Validation of Simple and Rapid UV-Spectrophotometric Method with Stress Degradation Study for Sildenafil Citrate
Shrikrishna Baokar*, Vinod Pawar, R.N. Patil, Rashmi Jagatap and Netrali Ekatpure
Shivnagar Vidya Prasarak Mandal’s College of Pharmacy, Malegaon (Bk II), Tal-Baramati, Dist- Pune, Maharashtra, India- 413115
Corresponding author: krishnabaokar@gmail.com
ABSTRACT:
A Simple, sensitive, specific, spectrophotometric method has been developed for the detection of Sildenafil citrate in pure form and Pharmaceutical formulations. The optimum condition for the analysis of the drug was established. Sildenafil citrate exhibiting absorption at 228nm and obeyed beers law in the concentration range 10 to 50µg/ml. The correlation coefficient was found to be 0.9998 and slope of line was found to be 0.07877. The percent S.D. for intra assay precision of the method was found to be 1.9976% whereas Inter assay precision was found to be 0.8332%. The sample solution was stable up to 24 hours. The assay results were found to be in good agreement with label claim. The degradation study was checked at different conditions like with acid, alkali, dry heat, oxidative and photolytic. All the results found in degradation study were satisfactory. So this proposed method was simple sensitive, precise, quick and useful for routine quality control.
KEYWORDS: Spectrophotometer, Sildenafil citrate, Validation.
INTRODUCTION:
Sildenafil citrate (Viagra®), used for the treatment of erectile dysfunction (ED), is rapidly becoming one of the most popular and widely used drugs throughout the United States and Europe. The extensive use of sildenafil is exemplified by the fact that 6 million prescriptions for this agent were written during the first 6 months following its introduction1. Contrary to popular belief, sildenafil is not an aphrodisiac, does not work in the absence of sexual arousal and does not make a potent man more virile2. Sildenafil citrate is designated chemically as 1-[[3-(6,7-dihydro-1-methyl-7-oxo-3-propyl-1H-pyrazolo[4,3-d]pyrimidin -5-yl)-4-ethoxyphenyl]sulfonyl]-4-methylpiperazine citrate.
Viagra is rapidly absorbed after oral administration, with absolute bioavailability of about 40% its pharmacokinetics are dose-proportional over the recommended dose range. It is eliminated predominantly by hepatic metabolism and is converted to an active metabolite with properties similar to the parent, Sildenafil3. Few papers have been reported for determination of sildenafil citrate include, voltametry4,5, liquid chromatography6, high performance liquid chromatography7,8 and spectrophotometry9-12. Derivative spectrophotometry has proved to be a good approach for determination of a drug in presence of matrix interference13, another drug14 or degradation product15.
Consequently, a first derivative would eliminate a constant interference and a second derivative would eliminate a linear interference.
Recently, one spectrophotometric method reported for determination of sildenafil citrate via its reaction with bromocresol green and with chromoxane cyanine R in aqueous acidic buffer solution12. Except this work, no data were available in the literature for the selective determination of the cited drug.
Fig No. 1 Structure for Sildenafil citrate
Moreover, the present work aims at developing a simple and rapid spectrophotometric method for the determination of sildenafil citrate (Viagra®) in bulk power and in pharmaceutical dosage forms with its stress degradation study.
EXPERIMENTATION:
INSTRUMENTATION:
A double–beam spectrophotometer Shimadzu UV-VIS 1700 pharmaspec serial no. 11024302261 LP was used for the detection of absorbance, AFCOSETFX-300 as Weighing balance and Ultra pure sonicator, borosil glass apparatus were used for experimental purpose.
CHEMICALS AND REAGENTS:
Sildenafil citrate working standard was supplied by Shalina Laboratories, jejuri. Marketed sample for the analysis which brought from local pharmacies Kamagra was manufactured by The Cipla Pharmaceuticals. All other chemicals used in the analysis were AR grade.
STANDARD SOLUTION:
Accurately 142.4 mg Sildenafil citrate was weighed and it was diluted with 100 ml methanol. From the resulting solution 1ml was taken and made up to 100 ml to give 10 ppm concentration of Sildenafil citrate and its absorbance was recorded.
