combinations with minor component based on isoabsorptive point’s leveling effect: An application on ciprofloxacin and fluocinolone acetonide in their recently delivered co-formulation,” Spectrochim. Acta - Part A Mol. Biomol. Spectrosc., vol. 219, pp. 186–194, 2019.

7.      H. M. Lotfy, N. Hassan, S. M. Elgizawy, and S. S. Saleh, “Comparative study of new spectrophotometric methods; An application on pharmaceutical binary mixture of ciprofloxacin hydrochloride and hydrocortisone,” J. Chil. Chem. Soc., vol. 58, no. 3, pp. 1892–1898, 2013.

8.      H. Mandil, A. A. Sakur, and B. Nasser, “Potentiometric determination of Gatifloxacin and Ciprofloxacin in pharmaceutical formulations,” Int. J. pharmacutical Sci., vol. 4, no. 4, pp. 537–542, 2012.

9.      B. Nasser, “Determination of Some Antibiotic (Ciprofloxacin, Norfloxacin and Gatifloxacin) in Pure Form and Pharmaceuticals Formulations Using Electrochemical Titrations" Ph. D. Thesis, Fac. of sci.,Aleppo Univ.  ; Supervisor H. MANDIL and A.A. SAKUR, 2012.

10.   R. Kazan, “Determination of some Antibiotic of fluoroquinolone derivative In Pharmaceutical preparations using Spectrophotometric Analysis”, M. Sc. Thesis, Fac. of sci.,Aleppo Univ.Supervisor H. MANDIL and A.A. SAKUR, 2009.

11.   K. A. M. Attia, M. W. I. Nassar, M. B. El-zeiny, and A. Serag, “Stability-indicating methods for the analysis of ciprofloxacin in the presence of its acid induced degradation product: A comparative study,” Spectrochim. Acta Part A Mol. Biomol., vol. 159, no. 2016, pp. 219–222, 2016.

12.   B. Aksoy, I. Küçükgüzel, and S. Rollas, “Development and validation of a stability-indicating HPLC method for determination of ciprofloxacin hydrochloride and its related compounds in film-coated tablets,” Chromatographia, vol. 66, no. 1, pp. 57–63, 2007.

13.   B. K. Singh, D. V. Parwate, S. Srivastava, and S. K. Shukla, “High performance thin-layer chromatographic selective and stability indicating method for assay of ciprofloxacin in pharmaceuticals,” Der Pharma Chem., vol. 2, no. 4, pp. 178–188, 2010.

14.   “Guidance for Industry Q2B Validation of Analytical Procedures: Methodology,” 1996.

15.   H. M. Lotfy, Y. M. Fayez, E. M. Morgan, E. Shokry, and M. Abdelkawy, “Comparative Study of the Spectral Resolution Efficiency of the Recently Developed and Conventional Spectrophotometric Methods in the Analysis of Severely Overlapped Zero-Order Absorption Spectra with the Same Geometrical Features,” Anal. Chem. Lett., vol. 6, no. 5, pp. 450–560, 2016.

16.   J. N. Miller and M. J. C., Statistics and Chemometrics for Analytical Chemistry, vol. Sixth Edit. 2010.

17.   S. S. Toubar, M. A. Hegazy, M. S. Elshahed, and M. I. Helmy, “Novel pure component contribution, mean centering of ratio spectra and factor based algorithms for simultaneous resolution and quanti fi cation of overlapped spectral signals: An appli,” Spectrochim Acta Part A, vol. 163, pp. 89–95, 2016.

18.   H. M. Lotfy, S. S. Saleh, N. Y. Hassan, and S. M. Elgizawy, “Univariate versus Multivariate Spectrophotometric Methods for Simultaneous Determination of Complex Binary Mixtures with Overlapped Spectra: A Comparative Study,” Anal Chem Lett, vol. 3, no. September 2013, pp. 70–84, 2013.

19.   J. J. Workman, P. R. Mobley, B. R. Kowalski, and R. Bro, “Review of chemometrics applied to spectroscopy: 1985-95, Part I,” Appl. Spectrosc. Rev., vol. 31, no. 1–2, pp. 73–124, 1996.

20.   S. El-malla, A. Kamal, and S. Hammad, “A review on UV spectrophotometric methods for simultaneous multicomponent analysis,” Eur. J. Pharm. Med. Res. Med. Res., vol. 3, no. 2, pp. 348–360, 2016.

21.   A. M. Mohsen, H. M. Lotfy, A. M. R. M. Badawey, H. Salem, and S. Z. Elkhateeb, “Application of Three Novel Spectrophotometric Methods Manipulating Ratio Spectra for Resolving A Pharmaceutical Mixture of Chlorphenoxamine Hydrochloride and Caffeine,” Int. J. Pharm. Pharm. Sci., vol. 5, no. 1, pp. 478–487, 2013.

22.   C. E. Miller, Chemometrics in Process Analytical Technology. 2010.

23.   M. Nebsen and G. M. Elsayed, “Determination of Ofloxacin and Dexamethasone in Dexaflox eye drops through different ratio spectra manipulating methods,” Bull. Fac. PHARMACY, CAIRO Univ., 2013.

24.   M. A. M. Hegazy, “A novel pure component contribution algorithm (PCCA) for extracting components’ contribution from severely overlapped signals; an application to UV-spectrophotometric data,” Spectrochim. Acta Part A Mol. Biomol. Spectrosc., vol. 151, pp. 405–414, Dec. 2015.

25.   R. H. Obaydo and A. A. Sakur, “A Green Analytical Method using Algorithm (PCCA) for Extracting Components’ Contribution from Severely Overlapped Spectral Signals in Pharmaceutical Mixtures,” Res. J. Pharm. Technol., vol. 12, no. 9, 2019.

26.   S. Mowaka, M. A. Hegazy, H. M. Lotfy, and E. H. Mohamed, “Novel pure component contribution algorithm (PCCA) and UHPLC methods for separation and quantification of amlodipine, valsartan, and hydrochlorothiazide in ternary mixture,” J. AOAC Int., vol. 100, no. 3, 2017.

27.   H. M. Lotfy and S. S. Saleh, “Recent Development in Ultraviolet Spectrophotometry Through The Last Decade (2006 – 2016): A Review,” interInternational J. Pharm. Pharm. Sci., vol. 8, no. 10, pp. 40–56, 2016.

28.   H. M. Lotfy and S. S. Saleh, “Testing the purity of spectral profiles: Finger-print resolution of complex matrices and extraction of absorbance signals,” Spectrochim. Acta Part A Mol. Biomol. Spectrosc., vol. 205, pp. 160–169, 2018.

29.   R. H. Obaydo and A. A. Sakur, “The Fingerprint Spectrophotometric Methods for the Determination of Co-Formulated Otic Solution of Ciprofloxacin and Fluocinolone Acetonide in their Challengeable Ratio,” Journal of Analytical Methods in Chemistry, vol. 2019, pp. 1–14, 2019.

 

 

 

 

Received on 30.12.2019           Modified on 06.03.2020

Accepted on 08.05.2020         © RJPT All right reserved

Research J. Pharm. and Tech. 2020; 13(12):5999-6006.

DOI: 10.5958/0974-360X.2020.01046.X