SAMPLE SOLUTION:
20 tablets were weighed and powdered. From this, powder equivalent to 0.5 gm of Sildenafil citrate was taken and it was extracted with methanol and then the resulting solution is made up to 200 ml with Methanol. From the resulting solution 10 ml was taken and made up to 100 ml. then pipette out 10 ml from above solution dilute up to 100 ml with methanol, then absorbance of resulting solution was recorded
SELECTION OF WAVELENGTH:
The wavelength for analysis of Sildenafil citrate was selected from UV spectrum which is shown in following fig. Highest absorbance was found at 228 nm.
Fig. No. 2 Selection of wavelength
METHOD VALIDATION
Validation of the analytical method for the determination of Sildenafil citrate in pure form and in pharmaceutical formulation was carried out as per ICH guidelines.
Table No. 1 Validation Parameters for Lumefantrine
|
Parameters |
Lumefantrine |
|
Measured Wavelength |
228 nm |
|
Linearity Range |
2-10 ppm |
|
Slope |
0.07877 |
|
Method Precision % RSD |
0.1175 |
|
Correlation Coefficient |
0.9998 |
LINEARITY:
The method was validated according to ICH Q2B guidelines for validation of analytical procedures in order to determine the linearity, sensitivity, precision and of the analyte16, 17, 18, 19. For Sildenafil citrate, five point calibration curves were generated with the appropriate volumes of the working standard solutions for UV method.
Table No. 2 Linearity
|
Sr. No. |
Concentration (ppm) |
Absorbance |
|
1 |
2 |
0.167 |
|
2 |
4 |
0.337 |
|
3 |
6 |
0.49 |
|
4 |
8 |
0.643 |
|
5 |
10 |
0.802 |
|
|
Mean |
0.4878 |
|
|
SD |
0.2229 |
|
|
RSD |
0.4569 |
|
|
Correl.coif. |
0.9998 |
|
|
Slope |
0.0788 |
|
|
Intercept |
0.015 |
Fig. No. 3 Linearity curve
ACCURACY
The accuracy of an analytical method is the closeness of the test results obtained by that method to the true value20. The accuracy of an analytical method is defined as the degree to which the determined value of analyte in a sample corresponds to the true value
PRECISION:
The precision of an analytical procedure expresses the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions21. Precision of the method was demonstrated by intraday and interday variation studies. In intraday variation six different solution of same conc. 20 ppm were analyzed three times in a day i. e. from morning, afternoon and evening and the absorbance is noted.
Table No. 3 Accuracy:-
|
Sr. No. |
Concentration (ppm) |
Absorbance |
Mean |
SD |
RSD |
%RSD |
|
1 |
80% |
2.241 |
2.2453 |
0.0045 |
0.002004 |
0.2004 |
|
2.250 |
||||||
|
2.245 |
||||||
|
2 |
120% |
3.436 |
3.429 |
0.0157 |
0.004578 |
0.4578 |
|
3.440 |
||||||
|
3.411 |
||||||
|
3 |
100% |
4.645 |
4.6616 |
0.0208 |
0.004461 |
0.4461 |
|
4.655 |
||||||
|
4.685 |
Table No. 4 Intra-assay Precision
|
Sr. No |
Conc. (ppm) |
Abs-I |
Abs-II |
Abs-III |
Avg %RSD |
|
1 |
7 |
0.580 |
0.585 |
0.596 |
|
|
2 |
7 |
0.596 |
0.586 |
0.590 |
|
|
3 |
7 |
0.584 |
0.599 |
0.587 |
|
|
4 |
7 |
0.599 |
0.582 |
0.585 |
|
|
5 |
7 |
0.582 |
0.585 |
0.588 |
|
|
6 |
7 |
0.587 |
0.580 |
0.581 |
|
|
|
Avg |
0.588 |
0.586 |
0.587 |
|
|
SD |
0.0794 |
0.00667 |
0.005037 |
||
|
RSD |
0.0135 |
0.01139 |
0.008580 |
||
|
%RSD |
1.35 |
1.1139 |
0.8580 |
0.8332 |
In the inter day variation study , solution of same conc. 20 ppm were analyze three time for the consecutive days and the absorbance result mean , standard deviation and % RSD was calculated and given in table.
Table No. 5 Inter-assay precision
|
|
|
Sr. No |
Conc. (ppm) |
Day -I |
Day-II |
Day III |
|
1 |
20 |
1.686 |
1.720 |
1.750 |
|
2 |
20 |
1.699 |
1.710 |
1.739 |
|
3 |
20 |
1.702 |
1.700 |
1.755 |
|
4 |
20 |
1.704 |
1.698 |
1.699 |
|
5 |
20 |
1.694 |
1.722 |
1.665 |
|
6 |
20 |
1.690 |
1.699 |
1.724 |
|
|
Avg |
1.698 |
1.7081 |
1.722 |
|
SD |
0.0070 |
0.0108 |
0.0344 |
|
|
RSD |
0.004122 |
0.006322 |
0.0199 |
|
|
%RSD |
0.4122 |
0.6322 |
1.9976 |
RUGGEDNESS:
The ruggedness of an analytical method is degree of reproducibility of test results obtained by the analysis of the same samples under a variety of normal test conditions, such as different laboratories, different analysts, different instruments, different lots of reagents, different elapsed assay times, different assay temperatures, different days etc.
Table No. 6 Ruggedness
|
Sr. No |
Parameter |
Set I |
Set II |
|
1 |
System |
Shimadzu-1700 |
Systronics -119 |
|
2 |
Sample |
Batch No-X |
Batch No –Y |
|
3 |
Day |
Monday |
Tuesday |
|
4 |
Date |
30/01/2011 |
31/01/2011 |
|
5 |
Time |
3.45pm |
3.45 pm |
|
6 |
Lab |
Analysis |
Chemistry |
|
7 |
Analyst |
19/07 |
24/08 |
|
8 |
Sample |
20ppm |
20ppm |
|
9 |
Absorbance |
1.686 |
1.680 |
|
10 |
Assay |
99.10% |
99.20% |
ROBUSTNESS:
The evaluation of robustness should be considered during the development and development is on the type of procedure under study. Robustness of the method was checked by making slight deliberate changes in selected conditions like change in wavelength. It was observed that there were no marked changes in chromatograms, which demonstrated that the developed method is robust22. Instruments are susceptible to variations in analytical conditions, should be suitability controlled or a precautionary statement should be including in the procedure.
DEGRADATION STUDIES:23
The International Conference on Harmonization (ICH) guideline entitled stability testing of new drug substances and products requires that stress testing be carried out to elucidate the inherent stability characteristics of the active substance. The aim of this work was to perform the stress degradation studies on the Sildenafil Citrate using the proposed method 4.
Hydrolytic degradation under acidic condition:
To 3 ml of stock solution (1000 μg/ml) of Sildenafil Citrate, 1 ml of 3 N HCl was added in 10 ml of volumetric flask and the volume was made up to the mark with methanol. Then, the volumetric flask was kept at normal condition for 90 minutes. After 60 min. time interval, 1 ml of solution was pipettes out from this flask, neutralized and diluted with methanol in order to make the volume up to 10 ml and the dilution was carried out to achieve the appropriate concentration (20 μg/ml). This solution was taken in cuvette. For the blank, 0.5 ml solution of 3N HCl and 0.5 ml solution of 3N NaOH were diluted with methanol in 10 ml of volumetric flask was repeated. After 90 min. time interval the same procedure is repeated and the dilution was carried out to achieve the appropriate concentration (20μg/ml). This solution was taken in cuvette and absorbance is recorded against blank solution.
Hydrolytic degradation under alkaline condition:
To 3 ml of stock solution of Sildenafil Citrate, 1 ml of 0.1 N NaOH was added in 10 ml of volumetric flask and made up the volume to the mark with methanol. Volumetric flask was kept at normal condition for 90 min. After 90 min time interval, 1 ml of solution was pipettes out from this flask, neutralized and diluted with methanol in order to make the volume up to 10 ml and the dilutions were carried out to achieve the appropriate concentration (20 μg/ml).
Table No. 7 Robustness
|
Sr. No |
Conc.(ppm) |
Set No. |
Absorbance |
Wavelength |
Mean |
SD |
RSD |
%RSD |
|
1
|
20
|
I |
1.192 |
223 |
1.1923 |
0.002517 |
0.002110 |
0.211 |
|
1.190 |
||||||||
|
1.195 |
||||||||
|
2
|
20
|
II |
1.122 |
228 |
1.123667 |
0.002082 |
0.0018528 |
0.185 |
|
1.123 |
||||||||
|
1.126 |
||||||||
|
3 |
20 |
III |
1.002 |
223 |
1.0033 |
0.001528 |
0.0015229 |
0.152 |
|
1.005 |
||||||||
|
1.003 |
The solution was then taken in cuvette. For the blank, 0.5 ml solution of 0.1N HCl and 0.5 ml solution of 0.1N NaOH diluted with methanol in 10 ml of volumetric flask. After 90 min. time interval the same procedure is repeated and the dilution was carried out to achieve the appropriate concentration (20μg/ml). This solution was taken in cuvette and absorbance is recorded against blank solution.
Dry heat induced degradation:
Sildenafil Citrate sample was taken in a Petri plate and exposed to a temperature of 70°c for 48 hours in an oven. After 48 hours, 10 mg of the sample was diluted with methanol in order to make the volume up to 10 ml. From this solution, dilutions were carried out to achieve the appropriate concentration (20μg/ml) and the solution was taken in cuvette for the UV-Vis analysis.
Oxidative degradation:
To 1.5 ml of the stock solution of Sildenafil Citrate (1000 μg/ml), 1 ml of 30 % w/v of hydrogen peroxide added in 10 ml of volumetric flask and the volume was made up to the mark with methanol . The volumetric flask was then kept at room temperature for 15 min. For the blank, 1 ml of the 30 % w/v of hydrogen peroxide was kept at normal condition for overnight in 10 ml of volumetric flask. Both solutions were heated on boiling water bath to remove the excess of hydrogen peroxide. Finally after 15 minutes dilutions were made from the stock solution to achieve the required concentration (30μg/ml). The solution was then taken in a cuvette and analyzed.
Photolytic degradation:
Sample of Sildenafil Citrate was exposed to near ultra violet lamp in photostablity chamber providing illumination of not less than 1.2 million lux hours. 10 mg sample was dissolved methanol and volume made up to 10 ml. From this solution appropriate dilution (20μg/ml) was made using methanol and taken in cuvette for the UV analysis.
Table No. 8 Stress Degradation study
|
Condition |
Time |
% Degradation |
|
Hydrolytic degradation of 3 N HCl |
60 min |
98.54 |
|
90 min |
93.43 |
|
|
Hydrolytic degradation of 0.1 N NaOH |
60 min |
97.87 |
|
90 min |
95.50 |
|
|
Photolytic degradation |
3 Hrs |
96.43 |
|
6 Hrs |
93.75 |
|
|
Dry Heat degradation |
48 Hrs. |
95.60 |
|
Oxidative degradation |
15 min |
97.32 |
RESULTS AND DISCUSSION:
Ø Linearity of the drug was obtained in the range of 10ppm to 50ppm of Sildenafil Citrate.
Ø The linearity coefficient and percentage curve fitting was found to be 0.9998 and 99.98% for Sildenafil Citrate.
Ø Accuracy of the method was determined through the recovery studies of the drugs. Recovery of the drugs was well within the acceptance limit (99% - 101%).
Ø Precision of the method was determined by analyzing the drug formulation by replicate concentrations.
Ø Percent RSD of the analyte was found to be within the limit of 2%, thus the developed method was found to provide high degree of precision and reproducibility.
Ø Ruggedness was determined by performing the assay with same condition on different days, by different analysts, different instrument and different day and time. The test results were found within limit 99 – 101%. The results were found to be reproducible, in spite of variations in conditions which could be normally expected from analysts to analysts.
Ø Robustness was determined by carrying out the assay during change Wavelength . Percent RSD was found to be within the limit NMT 2%. The values of RSD obtained with the change in wavelength ratio makes it possible to carry out the method for Sildenafil Citrate with a small variation in wavelength.
CONCLUSION:
Ø UV method is developed for the Sildenafil Citrate by using UV-Vis spectrophotometer
Ø Maxima for Sildenafil Citrate detected by UV detector at 228 nm.
Ø The developed method is applied for the Sildenafil Citrate from tablet form. The assay results are within the limit.
Ø The percentage purity of Sildenafil Citrate found to be 99.85
Ø The developed method is validated with various parameters as per ICH guidelines like accuracy, precision, linearity, specificity, ruggedness and robustness.
Ø The results obtained are within the acceptance criteria
Ø Hence the developed method is found to be satisfactory and would be used for the routine analysis in the laboratory for Sildenafil Citrate in pharmaceutical formulation
REFERENCES:
1. Schwarz U.R., Walter U., and Eigenthaler M., (2001), Biochem Pharmaco 62(9): 1153.
2. Terrett K., et.al., (1996), Bio. Med. Chem. Lett. 6(15): 1819.
3. Ichiro S., et.al., (2002), Europ. J. Pharmaco. 437(3): 155.
4. Berzas J.J., Rodriguez J., Castañeda G., and Villaseñor M.J., (2000), Anal Chimica Acta, 417(2): 143.
5. Rodr´guez J., Berzas J.J., Castan~eda G., and Rodr´guez N., (2004), Talanta 62:2427.
6. Dinesh N.D., Vishukumar B.K., Nagaraja P., Made Gowda N.M., and Rangappa K.S., (2002), J. Pharm. Biomed. Anal. 29(4): 743.
7. Daraghmeh N., Al-Omari M., Badwan A.A., and Jaber AMY, (2001), J. Pharm. Biomed. Anal., 25(3-4): 483.
8. Shim H., Lee E.J., Jung Y.H., Kim S.H., Kim S.H., Yoo M., Kwon J.W., Kim W.B., and Lee M.G., (2002), J. Pharm. Biomed. Anal. 30(3): 527.
9. Altokka G., Atkosar Z., Sener E., and Tunçel M., (2001), J. Pharm. Biomed. Anal., 25(2): 339.
10. Al-bojuk, et.al., (2001), Heterocycles. 55(9): 1789.
11. El-Abadelah, et.al., (2002), J. Heterocycl. Chem. 39(5): 1055.
12. Dinesh N.D., Nagaraja P., Made Gowda N.M., and Rangappa K.S., (2002), Talanta, 57(4): 757.
13. Mabrouk M.M., Salem H., Radwan M.F., and Kaood T.S., (2003), Egypt J. Biomed. Sci. 12: 141.
14. Salem H., (1999), Chinese Pharm. J., 51: 123.
15. Salem H., El-Maamli M., El-Sadek M., and Aboul-Kheir A., (1994), Zag. J. Pharm. Sci., 3(1): 115.
16. F.W. Fifield and D. Kealey, 5th Edition, Black Well Science Ltd. Principles and Practice of Analytical Chemistry, (2000) pp. 270 – 276.
17. C. Hokanson, A Life Cycle Approach to the Validation of Analytical Methods During Pharmaceutical Prod-uct Development, Part – II: Changes and the Need for Additional Validation, Pharm. Tech., 92 -100 (1994).
18. J.M. Green, A Practical Guide to Analytical Method Validation, Anal. Chem. News and Features, 305A – 309A (1996)
19. J. Vessman, Selectivity or Specificity? Validation of Analytical Methods from the Perspective of an Analytical Chemist in the Pharmaceutical Industry, J. Pharm and Biomed. Anal., 14, 867 – 869 (1996).
20. Shrikrishna B. Baokar, B. Shirke, V. Sivanand, G.K. Pratheesh, Analytical method development and validation for esteemation of sildenafil citrate from tablet dosage form by RP-HPLC, Int. J. Res. Phar.Sci. 2(2), 2011, 130-136.
21. Baokar Shrikrishna, Pawar Vinod , Sonawane S.H., High Performance Liquid Chromatographic Method Development and Validation of Cholesterol Inhibitor Drug.Journal of Pharmacy Research, 2011,4(7),2313-2316
22. Baokar Shrikrishna B., Erande R.S., Shaikh S.G. Analytical Method Development and Validation for Estimation of Ezetimibe from Tablet Dosage Form by Using RP-HPLC, Int. J. of Res. in Pharm. and Biomed. Sci., Vol. 2 (2), Apr – Jun 2011, 833-841
23. Suddhasattya dey, Y.Vikram reddy, T.Reddy, Method development abd validation for estimation of atazanvir in bulk and pharmaceutical dosage forms and its stress degradation studies using UV-VIS spectrophotometric method, Int. J. of Pharm. And Biosci, Vol.1, Issue 3, july-sep.2010.
Received on 27.09.2011 Modified on 11.10.2011
Accepted on 16.10.2011 © RJPT All right reserved
Research J. Pharm. and Tech. 5(2): Feb. 2012; Page 214-218